DOI: 10.4244/EIJV9I10A190

The scientific power of a “sham” arm?

Patrick W. Serruys, Editor in Chief

We recently received the disappointing news that SYMPLICITY HTN-3 failed to meet its primary efficacy endpoint defined as a difference of 10 mmHg office systolic blood pressure between the sham and active renal denervation groups while it met the primary safety endpoint1. In response to this, the Chairmen of the Resistant Hypertension Course issued a statement urging caution that we need to wait “until the peer-reviewed publication of clinical data, most likely around the end of March 2014, to have further information rather than speculating2.” Whilst sharing the disappointment of those involved, our thoughts turn to a reminiscence of previous experiences with the use of “sham” technology at the beginning of this current century.

In 2000, the group of Peter Fitzgerald, Renu Virmani, Frank Kodologie and Paul Yock presented data regarding the application of intravascular sonotherapy. They showed, in a swine model, that sonotherapy decelerated cellular proliferation and decreased the growth of neointimal hyperplasia after stenting. Naturally, these results were considered as an effective form of non-ionising energy to reduce in-stent restenosis. The data, and in particular I would refer here to figure 2 of their paper (which was ultimately published in Circulation), led to the birth of a new hype3. Very soon, leading interventionists such as Jeffery Moses and Antonio Colombo demonstrated –with live cases at TCT– the Pharmasonics (Sunnyvale, CA, USA) technology for antirestenotic therapy in humans.

In the same way, our group in Rotterdam was very much attracted by the potential of this treatment for in-stent restenosis. We performed a pilot study with 37 patients but, unfortunately, we came to the conclusion that a late lumen loss of 1.05 mm and a restenosis rate of 25% was in keeping with conventional treatments4.

Nevertheless, the great captain of industry, Menahem Massi, wanted to demonstrate with certainty the antirestenotic aspects of this technique. It was the first time in our career, both in the EU and the US, that a trial was designed using a “sham”. This “sham” meant that the catheter had to dwell for 10 minutes in the stented lesion, which in itself could generate an adverse event. We convinced the hospital ethics board that for a sham treatment arm we would need to cannulate the artery with the catheter without actually activating the sonotherapy. The beauty of the trial was that it was truly blind as both treatment arms were cannulated, and when the operator “pressed the button of the device” he was unable to know whether activation of the sonotherapy was actually performed. When the study was completed, no difference between the two groups could be demonstrated. We were unable to publish this trial in a major journal, with only the now obsolete International Journal of Cardiovascular Interventions willing to publish the data5. This is a lesson I never forgot.

When you have a mechanical device that is supposed to make a physiological or biological change, and when the assessment of the result is based on indirect measurements of bioclinical effects such as blood pressure, it is essential to show that the placebo effect of the device is fully masked with a sham treatment. With the sham treatment, you can fully isolate the safety of the therapy and, likewise, the catheter itself and the intubation of the vessel, as it is in this case concerning the renal artery. The ablation could generate side-effects: in the case discussed here, the simple introduction of the catheter into the renal artery without doing anything else could do this, and the sham by itself allows you to observe whether there are any detrimental effects.

By performing a sham, on one hand, you isolate the positive from the negative aspects of the catheter and, on the other hand, you judge the bioclinical effect in a very objective manner, similar to using a placebo medication in randomised pharmacological studies. Another classic example of “sham” is the treatment of refractory angina pectoris by transmyocardial revascularisation through the creation of a reptilian heart by means of endocardial perfusion. The hype was spectacular until the group of Martin Leon randomised this therapy with sham therapy. As the laser treatment was noisy, both patient groups had headphones with classical music to protect the blinding6.

With respect to the use of sham in the Symplicity HTN-3 trial – and awaiting the publication of the results – to correlate sham with the study’s failure, is at this moment, pure speculation. One could postulate that other factors may have influenced the results.

