Bioresorbable vascular scaffolds in left main coronary artery disease

Bert Everaert1, MD, PhD; Piera Capranzano2, MD; Corrado Tamburino2, MD, PhD; Ashok Seth3, MD, FRCP, DSc; Robert-Jan M. van Geuns1*, MD, PhD

Abstract

Currently, the use of bioresorbable vascular scaffolds (BRS) for the treatment of left main (LM) coronary artery disease has to be considered investigational. However, some early evidence from case reports supports the feasibility of BRS implantation in selected cases and shows good angiographic and clinical results with current-generation BRS devices. However, before the routine use of BRS for LM disease can be advocated, more data on long-term safety and efficacy and larger scaffold designs are essential.

Introduction

Percutaneous coronary intervention (PCI) has become a reliable revascularisation option for the treatment of ischaemic coronary artery disease (CAD)1. Several recent trials have suggested that PCI with metallic drug-eluting stents (DES) is feasible and safe and has equivalent long-term outcomes compared to coronary artery bypass grafting (CABG) for the treatment of LM stenosis in patients with a low to intermediate SYNTAX score (<33)2,3.

However, these permanent implants continue to remain a foreign body, even after vascular healing following the PCI, and this leads to rates of target vessel failure of more than 2% per year4. To eliminate this and other potential limitations of a permanent metallic implant, bioresorbable coronary stents or “scaffolds” (BRS) have been developed.

Currently, the use of BRS for the treatment of left main (LM) coronary artery stenosis has not been properly studied and is therefore considered investigational. However, limited evidence from case reports supports the feasibility of BRS implantation in these patients and shows good angiographic and clinical results in the short term. With the limited expansion capacity of currently available BRS devices there is a justified concern for BRS malapposition and subsequent BRS failure particularly in this patient population. Therefore, for routine use in LM disease, larger scaffold designs and long-term safety and efficacy data are essential.

Indications for LM revascularisation

CABG is still considered as the “gold standard” treatment for most patients with LM disease. However, several studies have indicated equally good clinical outcomes comparing CABG to PCI in patients with low to intermediate SYNTAX scores (≤32)2-4.

When considering PCI for the treatment of LM CAD, besides the SYNTAX score, other lesion characteristics are predictive of a favourable clinical outcome, such as lesion location (ostial or shaft), isolated LM disease, LM in combination with single-vessel disease, plaque distribution pattern favouring a single-stent crossover technique and limited calcium burden5. Importantly, when using DES, patients have to be able to comply with dual antiplatelet therapy (DAPT) for several months.

In conclusion, with the advent of first and second-generation DES, PCI has become a valuable treatment strategy in selected patients with LM disease.

Technical considerations for BRS implantation in LM disease

Several technical considerations, pertaining to the current generation of BRS, preclude its broad use in LM PCI (Table 1).

First and most important is the limited overexpansion capability of the current generation of BRS, impairing adequate stent apposition. As with other stent platforms6, BRS malapposition and/or incomplete lesion coverage are expected to increase the risk of BRS failure, i.e., by a higher incidence of acute and chronic scaffold thrombosis and in-scaffold restenosis. As the expansion limit of the largest commercially available Absorb BVS (Absorb 3.5 mm; Abbott Vascular, Santa Clara, CA, USA) is restricted to 0.5 mm, vessels with a diameter greater than 4.0 mm (quantitatively measured by IVUS or OCT, and 3.8 mm for QCA) should not be treated because of the high risk of malapposition. As LM diameters are often considerably larger than 4.0 mm, only a small proportion of patients are suitable for BRS implantation7. Other bioresorbable devices, such as the DESolve™ myolimus-eluting bioresorbable coronary scaffold system (Elixir Medical Corporation, Sunnyvale, CA, USA) or the magnesium-based drug-eluting absorbable metal scaffold platform DREAMS (Biotronik AG, Bülach, Switzerland), are currently under investigation but could overcome some of these issues. The DESolve BRS has been designed with unique expansion capabilities, such as the ability to expand within a wide safety margin without strut fracture (the 3.0 mm scaffold can be expanded to 4.5 mm)8, which could make this a potentially useful BRS platform for LM interventions in the future, when larger scaffold diameters of the DESolve become available. Also, the DREAMS platform could become an alternative to the polymeric scaffolds for LM CAD, because the mechanical properties of the magnesium alloy are comparable to those of permanent metallic stent platforms, resulting in an increased deliverability and a reduced incidence of strut fractures9. However, initial data on the bare metal version of this latter device showed significant late lumen loss and a higher rate of target vessel revascularisation. Data from the BIOSOLVE-II study investigating the new second-generation DREAMS platform are still not available. This stent is not commercially available and larger stent diameters would be needed to treat LM stenosis.

