DOI: 10.4244/EIJV14I9A175

The use of covered stents in the field of interventional procedures for congenital heart defects

Mario Carminati*, MD, FESC, FSCAI

The use of covered stents in the field of interventional procedures for congenital heart defects has become progressively more and more popular in recent years, either as a bail-out after complications related to angioplasty or bare metal stent implantation, or as a first-choice procedure to prevent such complications or to treat some defects electively, such as fenestration closure as described in the paper by Butera et al1.

In this issue of EuroIntervention, Y. Boudjemline describes his experience with the use of a new covered stent (BeGraft; Bentley Innomed GmbH, Hechingen, Germany), outlining the advantages as compared to other stents and, in particular, to the covered CP Stent® (NuMED Inc., Hopkinton, NY, USA), which was in fact the most commonly used covered stent in congenital interventional cardiology for many years2.

The enthusiasm of the author for the BeGraft stent is justified by the excellent results obtained; however, we have to remember that the covered CP Stent “did the job” with excellent results as well in hundreds of institutions all around the world for many years. We should be cautious of abandoning it in favour of new ones; more multicentre data and long-term follow-up are needed. In the recent past, a new stent (Advanta™; Maquet, Rastatt, Germany) seemed to be very promising but was quickly abandoned because of serious problems (infolding) that occurred in clinical practice, as correctly reported by the author in his paper. The characteristics of the BeGraft stent are in fact different and apparently safer and the stent performs better in our experience also. An example of BeGraft implantation in a case of “extreme” coarctation is shown in Figure 1.

Figure 1. BeGraft implantation in a case of “extreme” coarctation. A) Aortogram in a case of “extreme coarctation” (or aortic atresia); the contrast medium opacifies the ascending aorta and transverse arch but cannot progress beyond the atretic isthmus. B) Simultaneous contrast injection above (through the radial artery access) and below (through the femoral artery access) the atretic segment. C) The atretic segment was perforated with the transseptal needle and a Mullins sheath was advanced from the femoral artery up to the transverse aortic arch. The BeGraft covered stent mounted on the balloon is seen within the sheath. D) Aortogram after BeGraft covered stent expansion. The stent is perfectly apposed to the vessel wall, restoring normal flow.

On the other hand, the BeGraft stent is premounted on a proprietary single balloon; some operators may still prefer to use the CP Stent because it can be mounted on a BIB balloon and therefore the expansion of the inner balloon may allow (in some instances) a fine position adjustment before inflating the outer balloon for full stent expansion. Overall, the initial clinical experience with the BeGraft is very promising and it is likely to become more and more popular. To some extent the choice of material to be used is subjective and the final decision should be left in the hands of the individual operator.

Conflict of interest statement

The author has no conflicts of interest to declare.

Volume 14 Number 9
Oct 12, 2018
Volume 14 Number 9
View full issue


Key metrics

Suggested by Cory

SHORT REPORT

10.4244/EIJ-D-18-00287 Oct 12, 2018
Use of covered stents in the field of congenital heart diseases: the role of new players
Boudjemline Y
free

EXPERT REVIEW

10.4244/EIJV12I10A210 Nov 20, 2016
Coronary covered stents
Kilic D et al
free

EXPERT REVIEW

10.4244/EIJV11I6A133 Oct 20, 2015
The effectiveness of stenting of coarctation of the aorta: a systematic review
Hartman E et al
free

CLINICAL RESEARCH

10.4244/EIJY15M01_09 Mar 20, 2015
The VALEO® vascular stent for cardiovascular lesions in children
Ovaert C et al
free

10.4244/EIJV14I16A278 Mar 20, 2019
Should we implant a covered stent only when there is no other solution?
Colombo A and Mangieri A
free

10.4244/EIJ-D-19-00364L Feb 20, 2020
Very long-term outcome of coronary covered stents. Not all covered stents are the same
Jurado-Román A et al
free

Clinical research

10.4244/EIJ-D-19-00157 Feb 5, 2021
Early and midterm outcomes of bare metal stenting in small children with recurrent aortic coarctation
van Kalsbeek R et al
free
Trending articles
225.68

State-of-the-Art Review

10.4244/EIJ-D-21-00426 Dec 3, 2021
Myocardial infarction with non-obstructive coronary artery disease
Lindahl B et al
free
105.78

Expert consensus

10.4244/EIJ-E-22-00018 Dec 4, 2023
Definitions and Standardized Endpoints for Treatment of Coronary Bifurcations
Lunardi M et al
free
77.85

State-of-the-Art

10.4244/EIJ-D-23-00840 Sep 2, 2024
Aortic regurgitation: from mechanisms to management
Baumbach A et al
free
68.7

Clinical research

10.4244/EIJ-D-21-00545 Sep 20, 2022
Coronary lithotripsy for the treatment of underexpanded stents: the international; multicentre CRUNCH registry
Tovar Forero M et al
free
47.8

NEW INNOVATION

10.4244/EIJ-D-15-00467 Feb 20, 2018
Design and principle of operation of the HeartMate PHP (percutaneous heart pump)
Van Mieghem NM et al
free
45.3

Clinical research

10.4244/EIJ-D-18-01126 Aug 29, 2019
New-generation mechanical circulatory support during high-risk PCI: a cross-sectional analysis
Ameloot K et al
free
X

The Official Journal of EuroPCR and the European Association of Percutaneous Cardiovascular Interventions (EAPCI)

EuroPCR EAPCI
PCR ESC
Impact factor: 7.6
2023 Journal Citation Reports®
Science Edition (Clarivate Analytics, 2024)
Online ISSN 1969-6213 - Print ISSN 1774-024X
© 2005-2024 Europa Group - All rights reserved