Bioresorbable scaffolds, of which most experience has been obtained with the poly-L-lactide (PLLA)-based Absorb bioresorbable vascular scaffold (BVS; Abbott Vascular, Santa Clara, CA, USA), have the potential to overcome the limitations of metallic drug-eluting stents (DES), as they temporarily scaffold the diseased coronary vessel and then disappear. After degradation of the BVS, the coronary artery can restore its vasomotion and it may show favourable vascular remodelling1-3. During the last four years, more than 100,000 patients have been treated with BVS, while several million patients have been treated with metallic DES which showed favourable clinical outcomes in randomised clinical trials, including low target vessel revascularisation and stent thrombosis rates4,5.

So why has there not been a surge in the use of BVS in our daily clinical practice? A simple but important reason is the efficacy and proven long-term safety of the current-generation DES. When we recall the introduction of bare metal stents and – more than a decade later – DES, there were compelling reasons to introduce these devices rapidly, as in both cases the new devices solved or mitigated major clinical problems, for which there was no similarly effective alternative. BVS promise to overcome the problem of neoatherosclerosis formation in DES, which has been observed in both early and newer-generation DES6; however, considering the relatively low long-term event rates of patients treated with modern DES, this problem seems to be not overly common. The benefits of BVS are presumed to pay off in lower event rates after several years, but this remains to be demonstrated in ongoing large trials. While patients of all ages intuitively embrace the concept of a dissolving device, one may expect to see the greatest potential clinical benefit in the young and middle-aged.

In addition, recent clinical studies and meta-analyses have shown some signals of a potential increase in risk of (early) stent thrombosis with bioresorbable scaffolds7-9. While the significance of this finding is still unclear, the timing of scaffold thrombosis, early after implantation, suggests a relation with the relatively thick strut dimensions, the apposition to the vessel wall, and/or implantation techniques (e.g., aggressiveness of lesion preparation, adequate device sizing with or without intracoronary imaging, employment of post-dilatation, balloon size and pressures used). For thick-strut coronary stents and scaffolds alike, it is of paramount importance to appose the device well to the vessel wall9. In clinical practice, there is frequently a need for stent post-dilatation and overexpansion to achieve full device apposition10, and the same holds true for BVS. Severely oversized post-dilatations of metallic stents are generally well tolerated without destructive consequences to the metal frame, but may have a significant effect on stent geometry10. The consequences of BVS overexpansion have not been fully examined, so it was high time that a thorough evaluation of this matter was performed and published.

In the current issue of EuroIntervention, Foin et al report the results of meticulous bench-top experiments on the overexpansion of BVS with non-compliant balloons of increasing diameters11. While BVS of different sizes could be oversized by 1 mm in a non-constrained setting, only overexpansion with a non-compliant balloon 0.5 mm larger than the BVS size was feasible in a constraining eccentric arterial lesion model. Excessive overexpansion of BVS beyond that limit resulted in scaffold fractures and a decrease in mechanical support, as impressively demonstrated by mechanical point force measurements. This loss of mechanical support may trigger adverse clinical events. The work of Foin et al confirms the “no more than 0.5 mm larger” instruction for BVS post-dilatation with non-compliant balloons which was provided by the device manufacturer. In addition, interesting insights with 2D and 3D optical coherence tomography (OCT) in fractured BVS samples demonstrated an advantage of 3D OCT in detecting fracture sites. Partial overexpansion of BVS with non-compliant balloons beyond the safety limit of 0.5 mm resulted in fractures which were located, in particular, in the region between the post-dilated and non-post-dilated segments11. In parallel with these findings, previous studies of metallic DES demonstrated the greatest stent deformation to be present just in that region (Figure 1)10.

Figure 1. Micro-computed tomographic assessment following extremely oversized partial post-dilatation of metallic DES with an open-cell design. The maximum stent deformation with the largest cells (arrowheads) was seen in the transitional region between the 1.5 mm overexpanded and the non-post-dilated segments (A: TAXUS Liberté™; Boston Scientific, Natick, MA, USA; B: Resolute™; Medtronic, Santa Rosa, CA, USA). The region of maximum stent deformation corresponds with the scaffold region in the study by Foin et al11 which showed many BVS fractures after excessive partial overexpansion. Quantitative data on DES deformation have previously been reported10.

