Interventional post-myocardial infarction ventricular septal defect closure: a systematic review of current evidence

Florian Schlotter1, MD; Suzanne de Waha2, MD; Ingo Eitel2, MD; Steffen Desch2, MD; Georg Fuernau2, MD; Holger Thiele2*, MD

Abstract

Aims: Interventional closure of post-myocardial infarction ventricular septal defects (VSD) is an alternative treatment option to surgical repair. However, only limited evidence exists concerning the interventional closure of a VSD. This review seeks to establish an overview of the existing literature and to carry out a systematic analysis of the success rate and clinical outcome of this procedure.

Methods and results: We conducted a comprehensive systematic literature search to evaluate the existing evidence of percutaneous device closure of post-infarction VSD. Patient series with fewer than five reported cases were excluded. In total, 13 series were identified, with an overall inclusion of 273 patients. Mean patient age was 70 years. Cardiogenic shock was present in 48% of cases at the time of intervention. Device closure within the first 14 days (acute phase) after VSD detection was performed in 42% of cases. Technical success rate was high (>75%). Successful device implantation rate was 89%. Overall in-hospital/30-day mortality was 32%. Major complications included device embolisation, ventricular perforation and arrhythmias.

Conclusions: Percutaneous device closure of post-myocardial infarction VSD is a valuable alternative to surgical repair, with the advantage of immediate shunt reduction to prevent haemodynamic deterioration. A high rate of technically successful percutaneous procedures can be achieved; however, the mortality rate remains high, especially in cardiogenic shock patients.

Introduction

Post-infarction ventricular septal defect (VSD) is an infrequent but life-threatening complication following acute myocardial infarction (AMI). Data from thrombolytic management of AMI suggest an incidence of ventricular septal rupture of 0.2%, with a median time from myocardial infarction to VSD development of one day1. Only limited data exist on the frequency of post-infarct VSD in the interventional era. VSD complicating AMI frequently leads to cardiogenic shock (CS)2.

To date, three treatment options exist: 1) medical management, 2) surgical repair, and 3) interventional closure. Medical management carries a grim prognosis with a 94% mortality rate at 30 days1. Although open heart surgery is considered standard treatment for VSD, catheter-based intervention has emerged as a promising alternative3-5. The exact timing for closure remains a matter of debate. Current guidelines recommend urgent VSD closure, irrespective of haemodynamic status3-5, but surgeons often refrain from very early surgery due to extremely poor results. Instead, surgical repair is often only performed several weeks after myocardial infarction, which inevitably leads to selection of a healthier population eventually undergoing surgery (due to deaths while waiting for surgery).

In 1988, Lock and colleagues first reported on an interventional approach for post-infarct VSD closure6. Since this first report, transcatheter VSD closure has become a viable treatment option. Previously published case series of catheter-based VSD closure were usually restricted to an interventional approach during the subacute and chronic periods after VSD occurrence, with only one larger series demonstrating feasibility in the acute setting7. The general management of post-infarction VSD has recently been reviewed; however, it was not focused primarily on interventional management8.

Therefore, the aim of the present review was to provide a comprehensive overview of the current evidence including success rate and complications of interventional VSD closure in patients with post-infarction VSD.

Methods

DATA ACQUISITION

We conducted a comprehensive literature search of MEDLINE, EMBASE, CENTRAL and ClinicalTrials.gov, using the following subject headings: “ventricular septal defect”, “ventricular septal rupture”, “myocardial infarction”, “closure”, “septal occluder device”, “transcatheter”, “percutaneous”, in order to identify randomised trials, registries and case series.

No restrictions on publication date or language were applied. Duplicate papers were manually removed from the search results. Additionally, a manual search by examining reference lists from research papers and review articles was performed.

Three authors (FS, IE, SW) independently screened all titles and abstracts and excluded papers on the basis of predefined criteria. Disagreements were resolved by consensus. Case reports and case series including only a limited number of patients (<5) were excluded. Furthermore, case series reporting closure of non-infarction VSDs were omitted.

