Original Research

DOI: 10.4244/EIJ-D-23-00734

Early outcomes of transapical mitral valve implantation versus surgical replacement in matched elderly patients at intermediate surgical risk

Johannes A. Ziegelmueller1, MD; Melchior Burri1, MD; Andreas Stein2, MD; Peter Tassani-Prell2, MD, PhD; Markus Krane1,3, MD, PhD; Rüdiger Lange1, MD, PhD; Hendrik Ruge1, MD

Abstract

BACKGROUND: Data comparing transcatheter mitral valve implantation (TMVI) with surgical mitral valve replacement (SMVR) are lacking.

AIMS: This study sought to compare the 30-day Valve Academic Research Consortium (VARC)-3 device success of TMVI with that of SMVR.

METHODS: Matching protocol combined exact matching (sex, atrial fibrillation, previous surgical aortic valve replacement [SAVR] or coronary artery bypass grafting [CABG]), coarsened exact matching (age) and propensity score matching (body mass index, mitral valve pathology and concomitant tricuspid regurgitation).

RESULTS: A total of 40 Tendyne TMVI and 80 SMVR patients with similar baseline characteristics were analysed (TMVI vs SMVR): age (78 years [interquartile range [{IQR} 75; 80] vs 78 years [IQR 73; 80]; p=0.8), female (60% vs 60%; p=1.0), atrial fibrillation (67.5% vs 63.7%; p=0.8), previous SAVR (12.5% vs 10.0%; p=0.8), previous CABG (20.0% vs 16.2%; p=0.8), body mass index (25.54 kg/m² vs 25.24 kg/m²; p=0.7) and valve pathology (mitral regurgitation: 70.0% vs 73.8%, mitral stenosis: 7.5% vs 3.8%, and mixed disease: 22.5% vs 22.5%; p=0.6). Most baseline characteristics not included in the matching model were balanced among the TMVI/SMVR cohorts: European System for Cardiac Operative Risk Evaluation (EuroSCORE) II (5.8% [IQR 2.9; 7.5] vs 4.2% [IQR 2.4; 6.8]; p=0.3) and Society of Thoracic Surgeons Predicted Risk of Mortality (STS-PROM) score (5.2% [IQR 3.2; 8.6] vs 4.1% [IQR 3.3; 6.1]; p=0.076). Coronary artery disease (67.5% vs 32.5%; p<0.001) and previous percutaneous coronary intervention (47.5% vs 25.0%; p=0.023) differed among groups. Mitral VARC (MVARC) device success at 30 days was achieved in 82.5% of patients after TMVI and 57.5% of patients after SMVR (p=0.04). MVARC procedural success at 30 days was 75.0% after TMVI versus 52.5% after SMVR (p=0.07). Thirty-day mortality (2.5% vs 3.8%; p=0.47), technical success (97.5% vs 97.5%; p=1.0), major bleeding (17.5% vs 18.7%; p=0.087), stroke (5.0% vs 4.9%; p=1.0) and postoperative haemodialysis (7.5% vs 5.2%; p=0.4) were similar in both groups.

CONCLUSIONS: Patients with intermediate surgical risk, according to STS-PROM and EuroSCORE II, demonstrated higher rates of MVARC device at 30 days after TMVI compared to 30 days after SMVR. Rates of survival and procedural success, neurological, renal and bleeding complications were similar. Transfusion count and length of stay were lower after TMVI. For elderly patients at intermediate risk, a TMVI eligibility assessment may be considered.

Sign in to read
the full article

Forgot your password?
No account yet?
Sign up for free!

Create my pcr account

Join us for free and access thousands of articles from EuroIntervention, as well as presentations, videos, cases from PCRonline.com

Volume 20 Number 5
Mar 4, 2024
Volume 20 Number 5
View full issue


Key metrics

Suggested by Cory

Editorial

10.4244/EIJ-E-24-00002 Mar 4, 2024
Catheter-based versus surgical mitral valve intervention: choosing between the red pill and the blue pill
Bagur R and McInerney A
free

10.4244/EIJV13I9A153 Oct 11, 2017
Transcatheter mitral valve replacement: looking beyond the implant
Mylotte D
free

Clinical research

10.4244/EIJ-D-22-01037 Aug 21, 2023
Clinical outcomes of transcatheter mitral valve replacement: two-year results of the CHOICE-MI Registry
Ludwig S et al
free

10.4244/EIJV15I12A193 Dec 20, 2019
Predicting TMVR outcomes – the Tendyne experience
Piazza N et al
free

Research correspondence

10.4244/EIJ-D-21-00745 Apr 22, 2022
Early outcomes of transcatheter mitral valve replacement with the Tendyne system in severe mitral annular calcification
Gössl M et al
free
Trending articles
152.9

Clinical research

10.4244/EIJ-D-20-01125 Oct 20, 2021
An upfront combined strategy for endovascular haemostasis in transfemoral transcatheter aortic valve implantation
Costa G et al
free
47.8

NEW INNOVATION

10.4244/EIJ-D-15-00467 Feb 20, 2018
Design and principle of operation of the HeartMate PHP (percutaneous heart pump)
Van Mieghem NM et al
free
39.1

Clinical research

10.4244/EIJ-D-22-00558 Feb 6, 2023
Permanent pacemaker implantation and left bundle branch block with self-expanding valves – a SCOPE 2 subanalysis
Pellegrini C et al
free
X

The Official Journal of EuroPCR and the European Association of Percutaneous Cardiovascular Interventions (EAPCI)

EuroPCR EAPCI
PCR ESC
Impact factor: 7.6
2023 Journal Citation Reports®
Science Edition (Clarivate Analytics, 2024)
Online ISSN 1969-6213 - Print ISSN 1774-024X
© 2005-2024 Europa Group - All rights reserved