Image – Interventional flashlight

DOI: 10.4244/EIJ-D-20-00748

Fracture of the delivery catheter system during transcatheter aortic valve implantation

Xavier Armario1, MD; Liesbeth Rosseel1, MD; Faisal Sharif1,2, MBBS, PhD; Darren Mylotte1,2, MBBCh, MD, PhD

Figure 1. Delivery catheter system fracture during TAVI. A) Moderately calcified and tortuous iliofemoral arteries. B) & C) Moderately calcified and angulated trileaflet aortic valve. D) & E) Positioning, deployment and recapture of a 34 mm Evolut R valve. F) – H) Fractured delivery catheter system. Reproduced with permission from Medtronic.

Figure 1. Delivery catheter system fracture during TAVI. A) Moderately calcified and tortuous iliofemoral arteries. B) & C) Moderately calcified and angulated trileaflet aortic valve. D) & E) Positioning, deployment and recapture of a 34 mm Evolut R valve. F) – H) Fractured delivery catheter system. Reproduced with permission from Medtronic.

An 87-year-old male with severe symptomatic aortic stenosis was accepted for transcatheter aortic valve implantation (TAVI) via transfemoral approach (Figure 1). Multislice computed tomo-graphy demonstrated moderately calcified and tortuous iliofemoral arteries (minimal lumen diameter 10.4 mm) (Figure 1A), and a moderately calcified and angulated trileaflet aortic valve (aortic annulus perimeter 85.0 mm; aortic root angulation 45º) (Figure 1B, Figure 1C), suitable for implantation of a 34 mm Evolut™ R valve (Medtronic, Minneapolis, MN, USA) as per standard practice at our centre. After vascular pre-closure with two Perclose devices (Abbott Vascular, Santa Clara, CA, USA) and valve predilatation with a 24 mm NuCLEUS™ balloon (NuMED, Cornwall, ON, Canada), a 16 Fr EnVeo™ PRO delivery catheter system (DCS) (Medtronic) easily traversed the iliofemoral vessels and the aortic valve. Deployment of the transcatheter heart valve (THV) was challenging due to recurrent migration towards the left ventricle, despite guidewire manipulation and rapid ventricular pacing (Figure 1D, Figure 1E). During a third recapture manoeuvre, the THV would not completely re-enter the DCS capsule, and further rotation of the delivery handle (either direction) did not move the THV. The DCS was retrieved into the descending aorta where fluoroscopy revealed a fracture at the proximal end of the capsule (Figure 1F-Figure 1H). As the DCS was attached to its capsule, it was possible to remove the entire DCS en bloc via the transfemoral access by gentle retraction, without the requirement for additional manoeuvres. Surgical back-up was on stand-by in case of access-site complications. Haemostasis was achieved using the Perclose sutures. The patient suffered no additional complications and was successfully treated with a 29 mm SAPIEN 3 valve (Edwards Lifesciences, Irvine, CA, USA) one month later.

Technical analysis of the DCS by the manufacturer confirmed complete fracture of the capsule and delamination of the nitinol-reinforcing frame. The DCS was otherwise intact and functioning appropriately. As no definitive cause was confirmed, a multifactorial origin was suggested. Potential causes include excessive tension and torque on the DCS due to anatomical factors (vascular calcification, tortuosity and angulation) and repeated recapture or over-capture of the THV. As such, careful patient selection and device manipulation during deployment and retrieval are advised, and adherence to the manufacturer recommendations of a maximum of three recapture manoeuvres is advised.

Conflict of interest statement

D. Mylotte is a consultant for Medtronic, Boston Scientific, and MicroPort. The other authors have no conflicts of interest to declare.

Supplementary data

To read the full content of this article, please download the PDF.

Moving image 1. Delivery catheter system advance.

Moving image 2. Transcatheter heart valve deployment and recapture.

Moving image 3. Delivery catheter system removal.

Volume 17 Number 4
Jul 20, 2021
Volume 17 Number 4
View full issue


Key metrics

Suggested by Cory

IMAGE IN CARDIOLOGY

10.4244/EIJY15M06_01 Mar 18, 2016
How to overcome a scary complication: a mesh-trapped balloon during a CoreValve® implantation procedure
Bayet G et al
free

10.4244/EIJY14M09_04 Mar 20, 2015
How should I treat dislocation of a TAVI SAPIEN prosthesis into the left ventricle?
Tiroch K et al
free

Image – Interventional flashlight

10.4244/EIJ-D-19-00826 Sep 18, 2020
ACURATE neo valve perforation after valve-in-valve transcatheter aortic valve implantation
Rück A et al
free

10.4244/EIJV17I10A135 Nov 19, 2021
Bioprosthetic valve fracture: haemodynamically effective but safe and clinically necessary?
Rodés-Cabau J and Alperi A
free
Trending articles
87.2

State-of-the-Art

10.4244/EIJ-D-25-00266 Jan 19, 2026
Lesion stratification with intracoronary imaging
McGarvey M et al
free
47.45

NEW INNOVATION

10.4244/EIJ-D-15-00467 Feb 20, 2018
Design and principle of operation of the HeartMate PHP (percutaneous heart pump)
Van Mieghem NM et al
free
34.8

Original Research

10.4244/EIJ-D-25-01006 Mar 16, 2026
Clinical outcomes and haemodynamic response after blinded stress assessment of moderate aortic stenosis
Eerdekens R et al
22.2

Viewpoint

10.4244/EIJ-D-25-01066 May 4, 2026
Intracoronary imaging guidance for de novo coronary lesion treatment with drug-coated balloons
Amabile N et al
free
20.75

Flashlight

10.4244/EIJ-D-25-01014 Apr 6, 2026
Stent retriever-assisted coronary thrombectomy with continuous aspiration
Liabot Q et al
open access
19.5

Original Research

10.4244/EIJ-D-26-00032 May 15, 2026
Glucocorticoids to reduce permanent pacemaker implantation after TAVI: the GLUCO-TAVI randomised trial
Fuertes-Kenneally L et al
17.8

Expert Review

10.4244/EIJ-D-25-01316 Apr 20, 2026
Electrosurgical laceration and stabilisation of tricuspid edge-to-edge repair: the ELASTA-T technique
Alvarez-Covarrubias H et al
free
X

PCR
Impact factor: 9.5
2024 Journal Citation Reports®
Science Edition (Clarivate Analytics, 2025)
Online ISSN 1969-6213 - Print ISSN 1774-024X
© 2005-2026 Europa Group - All rights reserved