IMAGE IN CARDIOLOGY

DOI: 10.4244/EIJY14M07_07

Percutaneous extraction of pacing leads from the left coronary artery and left ventricle

Sammy Elmariah1, MD, MPH; Christopher J. Mutrie2, MD; Praveen Mehrotra3, MD; Brett Carter4, MD; Douglas E. Drachman1, MD; Jennifer D. Walker2*, MD

An 80-year-old female was transferred to our institution after undergoing permanent pacemaker implantation complicated by transarterial lead placement. Imaging revealed the misplaced ventricular lead within the left ventricle and the “atrial” lead within the left main coronary artery (Online Figure 1, Moving image 1). The ventricular lead was removed without complication. Angiography and echocardiography confirmed the “atrial” lead at the junction of the left anterior descending and ramus intermedius arteries (Figure 1, Moving image 2). We advanced two guiding catheters loaded with covered stents to the ascending aorta to prepare for possible coronary perforation. After removal of the “atrial” lead, a hazy filling defect was observed within the left main coronary artery which was retrieved using manual thrombectomy (Online Figure 2). The leads were then removed from the left subclavian artery and a covered stent deployed (Moving image 3).

Figure 1. Pacemaker leads within the left ventricle and left main coronary artery. A) Coronary angiography, and B) & C) transoesophageal echocardiography images demonstrating placement of the “atrial” lead within the left coronary system. Echocardiography was performed to monitor for the development of tamponade upon extraction of the lead. Images revealed both leads in the aortic root with one directed towards the left cusp (arrow) and the other crossing through the aortic valve (arrowhead).

Erroneous introduction of a pacing lead into the subclavian artery is a rare complication that in our case was in part due to the use of a micropuncture needle for vascular access. We demonstrate the efficacy of a multidisciplinary team approach by which malpositioned leads were safely explanted from the left ventricle and coronary system without the need for sternotomy and without evidence of myocardial injury.

Conflict of interest statement

The authors have no conflicts of interest to declare.

Online data supplement

Moving image 1. Fluoroscopic imaging during temporary right ventricular pacing wire implantation clearly demonstrates the right ventricular lead following the superior vena cava to right ventricle contour. In contrast, the malpositioned wires follow the aortic contour.

Moving image 2. Angiography revealed a trifurcating left main coronary artery with the malpositioned “atrial” pacemaker lead terminating at the branch point of the left anterior descending and ramus intermedius coronary arteries. The pacing lead screw appears fixated to the vessel wall with its distal end exiting the vessel.

Moving image 3. Under fluoroscopic guidance, the pacing leads were quickly removed from the left subclavian artery. A polytetrafluoroethylene (PTFE)-covered self-expanding 8 mm×5.0 cm stent was deployed via a long sheath from the right femoral artery to repair the subclavian arteriotomy.

Online Figure 1. Radiographic appearance of dual chamber pacemaker leads within the left heart. A) Frontal chest radiograph demonstrates the presence of a dual chamber pacemaker with the leads extending above the medial aspect of the left clavicle, consistent with a left subclavian arterial course. The atrial lead projects to the left of the thoracic spine (white arrow) and the ventricular lead projects over the expected position of the left ventricle (black arrow). B) Coronal reformatted images from a contrast-enhanced chest CT showed the atrial lead extending into the proximal left main coronary artery (white arrow). C) The ventricular lead extended across the aortic valve to terminate within the left ventricle (white arrow).

Online Figure 2. Filling defect on angiography and intravascular ultrasound within the left main coronary artery found to be large wire-associated thrombus. A) Upon removal of the pacing lead, a filling defect was observed within the mid portion of the left main coronary artery. B) Intravascular ultrasound confirmed the presence of a filling defect as opposed to a coronary artery dissection, perforation, or other pathologic process. C) Manual thrombectomy retrieved a large wire-associated thrombus from the left coronary system. FD: filling defect

Volume 11 Number 3
Jul 20, 2015
Volume 11 Number 3
View full issue


Key metrics

Suggested by Cory

IMAGE IN CARDIOLOGY

10.4244/EIJV11I7A169 Nov 20, 2015
An atypical clinical presentation of a broken guidewire left in the venous system
Montero-Cabezas J et al
free

10.4244/EIJV10I3A67 Jul 21, 2014
How should I treat a patient with an entrapped infected permanent pacemaker lead?
Asmarats L et al
free

IMAGE IN CARDIOLOGY

10.4244/EIJV12I6A121 Aug 20, 2016
Optical coherence tomography visualisation of burst balloon catheter trapped by coronary stent
Shimura T et al
free

10.4244/EIJV10I61A131 Oct 20, 2014
How should I treat an accidentally misplaced 8 Fr drainage catheter in the right ventricle?
Bakos Z and Harnek J
free

10.4244/EIJV7I4A83 Aug 30, 2011
How should I treat a patient to remove a fractured jailed side branch wire?
Owens C and Spence M
free

Jan 15, 2010
Femoral artery puncture
Haude M et al
free
Trending articles
318.1

State-of-the-Art Review

10.4244/EIJ-D-21-00695 Nov 19, 2021
Transcatheter treatment for tricuspid valve disease
Praz F et al
free
116.75

State-of-the-Art

10.4244/EIJ-D-24-00066 Apr 21, 2025
Management of complications after valvular interventions
Bansal A et al
free
108.3

Viewpoint

10.4244/EIJ-E-22-00007 May 15, 2022
TAVI at 20: how a crazy idea led to a clinical revolution
Eltchaninoff H et al
free
91.6

Image – Interventional flashlight

10.4244/EIJ-D-22-00344 Aug 5, 2022
First dedicated transcatheter leaflet splitting device: the ShortCut device
Tchétché D et al
free
72.4

State-of-the-art

10.4244/EIJ-D-22-00627 Feb 6, 2023
Left atrial appendage occlusion
Holmes D et al
free
68.9

State-of-the-Art

10.4244/EIJ-D-24-00992 Sep 15, 2025
Antithrombotic therapy in complex percutaneous coronary intervention
Castiello D et al
free
60.65

Clinical research

10.4244/EIJ-D-20-01155 Oct 20, 2021
A deep learning algorithm for detecting acute myocardial infarction
Liu W et al
free
49.55

CLINICAL RESEARCH

10.4244/EIJ-D-17-00962 Apr 6, 2018
A new optical coherence tomography-based calcium scoring system to predict stent underexpansion
Fujino A et al
free
43.45

State-of-the-Art Review

10.4244/EIJ-D-21-00145 Sep 20, 2021
Robotics, imaging, and artificial intelligence in the catheterisation laboratory
Beyar R et al
free
X

PCR
Impact factor: 9.5
2024 Journal Citation Reports®
Science Edition (Clarivate Analytics, 2025)
Online ISSN 1969-6213 - Print ISSN 1774-024X
© 2005-2025 Europa Group - All rights reserved