DOI:

Between-centre reproducibility of volumetric intravascular ultrasound radiofrequency-based analyses in mild-tomoderate coronary atherosclerosis: an international multicentre study

Jennifer Huisman1, MD, MSc; Rasmus Egede2, MD; Adam Rdzanek3, MD; Dirk Böse4, MD; Raimund Erbel4, MD; Janusz Kochman3, MD, PhD; Lisette Okkels Jensen2, MD, PhD; Job van de Palen5,6, MSc, PhD; Marc Hartmann1, MD, PhD; Gary S. Mintz7, MD; Clemens von Birgelen1,8*, MD, PhD

Abstract

Aims: To assess for the first time in a multicentre design the between-centre reproducibility of volumetric virtual histology intravascular ultrasound (VH-IVUS) measurements with a semi-automated, computer-assisted contour detection system in mild-to-moderately diseased coronary segments.

Methods and results: Analysts of four European IVUS centres performed independent IVUS analyses (in total 7,188 cross-sectional analyses) and obtained volumetric data to evaluate the reproducibility of volumetric VH-IVUS measurements in 36 coronary segments (length 20.0±0.4 mm) from patients with stable angina. Geometric and compositional VH-IVUS measurements were highly correlated for the different comparisons. Overall intraclass correlation for vessel, lumen, plaque volume and plaque burden were 0.98, 0.92 0.95, and 0.86, respectively; for fibrous, fibro-lipidic, necrotic core and calcified volumes overall intraclass correlations were 0.95, 0.93, 0.99, and 1.00, respectively. There were significant but small differences for vessel, lumen, fibrous and calcified volumes, and there was no significant difference for plaque volume. Of the plaque components necrotic core and calcified volume showed on average the highest reproducibility.

Conclusions: These findings underline the necessity to centrally analyse IVUS data obtained in multicentre studies addressing mild-to-moderately diseased coronary arteries. In addition, pooling VH-IVUS data from different studies, analysed at different centres, may be problematical.

Abbreviations

IVUS: intravascular ultrasound

VH: virtual histology

Introduction

Greyscale intravascular ultrasound (IVUS) permits accurate quantification of vessel and plaque dimensions, but it has significant limitations in the assessment of plaque composition.1-4 Virtual Histology-IVUS (VH-IVUS) incorporates analysis of radiofrequency IVUS data to provide a better assessment of the coronary plaque components such as necrotic core and densely calcified area.5-14 Volumetric measurements are known to have a higher measurement reproducibility than cross-sectional area measurements which is likely the result of averaging differences between individual cross-sectional area measurements.15,16

VH-IVUS includes a semi-automated analyse function (i.e., computer-assisted contour detection) to provide volumetric data that are increasingly used for the assessment of coronary atherosclerosis and as endpoints of pharmacological studies for the treatment of coronary atherosclerosis.1,17-21 However, because recent serial VH-IVUS studies such as the IBIS-2 trial revealed only small changes in plaque volume and composition20, the reproducibility of VH-IVUS becomes an important issue.1,14,22,23 In a previous study in mild-to-moderately diseased coronary segments, we have demonstrated that the inter-observer variability of volumetric geometric and compositional VH-IVUS measurements by two analysts from the «same» centre is very acceptable.22 But so far, the between-centre reproducibility of VH-IVUS measurements has never been investigated.

Therefore, in the present study we assessed the «between-centre» reproducibility of volumetric VH-IVUS measurements in mild-to-moderately diseased coronary segments by comparing data from repeated analyses as performed by independent analysts at four different European IVUS-centres.

Methods

Study population

To assess the between-centre reproducibility of segmental volumetric VH-IVUS data analysis of mild-to-moderately diseased atherosclerotic human coronary arteries, we performed IVUS analyses in 20 mm-long segments of non-target and non-treated coronary arteries in patients with stable angina pectoris. Segments had to meet the following inclusion criteria: absence of ambiguous or significant coronary lesions (i.e., mild-to-moderate atherosclerotic disease), severe plaque calcification, major vessel tortuosity, and/or major side-branches. In this multicentre study design, four experienced analysts independently performed VH-IVUS analyses of all segments included in the study. Each centre was invited to submit IVUS pullbacks of 10 coronary segments obtained from patients in sinus rhythm in whom a clinically driven cardiac catheterisation required IVUS assessment. Participating centres were Odense University Hospital, Odense, Denmark; Essen University Hospital, Essen, Germany; Central University Hospital, Warsaw, Poland; and Thoraxcentrum Twente, Enschede, The Netherlands.