We included publications on secondary VSD closures after primary surgical closure attempts with residual shunt if those were outlined in the same report as primary transcatheter closure cases. No further restrictions were applied.

Articles were rejected on initial screening if titles or abstracts were clearly irrelevant. The full text of potentially relevant articles was reviewed to assess eligibility for inclusion in the analysis.

This study was conducted in compliance with the Preferred Reporting Items for Systematic Reviews and Meta-Analyses (PRISMA) statement9. A standardised form was used to extract study data, including a description of the study population, patient characteristics, procedure characteristics, complications, as well as the nature and number of clinical events.

STATISTICAL ANALYSIS

As the search found no randomised controlled trials, we sought to establish a statistical approach to a systematic review of the individual studies within the limitations of a non-randomised data environment. The following clinical endpoints were included in the analysis: in-hospital/30-day mortality, complications, baseline characteristics (age, gender, number of diseased coronary vessels, infarct location, VSD location, time from AMI to VSD, time from VSD to closure, presence of CS, prior intra-aortic balloon pump [IABP] use, secondary percutaneous closure attempt), and procedural characteristics (successful device implantation, type of occluder). For determination of the mortality rate (in-hospital/30-day mortality if specified), we calculated the absolute number of events and their sum as a percentage of the total population investigated. Procedure-related complications are outlined in detail under the precondition that they were reported and considered procedure-related by the individual study authors. Continuous data are presented as means and standard deviations. A two-sided p-value <0.05 was considered statistically significant.

Results

The literature search identified 13 publications reporting on interventional VSD closure, representing the composite experience of 38 centres (Table 1). All series were published between 2004 and 2015. Underlying data were collected from 1988-2012. The total number of patients was 273. Most studies were retrospective10-19, with three prospective analyses7,20,21. No randomised trials were identified.

A secondary percutaneous approach after primary surgical VSD closure was performed in 16.5% of patients (45/273). In this patient population, 33 patients were treated for residual or recurrent VSD10,12,13,17, two patients for a surgical patch dehiscence11, and 10 patients without further notion of the reason for the secondary percutaneous approach20. The number of patients, year of publication, number of acute phase and chronic phase patients after VSD detection, 30-day or in-hospital mortality, number of CS patients, number of successful device implantations, and number of primary percutaneous VSD procedures are outlined in Table 2.

PATIENT CHARACTERISTICS

The mean patient age was 70.3 years. In total, 69.4% of patients were male (145/209). The rate of individuals in CS was 47.9% (114/238) at the time of intervention, and 58.4% (122/209) had an IABP placed prior to VSD occlusion. Anterior/apical myocardial infarction causing the VSD was present in 56.9% of patients (145/255). Coronary single-vessel, double-vessel and triple-vessel disease was present in 61.9% (128/207), 24.6% (51/207), and 13.5% (28/207) of patients, respectively.

DEVICES

Septal defect occluders used share a common morphology. The majority of studies used one of three Amplatzer™ (St. Jude Medical, St. Paul, MN, USA) device subtypes – the AMPLATZER™ Septal Occluder, the AMPLATZER™ Muscular VSD Occluder and the AMPLATZER™ P.I. Muscular VSD Occluder, the latter specifically designed for post-infarction VSD closure (Figure 1). One study summarising the early experiences with post-infarction VSD occlusion reported on the use of the Clamshell device (C.R. Bard, Murray Hill, NJ, USA), the CardioSEAL® (NMT Medical, Boston, MA, USA) device, and the STARFlex® device (NMT Medical)10. Only two more reports included devices of a non-Amplatzer occluder (modified double disc occluder [Shanghai Shape Memory Alloy Material Co., Ltd, Shanghai, China])21. The device size was determined using echocardiography in the majority of cases, with fluoroscopic balloon-based sizing performed in a minority of cases12,13. Figure 2 demonstrates an Amplatzer device implantation for VSD.

Figure 1. Amplatzer occluders. A) AMPLATZER Septal Occluder. B) AMPLATZER Muscular VSD Occluder. C) AMPLATZER P.I. Muscular VSD Occluder.

