DOI: 10.4244/EIJV11SVA44

When and how to use BRS in bifurcations?

Goran Stankovic1*, MD, PhD; Jens Flensted Lassen2, MD, PhD

Abstract

Bioresorbable coronary scaffolds (BRS) may offer potential advantages compared to metallic DES, aiming to restore vessel patency without implanting a permanent prosthesis, which may be especially important for bifurcation treatment. On the other hand, there are some inherent limitations, which may impact on the widespread use of BRS. In the current article we discuss the bench testing data and initial clinical results on BRS use in bifurcation lesions presented during European Bifurcation Club (EBC) meetings and review some of the limited number of published real-world registry results.

Introduction

Several bioresorbable coronary scaffold (BRS) designs are currently available, constructed either from polymers or from magnesium alloy1-3. BRS may offer potential advantages compared with metallic DES, aiming to restore vessel patency without implanting a permanent prosthesis, which may be especially important for bifurcation treatment. On the other hand, there are some inherent limitations, such as bulky profile and limited expansion capacity, length of time for complete resorption of the Absorb (Abbott Vascular, Santa Clara, CA, USA) BRS (approximately four years), creation of “neointimal bridge” on jailed struts4, as well as the signals for increased early scaffold thrombosis demonstrated recently by two large all-comers registries, the GHOST-EU and AMC5,6, which may impact on the widespread use of BRS. The BRS which has been evaluated most is the bioresorbable everolimus-eluting scaffold Absorb (Abbott Vascular), with the publication of several studies with long follow-up7-9. However, the Absorb BRS has been clinically evaluated in simple lesions and the available randomised trials have excluded bifurcations with side branches (SB) exceeding or equal to 2 mm in diameter. Therefore, bifurcation management with BRS remains challenging. In the current article we discuss bench testing data and initial clinical results on BRS use in bifurcation lesions presented during the 9th and 10th European Bifurcation Club (EBC) meetings, and review some of the limited number of published real-world registry results. The consensus reached represents a revision of the EBC recommendations for metallic DES10,11.

Provisional strategy utilising BRS

There is general agreement that a provisional strategy should be the default technique when BRS are used for bifurcation treatment11.

BRS SELECTION

Stent selection is still intensely debated: it seems to be a real dilemma as to which strategy should be recommended in different anatomies. The main problem is the restriction of dilatation of the BRS above 0.5 mm of the reference diameter. One approach is to select the stent according to the distal reference of the main vessel (MV), similar to the recommendation for metallic stents. This strategy, however, may have restrictions when post-dilating the proximal MV due to the mechanical limitations of expanding the stent. In suitable anatomies, another strategy could be to select the stent according to the proximal MV diameter and to deploy at low pressure in order to avoid damaging the distal vessel, followed by adequate post-dilatation of the proximal segment11,12. However, the decision on sizing the stent according to the proximal or distal reference diameter depends on the individual coronary anatomy of the patient and is, in the end, the decision of the operator. Importantly, an Absorb BRS should not be used in bifurcations in which the proximal MV diameter is greater than the maximal recommended diameter of the BRS12,13.

BRS DEPLOYMENT

Similar to non-bifurcation lesions, balloon pressure should be increased progressively (2 atm every five seconds) up to nominal pressure: inflation duration of 30 seconds or more is recommended.

OPTIMISATION STRATEGIES

Based on Murray’s law, the proximal optimisation technique (POT) is recommended with BRS but restricted to non-compliant (NC) balloons 0.5 mm larger than the reference (the maximal recommended diameter for the Absorb BRS because stretching with a larger diameter balloon may cause BRS strut fracture in the proximal vessel)13. While single strut fracture may have little clinical impact, multiple scaffold fractures and the protrusion of many struts into the lumen are likely to have adverse consequences14. To perform POT, an NC balloon is positioned with the distal balloon marker just proximal to the carina, and then inflated to expand to the diameter of the proximal segment11. This inflation decreases the angle of the ostium relative to the proximal MV segment, thus facilitating entry of a guidewire and balloon through the struts into the side branch. Bench testing has demonstrated that POT is safe and a mandatory step in provisional strategy15.

The next step is dilatation through the side of the scaffolds, which results in distortion inside the MV but also some protrusion of struts into the SB12,16. To avoid strut fracture, a side branch dilatation with a 3.0 mm NC balloon at a maximum of 10 atm could be used for a 3.0 mm Absorb BRS12. Final POT with an adequately sized NC balloon should be performed in the proximal MV to correct any scaffold malapposition created by side opening (the sequential strategy is “PSP”: POT+Side branch opening+final POT)15.

Routine final kissing balloon dilatation (FKBD) is not recommended. Recently, FKBD with minimal overlap of two balloons was proposed, positioning the proximal marker of the SB balloon in the MV immediately proximal to the SB ostium (“snuggle balloon dilatation” or “mini-FKBD”, depending on the length of SB balloon protrusion) and low-pressure inflation (safe threshold at or below 5 atm for each NC balloon sized 1:1 to the distal vessel reference)12,17,18. Higher pressures increase the risk of strut fracture12. Final POT with an adequately sized NC balloon is recommended (but not essential) to correct any scaffold deformation in the proximal MV (the kissing strategy is “PKP”: POT+Snuggle Kissing+final POT)15. In bench testing, the PSP and PKP strategies provided similar optical coherence tomography (OCT) and microcomputed tomography (micro-CT) results in the MV15. Intravascular imaging, especially OCT, is also recommended in the clinical setting to guide optimal Abosrb BRS implantation in bifurcation lesions11.