IVUS data acquisition

IVUS data were acquired with ECG-gating and commercially available phased-array IVUS catheters (Eagle Eye® Gold 2.9 Fr 20 MHz; Volcano Corporation, Rancho Cordova, CA, USA) with a dedicated console.2,24 The IVUS transducer was advanced ≥10 mm distal to the most distal side-branch. Angiographic cine runs were performed to define the position of the IVUS transducer. After intracoronary injection of 200 µg nitroglycerine, a continuous pullback of the IVUS catheter was performed using a motorised pullback device at 0.5 mm/s (TrackBack-II; Volcano Corporation, Rancho Cordova, CA, USA). IVUS image data were stored on digital video disk (DVD) for offline analysis.

VH-IVUS data analysis

Technique and the validation of VH-IVUS have previously been described.6,12 In brief, spectral IVUS radiofrequency data were reconstructed to generate tissue maps in order to classify the plaque in four major components: calcium; fibro-lipidic; fibrous; and necrotic core. Tissue components were displayed according to a colour code (white, light-green, green, and red, respectively).

VH-IVUS analysis of all cases was performed offline by one experienced IVUS analyst per centre (Figure 1).

Figure 1. Image acquisition and analysis. Each centre (A-D) performed motorized pullbacks at 0.5 mm/s, image frames were acquired at the time of the R-wave peak. These pullback sequences (obtained from 10 patients with stable angina) were exchanged in order to obtain a “pullback pool” of 40 pullbacks which each analyst independently analyzed (I). Automated contour detection of the lumen and vessel border was performed on all frames using the VH-images and software (II). Volumetric VH-IVUS data of the geometric and compositional data were automatically generated for the analyzed segment based on the application of the trapezium method to the cross-sectional area measurements (III).

The IVUS sequences were provided with a predefined region of interest that contained mild-to-moderate atherosclerotic disease (no target for percutaneous coronary interventions). Segments were located between two adjacent side-branches and contained no major calcification that could have limited reliable detection of the external vascular boundary.

IVUS measurements, both geometric and compositional measurements, were made using the VH-images and software. The lumen and vessel borders were traced using a semi-automated computer-assisted contour detection program (Volcano pcVH software program version 2.2; Corporation, Rancho Cordova, CA, USA); to exclude potential between version differences of the analysis software, all four centres used the same (latest) version of the analysis software. The leading edge of the blood-intima acoustic interface and the leading edge of the media-adventitia interface were automatically detected to define the lumen and vessel borders, respectively, and manually corrected as required. Plaque and media was used as a measure of atherosclerotic plaque and was calculated as the difference between vessel and lumen cross-sectional area. For each coronary segment, volumetric data of the vessel, lumen, plaque and plaque burden, as well as fibrous, fibro-lipidic, necrotic core, and calcified tissue were obtained from the application of the trapezium method to the cross-sectional area measurements. The pcVH software program uses this method that considers the exact spacing between image frames to calculate volumes from the cross-sectional area data in the most accurately way. In addition, the relative amount of the four plaque components was calculated.

In order to assess the between-centre reproducibility of volumetric VH-IVUS analyses, comparisons were performed on a 2-observer basis (i.e., centre A versus B, C and D respectively; centre B versus C and D respectively; and centre C versus D).

Statistical analysis

Statistical analysis was performed with SPSS 15.0 (SPSS Inc., Chicago, IL, USA). Dichotomous variables are presented as frequencies, while quantitative data are presented as mean ±1SD (or ±1SEM). Quantitative data were compared by use of a paired t-test after normality of the data was confirmed. A two-sided p-value <0.05 was considered as significant. According to Bland and Altman, the agreement between two measurements was assessed by determining the mean ±2SD of the between measurement differences.25 Intra-class correlations were calculated for comparisons at individual level. An intraclass correlation >0.90 was considered good.

Results

Study population

From the 40 segments (10 per centre) submitted for the purpose of this study, four segments did not meet the inclusion criteria (three major side branches and one patient with unstable angina) and were not included in the analysis. A total of 36 mild-to-moderately atherosclerotic coronary segments (10 right; 18 left anterior descending; and 8 left circumflex coronary arteries) from 36 patients with stable angina (27 males; 59±8 years old) were studied. Seventy-eight percent of the patients were on statin therapy prior to the cardiac catheterisation, and 17% were diabetics. There was no complication related to IVUS imaging.

VH-IVUS analysis and data

Each of the four analysts performed an independent analysis of each of the 36 coronary segments and obtained volumetric data to evaluate the reproducibility of volumetric VH-IVUS measurements (50±11 frames/segment). This approach required the analysis of a total of 7,188 cross-sectional frames (1,797 frames per analyst). Manual correction of the automatic border detection was required in almost all frames for a total analysis time of 59±5 min/segment. The length of the analysed segments was 20.0±0.4 mm (range: 19.2-20.7 mm). On average, the atherosclerotic segments contained predominantly fibrous (57.8%), fibro-lipidic (17.9%), and necrotic core (16.0%) tissue. Geometric and compositional IVUS data obtained by the analysts of the four different centres (A-D) are presented in Table 1.