Secondly, to minimise the risk of BRS malapposition in the LM, correct scaffold sizing based on a reliable assessment of LM dimensions should always include invasive imaging modalities such as IVUS or OCT. These techniques have proven to be superior to angiography in providing accurate estimations of LM diameter, lesion length and the involvement of the LM bifurcation and its distal side branches. Of note, OCT is particularly suited to optimising BRS implantation10,11 (i.e., to determine optimal scaffold size and length, identify optimal proximal and distal landing zones and guide post-dilatation of the scaffold to perfect strut apposition).

Thirdly, the radial strength of all PLLA-based scaffold designs is outperformed by cobalt-chromium and especially platinum-chromium stent platforms. LM PCI for ostial disease frequently suffers from acute stent recoil due to the fibroelastic properties of the aortic wall and the increased presence of calcium. To attain sufficient radial strength (on average one bar), the strut thickness of BRS has to be larger, which affects the scaffold’s crossing profile. However, with adequate predilatation and lesion preparation, this generally does not compromise the scaffold deliverability particularly in LM disease. While there is a concern that aggressive balloon dilatations within a highly calcified LM could potentially result in LM dissection with serious consequences, calcified lesions at other non-LM sites have regularly been treated successfully with BRS after adequate lesion preparation by high-pressure non-compliant balloon dilatation, cutting balloon or rotational atherectomy.

Fourthly, Absorb BVS have to be expanded gradually by increasing the inflation pressure by two atmospheres (atm) every five seconds, terminating with a long inflation of approximately 30 seconds. Thus, for a 3.5 mm Absorb BVS to be expanded to its expansion limit of 4.0 mm at 16 atm, in total approximately 65 s of occlusion of the LM are needed. In the setting of LV dysfunction and/or if the right coronary artery is diseased, these prolonged inflation times could result in ischaemic complications of severe hypotension or life-threatening arrhythmias.

Another technical concern, although minor, may be the challenging ostial positioning with no radiopaque scaffolds.

LM BIFURCATION LESIONS

Bifurcation lesions in appropriately selected patients are potentially good candidates for BRS treatment. However, as most (40% to 96% in different series) LM lesions involve the LM bifurcation and 70% to 80% of LM disease patients present with multivessel disease12, although theoretically sound, the LM bifurcation is less attractive for BRS treatment. Unprotected distal LM bifurcation PCI is always a challenging procedure and has worse long-term clinical outcomes than the more favourable results obtained with ostial- or shaft-LM lesions, especially when a two-stent approach is used13. In the SYNTAX study, 63% of LM lesions involved the left main bifurcation and required treatment. Of these, >90% had plaque extension into the left anterior descending (LAD) coronary artery and about 3/4 into the circumflex (LCx) territory.