Two-stent techniques

Bench studies and several case reports have demonstrated the feasibility of a number of planned two-stent techniques utilising either two BRS or a combination of BRS and DES6,19-25.

When SB stenting is required as a crossover from provisional, metallic DES or BRS implantation on the SB should be considered utilising for preference the T-stenting technique12,15. Preliminary experience with the TAP (T-stenting and small protrusion) technique has also demonstrated sufficient SB lumen area without compromising the MV18,24. A BRS may be considered for SB ≥2.5 mm and if the angle with the MV is favourable for the T technique. FKBD should be performed with sequential inflation (SB first followed by MV), and with sequential deflation (MV first followed by SB). FKBD should be performed at a low pressure to avoid scaffold disruption at the orifice of the SB, proximal MV scaffold overdilatation and excessive shift of the scaffold neo-carina and interference with the MV after MV balloon deflation25.

According to bench testing data with OCT and micro-CT evaluation, complex techniques like culotte and mini-crush stenting result in MV and SB lumen obstruction, and these techniques should be avoided in order to prevent scaffold strut fractures and excessive overlap related to the thickness of the struts11,12,15. Operators need to keep in mind that, in the case of culotte stenting, 300 μm of scaffold (Absorb BRS strut thickness ≈150 μm) remains in the overlapped region of the proximal MV and, in the case of mini-crush, the triple layer creates approximately 450 μm of scaffold on one side, though for a short distance24.

When a metallic DES is implanted in the SB as a first step in a planned two-stent strategy, the mini-crush and the sleeve techniques have been reported17,23. In conclusion, it is generally agreed that implantation of BRS in bifurcations with SB diameter larger than 2 mm should be evaluated in randomised trials to facilitate the rapid creation of firm data on efficacy and safety for this promising new technology.

Summary

Bench testing data and initial clinical results of BRS use in bifurcation lesions seem promising. Simple strategies, with MB stenting followed by POT, and, when needed, SB side opening and final POT or POT with mini-FKBD are recommended. Operators need to understand the difference in the mechanical characteristics of BRS and metallic DES and carefully select device and balloon size and pressures to be utilised. The present recommendations apply to the Absorb BRS and may need to be revised with other stent designs, including the magnesium alloy-based BRS which may have a more predictable behaviour, known from other metallic stents.

Conflict of interest statement

The authors have no conflicts of interest to declare.


References

Volume 11 Supplement V
May 19, 2015
Volume 11 Supplement V
View full issue


Key metrics

Suggested by Cory

10.4244/EIJV11SVA45 May 19, 2015
The future of BRS in bifurcations
Kraak R et al
free

10.4244/EIJV11SVA41 May 19, 2015
Lessons from the GHOST-EU registry
Latib A et al
free

10.4244/EIJV11SVA40 May 19, 2015
Bioresorbable scaffolds on the bench
Ormiston J et al
free

10.4244/EIJV11SVA38 May 19, 2015
Technology limitations of BRS in bifurcations
Fox J et al
free

10.4244/EIJV11SVA2 May 19, 2015
The EuroIntervention coronary bifurcation treatment supplement
Stankovic G et al
free

10.4244/EIJV9I7A128 Nov 29, 2013
Kissing vanishing stents: are we trading ephemeral benefit for permanent damage?
Di Mario C et al
free

10.4244/EIJV11SVA31 May 19, 2015
Is there a need for dedicated devices?
Wykrzykowska JJ et al
free

10.4244/EIJV11SVA42 May 19, 2015
Treatment of bioresorbable scaffold failure
Felix C et al
free
Trending articles
151.43

State-of-the-Art

10.4244/EIJ-D-22-00776 Apr 3, 2023
Computed tomographic angiography in coronary artery disease
Serruys PW et al
free
55.9

Clinical research

10.4244/EIJ-D-22-00621 Feb 20, 2023
Long-term changes in coronary physiology after aortic valve replacement
Sabbah M et al
free
54.9

Expert review

10.4244/EIJ-D-21-01010 Jun 24, 2022
Device-related thrombus following left atrial appendage occlusion
Simard T et al
free
43.75

Clinical Research

10.4244/EIJ-D-21-01091 Aug 5, 2022
Lifetime management of patients with symptomatic severe aortic stenosis: a computed tomography simulation study
Medranda G et al
free
39.95

Clinical research

10.4244/EIJ-D-22-00558 Feb 6, 2023
Permanent pacemaker implantation and left bundle branch block with self-expanding valves – a SCOPE 2 subanalysis
Pellegrini C et al
free
X

The Official Journal of EuroPCR and the European Association of Percutaneous Cardiovascular Interventions (EAPCI)

EuroPCR EAPCI
PCR ESC
Impact factor: 7.6
2023 Journal Citation Reports®
Science Edition (Clarivate Analytics, 2024)
Online ISSN 1969-6213 - Print ISSN 1774-024X
© 2005-2024 Europa Group - All rights reserved