Geometry

Geometric measurements were highly correlated for the different comparisons. There were significant differences for vessel and lumen volumes; however, these differences were relatively small (Table 2).

Plaque volume showed no significant between-centre difference. Plaque burden showed no significant differences for the comparison between centre A versus C and D, respectively, and B versus C; the other comparisons between the different centres were significant (P≤0.03), however, the (absolute) differences were small, ranging from 0.6-1.1%. Overall intraclass correlation and range of the individual between-centre comparisons (within brackets) were for vessel, lumen, plaque volume and plaque burden 0.98 (0.97-0.99), 0.92 (0.86-0.95), 0.95 (0.94-0.98) and 0.86 (0.78-0.92), respectively. The limits of agreement (i.e., 2 SD of mean difference) were small and similar for the different comparisons. The range of the standard deviation of the relative between centre differences for vessel, lumen, plaque volume and plaque burden were 2.5-3.0%, 6.4-9.3%, 6.7-10.1% and 5.9-9.3%, respectively. An example of the agreement is presented in Figure 2; the measurement reproducibility of vessel volume was particularly high. Measurement reproducibility was not influenced by the size of the measured volumes.

Figure 2. Example of the agreement of repeated VH-IVUS measurements of geometric and compositional volumes for two centres. Corresponding analyses were also performed for the other possible comparisons between the four centres, showing similar results (see text for further results).

Composition

VH-IVUS assessments of plaque composition were highly correlated for the different comparisons. Overall intraclass correlation and range of the individual between-centre comparisons (within brackets) for (absolute) fibrous, fibro-lipidic, necrotic core and calcified volume were 0.95 (0.92-0.97) 0.93 (0.90-0.96), 0.99 (0.98-1.00), and 1.00 (1.00-1.00), respectively. There were small significant differences for fibrous and calcified volumes (Table 2). The limits of agreement (i.e., 2 SD of mean difference) for compositional data were somewhat higher, but similar for the different comparisons. The range of the standard deviation of the relative between centre differences for fibrous, fibro-lipidic, necrotic core and calcified volume were 10.6-15.9%, 20.2-28.9%, 8.9-13.9% and 6.4-13.9%, respectively. An example of the agreement is presented in Figure 2. Necrotic core and calcified volumes showed the highest measurement reproducibility. Smaller compositional volumes showed a slightly higher between-centre variability.

Discussion

In the present multicentre study we assessed the between-centre reproducibility of measurements of vessel geometry and plaque composition in mild-to-moderately diseased atherosclerotic coronary segments. The overall intraclass correlations were high for geometric as well as compositional data (>0.86). The relative between-centre differences were low for vessel, lumen, and plaque volume (<4%, <9%, and <3%, respectively) as well as for fibrous, fibro-lipidic, necrotic core, and calcified plaque components (<6%, <11%, <3%, and <8%, respectively) demonstrating mild but systematic differences, most likely as a result of manual contour editing. Plaque burden measurements showed more differences (<5%), resulting in a higher variability and somewhat lower, but still good, intraclass correlations. This higher variability might be explained by the addition of variabilities of the variables themselves that are required for the calculation of plaque burden.

Measurement reproducibility

Serial VH-IVUS is increasingly used for the assessment of the natural history of atherosclerosis, vascular remodelling, and pharmacological interventions.1,4,17-21 As changes in plaque size and composition are often small,20 good measurement reproducibility is important.9,10,14,26

Rodriguez-Granillo et al investigated the measurement variability of VH-IVUS cross-sectional data obtained from 16 non-significant coronary atherosclerotic lesions in a single-centre study;14 inter-observer variability of geometric and compositional data was larger than inter-catheter variability particularly for indirect measurements such as plaque area. For geometric cross-sectional area measurements the relative intra-observer difference was ≤ 3% for lumen and vessel cross-sectional area and <11% for plaque cross-sectional area; conversely, compositional measurements showed larger measurement variability of up to 24%.14

Prasad et al assessed VH-IVUS measurements in significant lesions prior to and following percutaneous coronary intervention and demonstrated that volumetric VH-IVUS measurements of coronary segments had a high measurement reproducibility.23 The agreement between repeated geometric cross-sectional IVUS analyses was somewhat higher than that of cross-sectional compositional VH-IVUS analyses.23

We previously demonstrated a relatively high intra- and inter-observer reproducibility of volumetric geometric and compositional VH-IVUS data in 33 mild-to-moderately diseased coronary segments.22 Nevertheless, all these previous studies investigated the measurement variability of analysts from a single centre (or core lab) who were trained in one and the same way.