At this moment the data on BRS bifurcation techniques are still limited compared to metallic DES bifurcation techniques, although even for metal DES there are no good systematic data supporting one bifurcation technique over another for LM bifurcation stenting14. For use of BRS in a coronary bifurcation, the European Bifurcation Club (EBC) proposed the strategy of provisional side branch crossover scaffolding with a proximal optimisation technique (POT) for proper apposition, side branch dilatation in case of TIMI flow <2 or symptoms of angina at low pressures (max. 8 atm), and a second stent (DES or scaffold) as bail-out with final POT (Table 2)15. However, bench testing using bifurcation phantoms has been performed to assess the safety and efficacy of side branch dilatation through the BRS and main branch post-dilatation techniques. Strut fractures were not observed at low inflation pressures, but high inflation pressures or larger side branch balloons frequently caused BRS strut fractures and/or lumen compromise16. When a two BRS strategy has to be used, a provisional T-stenting or TAP technique is recommended in the majority of cases. However, at least in a bifurcation phantom model, other contemporary bifurcation techniques, such as the culotte, double-crush or mini-crush, performed well using the Absorb BVS17. These techniques are generally not advocated for use with BRS in small vessel bifurcations, as they would result in ≥2 or 3 layers of scaffold struts leading to a serious luminal reduction and a high chance of delayed healing. However, in particular in the setting of an LM bifurcation with a main branch lumen of 3.5-4.0 mm, such a luminal reduction over a short coronary segment may be tolerated (Figure 1, Moving image 1-Moving image 16). As the expansion capabilities of the Absorb BVS are limited, simultaneous kissing balloon post-dilatation is usually not recommended while sequential POT, side branch post-dilatation and final POT were proposed at the last EBC meeting. If simultaneous kissing dilatation is indicated, only low pressures (e.g., 5 atm) and preferably with the side branch balloon protruding just outside the side branch, “snuggling” against the main branch balloon, are advised18. For POT, non-compliant balloons with a diameter no greater than 0.5 mm above nominal scaffold diameter should be used.

Figure 1. LM bifucation lesion treated with mini-crush technique and final kissing balloon post-dilatation. Distal LM lesion before (A) and after treatment with two Absorb BVS (LCx and LM-LAD), mini-crush and final kissing balloon post-dilatation (B). OCT of LM-LAD showing good stent apposition (C). Red dotted lines: OCT images of LM bifurcation (D) and LM carina (E). Three layers of Absorb BVS struts are visible at the carina.

Also, intracoronary imaging with OCT should always be considered to determine optimal stent size and landing zones and is highly recommended to ensure adequate stent expansion and apposition in all LM bifurcation segments after the procedure.

PATIENT CONSIDERATIONS IN LM DISEASE

Patients presenting with LM disease tend to be older than the average PCI patient population2, leading to a higher bleeding risk and less preference for long-term DAPT. However, as a shorter DAPT duration for second-generation DES was recently shown to be safe19, DAPT for ≥12 months is still advocated for BRS because of the strut thickness and concerns about stent thrombosis20, making this technology less appropriate for patients with an increased risk of bleeding complications.

Case reports on BRS use in LM disease

Several authors have already reported on the successful use of Absorb BVS in LM disease. However, information on long-term follow-up is generally lacking. Fernandez et al described the case of a 56-year-old male patient with a history of CABG, who presented with an NSTEMI due to a severe lesion in the distal LM. The LAD was protected by a LIMA graft and was ostially occluded. The circumflex territory was, however, not protected. After predilatation, a 3.5×18 mm Absorb BVS was expanded to 3.94 mm at 14 atm with good final angiographic result and no events at 30-day follow-up21. Two successful PCI cases, using a bifurcation crossover technique with stenting from LM towards the LAD and subsequent fenestration of the scaffold struts towards the circumflex artery, were published22,23. Cortese et al described nine patients with unprotected LM disease involving the bifurcation and a low SYNTAX score. A single-scaffold strategy was performed in all patients. Invasive imaging with IVUS showed four cases of scaffold underexpansion and one case of acute recoil. One patient developed recurrent angina 12 months after the index PCI due to late scaffold recoil and underwent CABG24. Furthermore, V-stenting with two Absorb BVS scaffolds for an LM bifurcation lesion has been described25, but after five months the patient presented with in-scaffold restenosis treated by PCI, as reported by Miyazaki et al26. Here, we report a new successful case of the V-stenting approach for treating a Medina 0,1,1 LM bifurcation stenosis (Figure 2). Finally, in the GHOST-EU trial, 17 of the 1,189 Absorb BVS-treated patients (1.2%) received a scaffold for LM disease. Although one of these patients presented with a scaffold thrombosis at day 84, unfortunately no further description of the implantation technique or of the individual patient outcomes was reported for the remaining 16 patients20.

