Three-dimensional optical coherence tomography assessment of coronary wire re-crossing position during bifurcation stenting

Takayuki Okamura*, MD, PhD; Jutaro Yamada, MD, PhD; Tomoko Nao, MD, PhD; Takeshi Suetomi, MD; Takao Maeda, MD; Kohzoh Shiraishi, MD; Toshiro Miura, MD, PhD; Masunori Matsuzaki, MD, PhD

In a bifurcation intervention, coronary wire re-crossing through an appropriate cell of the jailing strut is important before kissing balloon post-dilation (KBPD). We assessed the re-crossing position of the wire employing 3-dimensional optical coherence tomography (3D-OCT) image reconstruction.

A 69-year-old male who had presented with angina pectoris underwent percutaneous coronary intervention (PCI). Baseline coronary angiography revealed a napkin-ring narrowing at the ostium of the left circumflex coronary artery (LCx, Figure 1A). After deployment of a 3.0/15 mm zotarolimus-eluting Endeavor® stent (Medronic Inc., Santa Rosa, CA, USA) across the left anterior descending coronary artery (LAD), two wires were re-crossed into the LAD and the high lateral (HL) branch (Figure 1B).

Figure 1. Coronary angiography and on-site 3D-OCT reconstruction. A) Baseline angiography shows a napkin-ring stenosis at the ostium of the circumflex coronary artery (LCx) (Arrow). B) After stent deployment, two wires were passed into the left anterior descending coronary artery (LAD) and the high lateral (HL) branch. C) On-site 3D-OCT image and schematic illustration. 3D-OCT shows that the wire was passed into the LAD, apparently through the appropriate cell. D) Final angiogram.

Fourier-domain OCT images (ILUMIEN; St Jude Medical Inc., St. Paul, MN, USA) were obtained at pullback speeds of 10 mm/s. Automatic strut detection (in-house software from Yamaguchi University, Ube, Japan) and 3D-OCT image reconstruction (INTAGE realia; Cybernet, Tokyo, Japan) for the bifurcation segment were achieved using the dedicated off-line workstation. The 3D-OCT clearly demonstrated that the coronary wire had been successfully passed into the LAD through the distal cell of the jailing strut (Figures 1C and 2A, Moving image 1). After KBPD, OCT images were obtained at the pullback speed of 10 mm/s. 3D-OCT demonstrated that the jailing strut was adequately opened and no protruding struts were seen in front of the LAD ostium (Figure 2B, Moving image 2). The final angiogram confirmed a good result (Figure 1D).

Figure 2. 3D-OCT reconstruction after procedure. A) After guidewire (GW) re-cross. The LAD ostium is jailed by the strut. Moving image 1. B) After kissing balloon post-dilation. The jailing struts are adequately opened with scaffolding to the vessel wall against the carina (yellow arrow). Moving image 2.

In order to visualise the stent clearly in 3D-OCT, an enhancement of the stent strut is necessary. However, manual strut detection is time-consuming, usually requiring one to two hours per stent. The novel automatic strut detecting program allows rapid 3D stent image reconstructions (about 10 minutes) during procedure. 3D-OCT guidance of the wire re-crossing the jailed side branch before KBPD is feasible in the interventional suite, and may improve bifurcation stenting.

Conflict of interest statement

The authors have no conflict of interest to declare.

Online data supplement

Moving image 1. 3D-OCT after guidewire re-cross.

Moving image 2. 3D-OCT after kissing balloon post-dilation.

Volume 7 Number 7
Nov 30, 2011
Volume 7 Number 7
View full issue


Key metrics

Suggested by Cory

Image – Interventional flashlight

10.4244/EIJ-D-21-00393 Apr 22, 2022
Napkin ring formation in culotte stenting using current-generation drug-eluting stents in left main coronary artery bifurcation
Murasato Y et al
free

10.4244/EIJV11SVA14 May 19, 2015
OCT for bifurcation stenting: what have we learned?
Holm N et al
free
Trending articles
152.9

Clinical research

10.4244/EIJ-D-20-01125 Oct 20, 2021
An upfront combined strategy for endovascular haemostasis in transfemoral transcatheter aortic valve implantation
Costa G et al
free
47.8

NEW INNOVATION

10.4244/EIJ-D-15-00467 Feb 20, 2018
Design and principle of operation of the HeartMate PHP (percutaneous heart pump)
Van Mieghem NM et al
free
39.1

Clinical research

10.4244/EIJ-D-22-00558 Feb 6, 2023
Permanent pacemaker implantation and left bundle branch block with self-expanding valves – a SCOPE 2 subanalysis
Pellegrini C et al
free
X

The Official Journal of EuroPCR and the European Association of Percutaneous Cardiovascular Interventions (EAPCI)

EuroPCR EAPCI
PCR ESC
Impact factor: 7.6
2023 Journal Citation Reports®
Science Edition (Clarivate Analytics, 2024)
Online ISSN 1969-6213 - Print ISSN 1774-024X
© 2005-2024 Europa Group - All rights reserved