Figure 2. Device implantation. A) Compressed double umbrella device advanced over the guidewire via the VSD into the left ventricle after release of the left ventricular double umbrella. B) Left ventricular disc fully developed, right-sided disc partially opened. C) Left and right ventricular disc fully opened and released from the delivery system.

PATIENT SELECTION

In the retrospective report by Assenza et al, representing the longest time interval under investigation (1988-2008), 30 patients with VSD were included (12 primary, 18 secondary device closures). The indication for device closure was based on the presence of heart failure and pulmonary overcirculation that did not respond to medical therapy. Eligibility was not based on prospective criteria and a surgical option was denied by consensus in all patients. Multiorgan failure had a very high mortality rate irrespective of the closure attempt10.

The first five consecutive cases of interventional VSD closure of the Auckland City Hospital were described in another report. The authors included two cases of prior surgical closure with residual shunt and one acute phase patient treated on day two after AMI. In this patient, device delivery was unsuccessful due to an inability to cross the defect and the patient died shortly afterwards11.

In the largest series to date, the UK experience is summarised retrospectively13. Electronic case records were circulated to centres that reported VSD device closure cases in a national database. No further selection criteria were applied.

Bialkowski et al reported on 19 patients receiving percutaneous VSD device closure, two for closure after surgery and 17 for primary VSD. The authors stratified the outcome according to closure attempt in the acute, subacute and chronic phases of VSD, while no restrictions on eligibility were outlined12.

Immediate device therapy was the preferred primary therapy in all patients with VSD in the prospective study by Thiele et al. The authors reported a high rate of CS patients (55%). Only morphological echocardiographically evaluated exclusion criteria were applied. These included a VSD size of >35 mm, an apical VSD without a suitable rim and a basal VSD too close to the valvular apparatus. However, these criteria did not lead to any exclusion of patients7.

Martinez et al reported the retrospective Mayo Clinic experience of percutaneous VSD occlusion. No further restrictions were applied17.

Only VSD patients with a significant left-to-right shunt (Qp:Qs ≥1.5) and a defect size as well as location suitable for device therapy were considered in another series comprising nine patients14.

Holzer et al evaluated patients for interventional closure on a case-by-case basis and applied no general clinical or haemodynamic exclusion criteria. A defect size exceeding 24 mm was considered unsuitable. In cases of prior unsuccessful surgical attempts, technically challenging VSD location or in patients with high surgical risk, device closure was preferred to surgery20.

The experience of the Montreal Heart Institute is summarised in another study by Maltais et al. Percutaneous treatment was considered on an individual basis in all patients with a VSD diameter <15 mm and in unstable patients with high surgical risk. The authors proposed a new treatment algorithm, rendering patients with a small or medium VSD suitable for percutaneous occlusion, facilitating delayed surgical correction in case of a significant residual VSD15. Initial single-centre results of eight patients were reported in another study, with percutaneous closure attempts performed within one week of myocardial infarction and haemodynamic deterioration in all but two patients16.

Zhu et al prospectively evaluated the feasibility of concomitant VSD device closure and coronary revascularisation in a 35-patient multicentre registry. A defect size in excess of the maximum available occluder device was considered inappropriate for percutaneous closure21. In a relatively large analysis, 42 patients ineligible for cardiac surgery entered a retrospective registry. Patients with congenital muscular VSD were excluded19.

Finally, in the analysis of Trivedi et al data, all patients undergoing treatment for VSD following acute myocardial infarction in three French hospitals were analysed. No further inclusion or exclusion criteria were defined18.

TIMING

Great heterogeneity concerning the timing of the interventional approach was observed. In summary, 46.4% of patients (110/237) were treated percutaneously within the first two weeks after VSD detection. A linear relationship exists between the number of acute phase cases (occlusion <14 days after AMI) included in the individual studies and a higher mortality (r=0.72, p=0.01) (Figure 3). Only one study specifically targeted the acute phase of VSD occurrence7.