For example, on a technical level, the cannulation of the catheter is relatively straightforward, however, the renal nerves are widely distributed and perhaps catheter orientation issues appeared. As in all new therapies, a learning curve exists, perhaps this was a contributing factor. The majority of the first studies, and the Symplicity HTN-1 and 2, were performed at selected sites in Europe and Australia, whereas the Symplicity HTN-3 trial was performed at selected sites in the United States and Australia. Despite the strict inclusion/exclusion criteria in HTN-3, did the patients truly have resistant hypertension? How did the investigators confirm medication adherence when a recent publication recently reported that 43 to 65.5% of patients with presumed resistant hypertension are non-adherent7? What about changes in antihypertensive drug treatment close to randomisation or within the study? Was the HTN-3 primary efficacy endpoint of a 10 mmHg systolic pressure difference between the two groups overly ambitious? Some experts explain that a long-term sustainable 5 mmHg decrease in systolic pressure is still beneficial for the patient group, suggesting that this decrease is the equivalent of adding one additional tablet.

Is renal denervation a hype that is now over? Only with a robust analysis of the HTN-3 results will we know, and it is at that time that we will have to ask ourselves –will we ever really know? Is the outcome of Symplicity HTN-3 related to the first generation device that has been used? What about the other investigational devices with different design and energy sources? Perhaps the focus will be redirected to the early work on various secondary indications, such as the role of renal denervation in the treatment of cardiac arrhythmias, chronic heart failure, chronic kidney disease and diabetes8.

Ultimately, the evolution of these treatment options for resistant hypertension, be it through pharmacological and/or non-pharmacological pathways, will always have a prominent place in cardiovascular research. The need for progress and further development is clearly justified by the immense patient group involved, and their significant risk for stroke, myocardial infarction and kidney failure.


References

Volume 9 Number 10
Feb 20, 2014
Volume 9 Number 10
View full issue


Key metrics

Suggested by Cory

10.4244/EIJV14I12A225 Dec 20, 2018
Renal artery denervation: a lot done and more to do
Sharif F
free

10.4244/EIJV12I18A355 Apr 7, 2017
Renal denervation for treatment of hypertension – will 2017 be the year of enlightenment?
Mahfoud F and Edelman ER
free

10.4244/EIJV10I10A187 Feb 20, 2015
Renal denervation reloaded: where to go from here?
Mahfoud F and Serruys PW
free

10.4244/EIJV10I2A30 Jun 20, 2014
Renal denervation revisited: comparative appraisal of safety and efficacy
Sakakura K and Joner M
free
Trending articles
225.68

State-of-the-Art Review

10.4244/EIJ-D-21-00426 Dec 3, 2021
Myocardial infarction with non-obstructive coronary artery disease
Lindahl B et al
free
105.78

Expert consensus

10.4244/EIJ-E-22-00018 Dec 4, 2023
Definitions and Standardized Endpoints for Treatment of Coronary Bifurcations
Lunardi M et al
free
77.85

State-of-the-Art

10.4244/EIJ-D-23-00840 Sep 2, 2024
Aortic regurgitation: from mechanisms to management
Baumbach A et al
free
68.7

Clinical research

10.4244/EIJ-D-21-00545 Sep 20, 2022
Coronary lithotripsy for the treatment of underexpanded stents: the international; multicentre CRUNCH registry
Tovar Forero M et al
free
45.3

Clinical research

10.4244/EIJ-D-18-01126 Aug 29, 2019
New-generation mechanical circulatory support during high-risk PCI: a cross-sectional analysis
Ameloot K et al
free
43.4

Clinical research

10.4244/EIJ-D-23-00590 Dec 4, 2023
Prognostic impact of cardiac damage staging classification in each aortic stenosis subtype undergoing TAVI
Nakase M et al
free
X

The Official Journal of EuroPCR and the European Association of Percutaneous Cardiovascular Interventions (EAPCI)

EuroPCR EAPCI
PCR ESC
Impact factor: 7.6
2023 Journal Citation Reports®
Science Edition (Clarivate Analytics, 2024)
Online ISSN 1969-6213 - Print ISSN 1774-024X
© 2005-2024 Europa Group - All rights reserved