Original Research

DOI: 10.4244/EIJ-D-23-00802

Long-term cardiac computed tomography follow-up after left atrial appendage occlusion

Lasse Hubertus Tiroke1,2, BSc; Anders Kramer1,2, MD; Mette Wørmer Poulsen1,2, BSc; Caroline Damsgaard Jensen1,2, BSc; Jesper Møller Jensen1, MD, PhD; Bjarne Linde Nørgaard1,2, MD, DMSc, PhD; Kasper Korsholm1,2, MD, PhD; Jens Erik Nielsen-Kudsk1,2, MD, DMSc

Abstract

BACKGROUND: Left atrial appendage occlusion (LAAO) is performed increasingly, but long-term follow-up imaging data are lacking.

AIMS: The aim of this study was to evaluate the safety and durability of the Amplatzer Amulet device >4 years after LAAO.

METHODS: This was a prospective observational cohort study including 52 patients implanted with the Amplatzer Amulet device at Aarhus University Hospital, Denmark. A >4-year follow-up cardiac computed tomography (CT) scan after LAAO was performed and compared with the results from the 2-month and 12-month scans. The primary outcome was left atrial appendage (LAA) sealing based on distal LAA contrast patency and peridevice leakage (PDL), stratified into complete occlusion (grade 0 [G0]) and grade 1-3 leakage (G1-3), respectively. Secondary outcomes were low- and high-grade hypoattenuated thickening (HAT), device-related thrombosis (DRT) and device durability.

RESULTS: The median (interquartile range [IQR]) follow-up time from LAAO to the latest CT scan was 5.8 years (4.5; 6.3). At 2-month (n=52), 12-month (n=27) and >4-year CT follow-ups (n=52), rates of both complete occlusion (33%, 37%, 35%) and G2 leaks (52%, 52%, 48%) remained stable. Rates of G1 leaks varied (14%, 4%, 6%) and G3 leaks rose (2%, 7%, 12%) from earliest to latest follow-up. The median left atrial (LA) volume increased from 127 mL (96; 176) to 144 mL (108; 182) and 147 mL (107; 193). No DRT was found. The structural device integrity was preserved.

CONCLUSIONS: This study indicates a stable LAA sealing status throughout the follow-up period, emphasising the importance of the procedural result in avoiding PDL. Few patients displayed PDL progression, which might partly be related to LA remodelling with increasing volume. The long-term device durability appears excellent. Larger studies are warranted to confirm these findings.

Sign in to read
the full article

Forgot your password?
No account yet?
Sign up for free!

Create my pcr account

Join us for free and access thousands of articles from EuroIntervention, as well as presentations, videos, cases from PCRonline.com

Volume 20 Number 11
Jun 3, 2024
Volume 20 Number 11
View full issue


Key metrics

Suggested by Cory

Clinical research

10.4244/EIJ-D-21-00286 Dec 17, 2021
Safety and feasibility of peri-device leakage closure after LAAO: an international, multicentre collaborative study
Piayda K et al
free

Original Research

10.4244/EIJ-D-23-00684 Mar 4, 2024
Single antiplatelet therapy following Amplatzer left atrial appendage occlusion
Kramer A et al

EXPERT REVIEW

10.4244/EIJ-D-17-00469 Nov 20, 2017
Residual leaks following percutaneous left atrial appendage occlusion: assessment and management implications
Raphael CE et al
free
Trending articles
152.9

Clinical research

10.4244/EIJ-D-20-01125 Oct 20, 2021
An upfront combined strategy for endovascular haemostasis in transfemoral transcatheter aortic valve implantation
Costa G et al
free
47.8

NEW INNOVATION

10.4244/EIJ-D-15-00467 Feb 20, 2018
Design and principle of operation of the HeartMate PHP (percutaneous heart pump)
Van Mieghem NM et al
free
39.1

Clinical research

10.4244/EIJ-D-22-00558 Feb 6, 2023
Permanent pacemaker implantation and left bundle branch block with self-expanding valves – a SCOPE 2 subanalysis
Pellegrini C et al
free
38.95

State-of-the-Art

10.4244/EIJ-D-23-00912 Oct 7, 2024
Optical coherence tomography to guide percutaneous coronary intervention
Almajid F et al
free
X

The Official Journal of EuroPCR and the European Association of Percutaneous Cardiovascular Interventions (EAPCI)

EuroPCR EAPCI
PCR ESC
Impact factor: 7.6
2023 Journal Citation Reports®
Science Edition (Clarivate Analytics, 2024)
Online ISSN 1969-6213 - Print ISSN 1774-024X
© 2005-2024 Europa Group - All rights reserved