As demonstrated in the present multicentre study, measurements by analysts from four different IVUS centres with different training programs showed somewhat higher measurement variability and minor (though often significant) measurement differences. Differences may have resulted from slight systematic over- or underestimation of the lumen and/or vessel borders and/or differences in the interpolation of the vessel contour at the site of side branches. Necrotic core and calcified plaque volume (often located more centrally in the plaque) showed a particularly high measurement reproducibility while fibrous and fibro-lipidic volume showed more variability. This is in agreement with previous observations suggesting that variation in tracing the lumen contour (i.e., inclusion of the lumen in plaque and media volume) may result in misinterpretation of blood inside the lumen as fibrous and fibro-lipidic tissue. This may explain why these two tissue types show the largest relative difference between two centres. The centres with small lumen volume measured the largest content of fibrous and fibro-lipidic tissue (Table 2). Misinterpretation of thrombus and soft plaque adjacent to the lumen could theoretically have introduced a large measurement variability; however, this is extremely unlikely in the examined patient population with stable angina pectoris.2,6,9,12,27

Our present data underline the necessity to analyse such data at a single site - ideally in an experienced core-lab. As an alternative, one may consider that analysts of different centres undergo an identical training to assure a very similar (or identical) way of analysis. In addition, our findings suggest that pooling VH-IVUS data from different studies that are analysed at different IVUS-centres may be problematic. Awareness of the inter-observer differences may be particularly important in serial pharmacological intervention trials where small changes in plaque and/or vessel size as well as plaque composition may be expected.20

Limitations

In the present study, IVUS analyses were performed offline in coronary segments obtained from patients with stable angina pectoris; therefore our findings cannot be extrapolated to VH-IVUS reproducibility studies in an online setting or in patients with acute coronary syndromes. The coronary segments of the study population contained mild-to-moderate atherosclerotic disease. The study did not address significant coronary lesions with particularly small lumen dimensions (e.g., a lumen area <4.0mm2), in which the recognition of the lumen border may be more difficult and measurement variability may be higher.

Similar to other studies with IVUS and VH-IVUS, we excluded very tortuous and severely calcified vessels that could have led to non-uniform pullbacks and/or inability to detect the (external) vessel contour.14,22 The variability between different IVUS catheters or pullback devices, repeated pullbacks, and independent selections of the segment of interest were not addressed in the present multicentre study that focused on the assessment of measurement reproducibility in predefined segments with mild-to-moderate coronary disease.

Conclusion

These findings underline the necessity to centrally analyse IVUS data obtained in multicentre studies that address mild-to-moderately diseased coronary arteries. However, pooling VH-IVUS data from different studies analysed at different centres may be problematic.


References

Volume 5 Number 8
Apr 12, 2010
Volume 5 Number 8
View full issue


Key metrics

Suggested by Cory

10.4244/EIJV8I4A79 Aug 24, 2012
Non-invasive fractional flow reserve: scientific basis, methods and perspectives
Serruys PW et al
free
Trending articles
225.68

State-of-the-Art Review

10.4244/EIJ-D-21-00426 Dec 3, 2021
Myocardial infarction with non-obstructive coronary artery disease
Lindahl B et al
free
105.78

Expert consensus

10.4244/EIJ-E-22-00018 Dec 4, 2023
Definitions and Standardized Endpoints for Treatment of Coronary Bifurcations
Lunardi M et al
free
77.85

State-of-the-Art

10.4244/EIJ-D-23-00840 Sep 2, 2024
Aortic regurgitation: from mechanisms to management
Baumbach A et al
free
68.7

Clinical research

10.4244/EIJ-D-21-00545 Sep 20, 2022
Coronary lithotripsy for the treatment of underexpanded stents: the international; multicentre CRUNCH registry
Tovar Forero M et al
free
47.8

NEW INNOVATION

10.4244/EIJ-D-15-00467 Feb 20, 2018
Design and principle of operation of the HeartMate PHP (percutaneous heart pump)
Van Mieghem NM et al
free
45.3

Clinical research

10.4244/EIJ-D-18-01126 Aug 29, 2019
New-generation mechanical circulatory support during high-risk PCI: a cross-sectional analysis
Ameloot K et al
free
X

The Official Journal of EuroPCR and the European Association of Percutaneous Cardiovascular Interventions (EAPCI)

EuroPCR EAPCI
PCR ESC
Impact factor: 7.6
2023 Journal Citation Reports®
Science Edition (Clarivate Analytics, 2024)
Online ISSN 1969-6213 - Print ISSN 1774-024X
© 2005-2024 Europa Group - All rights reserved