Figure 3. Acute phase device closure and mortality. Correlation analysis between number of acute phase patients included (<14 days after AMI) and 30-day/in-hospital mortality.

COMPLICATIONS

A number of complications occurred in this severely compromised patient population. There was a total of 36 procedure-related complications. Among those reported were arrhythmias of differing degrees (three transitory 3rd degree atrioventricular blocks)10,12 and one 3rd degree atrioventricular block after crossing the VSD with the delivery sheath with consecutive death before implementation of pacing despite cardiopulmonary resuscitation7, one electromechanical dissociation with consecutive death10, eight reports of ventricular fibrillations/tachycardias11,12,19,21, and one report of atrial flutter terminated by cardioversion11. Furthermore, there was a report of two bradycardias associated with the implantation of the VSD occluder, which were not further specified20.

Appropriate sizing of the VSD occluder remains a matter of debate due to the broad variety of anatomy and tissues found in VSD. As a result, in total seven device embolisations have been reported7,10,13,20. One device embolised into the pulmonary artery and led to the death of the patient15.

Thiele et al described three cases of left ventricular rupture after closure7. There was one case of an iatrogenic right ventricular perforation15. One case of left ventricular dysfunction after the procedure was reported20. VSD occluder-related haemolysis was reported in one case12. Procedure-related blood loss or haematoma necessitating blood transfusion occurred in nine cases11,13,19,20. Finally, ventricular rupture related to the manipulation of the VSD positioning was reported in two patients enrolled in the same cohort19.

SUCCESSFUL DEVICE IMPLANTATION/MORTALITY

The overall successful device implantation rate was 89.0% (range: 80-100%)7,11-14,17,20,21. The total number of deaths amounted to 86, representing an in-hospital or 30-day mortality rate of 31.5%. The studies by Thiele et al and Marinakis et al reported the highest mortality rates (66% and 75%, respectively)7,16. A logistic EuroSCORE of 70% was reported in the study by Marinakis et al, representing a severely compromised status prior to implantation16. Both studies had CS rates >50%, and procedural success was 86% in the report by Thiele et al7. The majority of studies did not provide a detailed description of the causes of death.

PREDICTORS OF ADVERSE OUTCOMES

Ahmed et al reported a case series of five patients and indicated that a better outcome may be achieved if the procedure is delayed relative to the incident event11. Assenza et al used univariate analyses to predict the risk of mortality and identified the presence of CS, the use of the 43 mm STARFlex device, and increasing pulmonary/systemic flow ratio (at time of VSD closure) as being related to an increase in the mortality risk. Furthermore, the baseline modified Model for End-Stage Liver Disease (MELD) score, excluding the international normalised ratio (MELD)-XI score at the time of VSD closure, proved to be a strong predictor of increased death. A MELD-XI score of 20 had 100% sensitivity and 87% specificity for predicting 30-day mortality10. By means of multivariate logistic regression analysis, Maltais et al were able to identify residual VSD, time from AMI to VSD diagnosis and time delay between VSD diagnosis and either surgical or percutaneous VSD closure as statistically significant predictors of 30-day and in-hospital mortality15. Calvert et al demonstrated increased hazard ratios of death in association with female sex, the presence of CS, the use of inotropes, higher creatinine levels, larger defect size and absence of revascularisation therapy for presenting AMI, and a shunt reduction of less than two thirds13. Advanced age was identified as a risk factor for VSD occurrence and reduced survival after VSD occlusion13.

Discussion

The results of this systematic review indicate that interventional closure of VSD is a feasible treatment option in carefully selected high-risk patients with VSD. However, these pooled data clearly indicate that interventional VSD closure after AMI remains a high-risk procedure.

The overall mortality rate of percutaneous VSD closure was below or similar to that of several surgical VSD repair series reporting mortality rates ranging between 36 and 81%2,22-24. However, a direct comparison is not possible owing to the non-randomised design and an inherent selection bias with all case series of either surgical or interventional closure. Nevertheless, percutaneous VSD closure may permit avoidance or delay of surgical repair, while providing potentially life-saving therapy to high-risk patients with VSD. In addition, residual shunts of either a primary surgical or primary percutaneous closure may be treated, avoiding re-operation7,25,26.

