IMAGE IN CARDIOLOGY

DOI: 10.4244/EIJY15M07_09

Percutaneous Melody valve implantation for tricuspid valve regurgitation with very large annulus: the diabolo-shaped bridge for valve-in-stent technique

Sophie Malekzadeh-Milani1,2, MD; Mehul Patel1,2, MD; Younes Boudjemline1,2,3*, MD, PhD

Tricuspid valve insertion of the Melody® (Medtronic, Minneapolis, MN, USA) valve has been described in patients with a bioprosthesis. However, size limitations restrict its use. We report implantation of a Melody valve in a patient with severe regurgitation of a 33 mm Carpentier-Edwards PERIMOUNT valve (Edwards Lifesciences, Irvine, CA, USA). The procedure was conducted from the right femoral vein. Balloon calibration showed an inner diameter of 28 mm. The platform for the Melody valve was created using a short ev3 Mega™ LD (ev3 Endovascular Inc., Plymouth, MN, USA) fitted over a CP covered stent (8Z39; NuMED, Inc., Hopkinton, NY, USA), the two stents being crimped on a 28 mm NuCLEUS™ balloon (NuMED). The assembly was delivered in the bioprosthesis (Figure 1). This allowed full expansion and flaring of the extremities of the CP stent and only partial opening of the centre of the CP stent, creating a diabolo-shaped platform. A Melody valve mounted on a 22 mm delivery catheter was implanted and post-dilated with a 24 mm balloon (Moving image 1). Five months post implantation, there was no valvular or paravalvular regurgitation with a mean gradient of 3 mmHg.

Figure 1. Bench testing of the valve-in-stent technique. The properties of two different stents were used with one fitted on the second one to create a platform: 1) a bare metal stent with limited expansion to create a restrictive region; 2) a covered stent without limited expansion to allow anchoring to the large surgically inserted bioprosthesis. A) Lateral view of the two stents after inflation of the balloon. B) Sizing with PTS balloon. C) Inflation of the assembly on the balloon. D) The Melody valve is advanced within the bioprosthesis. E) Post-dilatation with a balloon. F) Post-implantation angiogram: no regurgitation.

Conflict of interest statement

Y. Boudjemline acts as a proctor for Medtronic Inc. The other authors have no conflicts of interest to declare.

Supplementary data

Moving image 1. Creation of a diabolo-shaped platform and percutaneous implantation of a Melody valve. A) Right ventricle angiogram: severe tricuspid regurgitation. B) Calibration with PTS balloon. C) & D) Pre-stenting: creation of a diabolo-shaped platform. E) Melody valve is advanced. F) Post-dilatation.

Volume 11 Number 13
Apr 20, 2016
Volume 11 Number 13
View full issue


Key metrics

Suggested by Cory

10.4244/EIJV14I9A174 Oct 12, 2018
A tune of triumph
Schmid M and Vejlstrup N
free

CLINICAL RESEARCH

10.4244/EIJV10I8A168 Dec 22, 2014
Transcatheter valve-in-valve implantation in the tricuspid position
Tzifa A et al
free

CLINICAL RESEARCH

10.4244/EIJ-D-18-00054 Oct 12, 2018
Up to 11 years of experience with the Melody valved stent in the right ventricular outflow tract
Cools B et al
free

Image – Interventional flashlight

10.4244/EIJ-D-20-00427 Feb 19, 2021
Intravascular ultrasound guidance for valve-in-valve transcatheter pulmonary valve replacement
Kalinczuk L et al
free

10.4244/EIJV10I8A156 Dec 22, 2014
Transcatheter mitral valve-in-ring with the Melody prosthesis: a new therapeutic opportunity
Borger MA and Kodali SK
free

CLINICAL RESEARCH

10.4244/EIJ-D-16-00853 Apr 20, 2017
Surgical mitral valve replacement with the Melody valve in infants and children: the Italian experience
Frigiola A et al
free

10.4244/EIJV9I10A207 Feb 20, 2014
Folded Melody valve technique for complex right ventricular outflow tract
Malekzadeh-Milani S et al
free
Trending articles
151.43

State-of-the-Art

10.4244/EIJ-D-22-00776 Apr 3, 2023
Computed tomographic angiography in coronary artery disease
Serruys PW et al
free
55.9

Clinical research

10.4244/EIJ-D-22-00621 Feb 20, 2023
Long-term changes in coronary physiology after aortic valve replacement
Sabbah M et al
free
54.9

Expert review

10.4244/EIJ-D-21-01010 Jun 24, 2022
Device-related thrombus following left atrial appendage occlusion
Simard T et al
free
43.75

Clinical Research

10.4244/EIJ-D-21-01091 Aug 5, 2022
Lifetime management of patients with symptomatic severe aortic stenosis: a computed tomography simulation study
Medranda G et al
free
39.95

Clinical research

10.4244/EIJ-D-22-00558 Feb 6, 2023
Permanent pacemaker implantation and left bundle branch block with self-expanding valves – a SCOPE 2 subanalysis
Pellegrini C et al
free
X

The Official Journal of EuroPCR and the European Association of Percutaneous Cardiovascular Interventions (EAPCI)

EuroPCR EAPCI
PCR ESC
Impact factor: 7.6
2023 Journal Citation Reports®
Science Edition (Clarivate Analytics, 2024)
Online ISSN 1969-6213 - Print ISSN 1774-024X
© 2005-2024 Europa Group - All rights reserved