Figure 2. LM bifurcation stenosis treated with V-stenting. LM bifurcation stenosis (Medina 0,1,1) and diffuse calcification of the proximal and mid LAD (A). Rotational atherectomy of the LAD (B). V-stenting with a 3.5×18 mm Absorb BVS in both LAD and LCx (C). Final result after implantation of four Absorb BVS (three in the LAD, one in the LCx) (D).

Conclusions

Taking everything into consideration, the use of BRS for the treatment of LM disease is presently generally not recommended and is considered to be investigational with current BRS platforms. However, in highly selected cases, the current BRS devices could still be applied with favourable results, especially in young patients with non-calcific LM lesions of ≤4.0 mm diameter with ostial or body disease or distal LM disease involving the LAD alone where a single crossover scaffold strategy can be applied. Before the use of BRS for LM disease can be advocated, more data on long-term safety and efficacy and larger scaffold designs are urgently needed.

Conflict of interest statement

C. Tamburino and R-J. van Geuns have received speaker’s honoraria from Abbott Vascular. R-J. van Geuns has received research grants (institutional) from Abbott Vascular. The other authors have no conflicts of interest to declare.

Online data supplement

Moving image 1-Moving image 16. LM bifucation lesion treated with mini-crush technique and final kissing balloon post-dilatation.

Supplementary data

To read the full content of this article, please download the PDF.

Moving image 1

Moving image 2

Moving image 3

Moving image 4

Moving image 5

Moving image 6

Moving image 7

Moving image 8

Moving image 9

Moving image 10

Moving image 11

Moving image 12

Moving image 13

Moving image 14

Moving image 15

Moving image 16


References

Volume 11 Supplement V
May 19, 2015
Volume 11 Supplement V
View full issue


Key metrics

Suggested by Cory

10.4244/EIJV11SVA44 May 19, 2015
When and how to use BRS in bifurcations?
Stankovic G and Lassen JF
free

10.4244/EIJV11SVA40 May 19, 2015
Bioresorbable scaffolds on the bench
Ormiston J et al
free

Debate

10.4244/EIJ-E-23-00015 Jun 19, 2023
Bioresorbable coronary scaffolds are ready for a comeback: pros and cons
Stone GW et al
free

10.4244/EIJV11SVA45 May 19, 2015
The future of BRS in bifurcations
Kraak R et al
free

10.4244/EIJV11SVA29 May 19, 2015
Dedicated stents for distal left main stenting
Grundeken M.J et al
free

EXPERT REVIEW

10.4244/EIJ-D-17-00499 Aug 25, 2017
State of the art: the inception, advent and future of fully bioresorbable scaffolds
Katagiri Y et al
free

10.4244/EIJV11SVA41 May 19, 2015
Lessons from the GHOST-EU registry
Latib A et al
free

10.4244/EIJV9I7A128 Nov 29, 2013
Kissing vanishing stents: are we trading ephemeral benefit for permanent damage?
Di Mario C et al
free

10.4244/EIJV11SVA38 May 19, 2015
Technology limitations of BRS in bifurcations
Fox J et al
free
Trending articles
153.65

Clinical research

10.4244/EIJ-D-20-01125 Oct 20, 2021
An upfront combined strategy for endovascular haemostasis in transfemoral transcatheter aortic valve implantation
Costa G et al
free
54.9

Expert review

10.4244/EIJ-D-21-01010 Jun 24, 2022
Device-related thrombus following left atrial appendage occlusion
Simard T et al
free
43.25

Clinical Research

10.4244/EIJ-D-21-01091 Aug 5, 2022
Lifetime management of patients with symptomatic severe aortic stenosis: a computed tomography simulation study
Medranda G et al
free
X

The Official Journal of EuroPCR and the European Association of Percutaneous Cardiovascular Interventions (EAPCI)

EuroPCR EAPCI
PCR ESC
Impact factor: 7.6
2023 Journal Citation Reports®
Science Edition (Clarivate Analytics, 2024)
Online ISSN 1969-6213 - Print ISSN 1774-024X
© 2005-2024 Europa Group - All rights reserved