From the first feasibility attempts at percutaneous device VSD closure in the 1980s6, a quarter of a century has passed and the procedure has received increasing attention in spite of decreasing VSD occurrence27.

FEASIBILITY

Recent reports have demonstrated that interventional VSD closure can be performed successfully7,10-17,20,21. However, currently there is no general consensus on establishing criteria for appropriate suitability. Based on the reports incorporated in this systematic review, the majority of VSD can be treated interventionally. In addition, various different subgroups within the VSD patient population have been investigated, including secondary VSD closure for residual shunt after surgical VSD closure10-13,17,20, concomitant percutaneous coronary intervention21, and immediate primary percutaneous closure7.

The multicentre, prospective GUSTO-I trial and the SHOCK registry reported mortality rates of 94% and 96%, respectively, in medically managed VSD patients, representing numbers estimated to be close to the natural history of VSD outcomes1,2. The advent of surgical closure increased survival rates dramatically, while a significant survival difference exists between surgical closure in the first days after AMI (<7 days after AMI, 54.1% operative mortality [in-hospital or within 30 days of the index surgical closure]) versus the time thereafter (>7 days after AMI, 18.4% operative mortality)22. Survival rates of interventional VSD device closure were shown to range from 25 to 86%, primarily depending on the presence of CS and the amount of time from AMI because of the inherent selection bias7,10-17,20,21. This observation is in line with the surgical closure approach.

In the GUSTO-I trial, all VSD patients with a Killip class of III or IV died1. Thus, prevention and timely reversal of CS are of high prognostic relevance. Even in CS patients, immediate primary closure attempts seem to be superior to the devastating results of medically treated patients7. The feasibility of percutaneous VSD occlusion is reflected by a class IIb recommendation in current European guidelines on revascularisation5.

PATIENT SELECTION

The primary objective of VSD therapy either surgically or percutaneously is left-to-right shunt reduction. This can be difficult because of the often meandering and dynamic VSD shape. Defect size can vary greatly and more than one VSD is often present. Failure to reduce the shunt results in worse outcomes13,15.

While survival to discharge is associated with favourable long-term outcomes7,13, procedural success does not necessarily translate into increased survival rates. This was demonstrated in the study by Thiele et al where the 30-day survival rate was only 35% despite a high percentage of technically successful device implantations (86%)7.

There is reason to believe that there may be a selection bias in percutaneous closure in relation to surgical closure. Percutaneous closure may have been carried out earlier after VSD detection but anatomically more complex lesions may have been referred to surgical correction.

The timing of VSD closure should be based on the haemodynamic status of the patient and the severity of the condition. Early closure of post-AMI VSD is advisable before establishment of CS and multiorgan failure. In case of established CS, either surgical or interventional closure is strongly indicated to ameliorate the otherwise grim prognosis. If the patient is rendered unsuitable for surgical closure, a percutaneous device therapy should be considered.

Acute phase patients without CS should undergo an immediate closure attempt as haemodynamic status allows. The role of primary percutaneous versus primary surgical closure remains unclear in this patient population. Percutaneous device closure may have a role as a bridge to surgery in selected patients28. If IABP fails5,29, active left ventricular assist devices may be useful in selected patients as a bridge to definitive treatment30,31.

AMI and consecutive VSD location may also play a crucial role because of the proximity of the valvular apparatus and possible interference with valvular function after device deployment in inferior/posterior VSD.

TREATMENT ALGORITHM

A practical treatment algorithm is depicted in Figure 4. Haemodynamic deterioration or a new murmur on auscultation should warrant prompt echocardiographic evaluation. After confirmation of a VSD, immediate closure irrespective of haemodynamic status is indicated. In the absence of randomised trials and guideline recommendations, we recommend that the decision whether to operate or to perform percutaneous device closure should be evaluated by a Heart Team based on the complexity of the defect, the maximum available occluder size, the fastest available option and the individual degree of interventional experience.

Figure 4. Potential treatment algorithm. Suggested treatment algorithm of ventricular septal defect (VSD) closure.

If a significant shunt reduction is achieved by either means of occlusion, close follow-up is warranted. If shunt reduction remains insufficient, a secondary percutaneous approach, switch to open heart surgery or in the case of a primary surgical approach a re-operation should be considered.

Study limitations

Our analysis does not reach the statistical accuracy of a meta-analysis since the underlying data lack a control group and are non-randomised. However, due to ethical issues and the infrequency of post-infarction VSD, randomised, controlled trials are difficult to perform in this patient population. Therefore, a direct comparison between surgical and interventional, as well as early versus late closure, is not possible owing to the non-randomised design and an inherent selection bias with all case series. Finally, although all VSD closure devices share a common morphology, the performance of different devices might also influence outcome. However, most patients were treated using Amplatzer devices and thus a sub-analysis according to the type of septal defect occluder was not possible.

Conclusion

In conclusion, percutaneous device closure of post-myocardial infarction VSD may represent a valuable alternative to surgical repair, with the advantage of immediate shunt reduction to prevent haemodynamic deterioration. A high rate of technically successful percutaneous procedures can be achieved. However, especially in CS patients, the mortality rate remains high.

Impact on daily practice

If left untreated, post-infarction VSD is associated with a devastating prognosis. Percutaneous device closure of post-myocardial infarction VSD is a valuable alternative to surgical repair with high rates of procedural success. Due to the lack of randomised data, the clinical decision regarding the type of treatment (surgical versus interventional approach) needs to be evaluated by a multidisciplinary team based on the complexity of the defect, the maximum available occluder size, the fastest available option and the individual degree of interventional experience.

Conflict of interest statement

The authors have no conflicts of interest to declare.


References

Volume 12 Number 1
May 16, 2016
Volume 12 Number 1
View full issue


Key metrics

Suggested by Cory

10.4244/EIJV12SXA4 May 16, 2016
Management of post-myocardial infarction ventricular septal rupture
Wilson W and Horlick E
free

Image – Interventional flashlight

10.4244/EIJ-D-18-00867 Jun 12, 2019
First European experience of percutaneous closure of ventricular septal defects using a new CE-marked VSD occluder
Schubert S et al
free

CLINICAL RESEARCH

10.4244/EIJV10I4A87 Aug 19, 2014
Percutaneous closure of ruptured sinus of Valsalva aneurysm: results from a multicentre experience
Liu S et al
free

10.4244/EIJV10I61A131 Oct 20, 2014
How should I treat an accidentally misplaced 8 Fr drainage catheter in the right ventricle?
Bakos Z and Harnek J
free
Trending articles
151.43

State-of-the-Art

10.4244/EIJ-D-22-00776 Apr 3, 2023
Computed tomographic angiography in coronary artery disease
Serruys PW et al
free
55.9

Clinical research

10.4244/EIJ-D-22-00621 Feb 20, 2023
Long-term changes in coronary physiology after aortic valve replacement
Sabbah M et al
free
54.9

Expert review

10.4244/EIJ-D-21-01010 Jun 24, 2022
Device-related thrombus following left atrial appendage occlusion
Simard T et al
free
43.75

Clinical Research

10.4244/EIJ-D-21-01091 Aug 5, 2022
Lifetime management of patients with symptomatic severe aortic stenosis: a computed tomography simulation study
Medranda G et al
free
39.95

Clinical research

10.4244/EIJ-D-22-00558 Feb 6, 2023
Permanent pacemaker implantation and left bundle branch block with self-expanding valves – a SCOPE 2 subanalysis
Pellegrini C et al
free
X

The Official Journal of EuroPCR and the European Association of Percutaneous Cardiovascular Interventions (EAPCI)

EuroPCR EAPCI
PCR ESC
Impact factor: 7.6
2023 Journal Citation Reports®
Science Edition (Clarivate Analytics, 2024)
Online ISSN 1969-6213 - Print ISSN 1774-024X
© 2005-2024 Europa Group - All rights reserved