The safety limit of BVS overexpansion emphasises the fact that an optimised sizing of BVS is of paramount importance. In a way, this is an echo from the past, when intravascular ultrasound (IVUS) measurements taught the interventional community that bare metal stents had to be implanted and often post-dilated with larger balloons and at higher pressures to improve stent expansion and apposition12,13. IVUS use then resulted in an impressive reduction of stent thrombosis risk12. Careful pre-interventional assessment of reference vessel size and examinations of BVS implanted are likely to improve the clinical results in BVS too. In most of the recent BVS studies, quantitative coronary angiography was used for BVS sizing, but in some studies IVUS was also performed2. Pre-interventional IVUS can be highly valuable for the assessment of true vessel size and both vascular remodelling and tapering14. Imaging with OCT can also help to optimise BVS safely and to detect device fractures, which –of course– should be prevented based on pre-interventional measurements of vessel size11,15.

Does the current study provide answers to all relevant questions about the overexpansion of bioresorbable scaffolds? Well, probably not. While the authors obtained highly valuable data about Absorb BVS, they did not study other polymer- or magnesium-based bioresorbable scaffolds which have different physical properties16,17. For instance, a PLLA-based scaffold, which regains its diameter after acute recoil due to the “self-correcting” properties of the device, was previously shown to have a much greater tolerance to significantly oversized post-dilations than the Absorb BVS16. Finally, Foin et al used expired BVS for their experiments, and polymer ageing could have affected the physical properties of the devices tested18. Therefore, it would be of great interest to perform additional in vitro studies with fresh samples of the Absorb BVS and other bioresorbable scaffolds, using the same experimental set-up as applied in the current study.

To conclude, well-controlled post-dilatation and overexpansion of current BVS is likely to prevent a substantial proportion of potential scaffold thromboses by improving the BVS apposition to the vessel wall. On the other hand, excessive overexpansion can lead to fractures of the polymer that may induce adverse events such as scaffold thromboses. 3D OCT, a technique which Foin et al used in their current bench-top study to identify BVS fractures, could not only help reveal the mechanisms of thrombus formation in BVS, but has – together with IVUS and other advanced quantitative coronary imaging techniques – the potential to reduce the risk of fracturing struts.

Funding

The Research Department of Thoraxcentrum Twente has received research grants from AstraZeneca, Biotronik, Boston Scientific, and Medtronic.

Conflict of interest statement

C. von Birgelen has been a consultant for Boston Scientific and Medtronic, and has received speaker’s honoraria from MSD and AstraZeneca. The other author has no conflicts of interest to declare.


References

Volume 11 Number 12
Mar 18, 2016
Volume 11 Number 12
View full issue


Key metrics

Suggested by Cory

Debate

10.4244/EIJ-E-23-00015 Jun 19, 2023
Bioresorbable coronary scaffolds are ready for a comeback: pros and cons
Stone GW et al
free

EXPERT REVIEW

10.4244/EIJ-D-17-00499 Aug 25, 2017
State of the art: the inception, advent and future of fully bioresorbable scaffolds
Katagiri Y et al
free

10.4244/EIJV13I7A115 Sep 20, 2017
Bioresorbable scaffolds: reflections after a setback - losing a battle does not mean losing the war!
Colombo A and Azzalini L
free

Dec 30, 2009
A.R.T.: concept of a bioresorbable stent without drug elution
Lafont A and Durand E
free

CLINICAL RESEARCH

10.4244/EIJV10I4A78 Aug 19, 2014
Lessons learned from acute and late scaffold failures in the ABSORB EXTEND trial
Ishibashi Y et al
free

10.4244/EIJV15I1A5 May 20, 2019
Should we stop using bioresorbable scaffolds in coronary revascularisation?
Mangieri A and Colombo A
free

10.4244/EIJV16I2A16 Jun 12, 2020
Bioresorbable scaffolds: did we jump the gun?
Waksman R and Forrestal B
free

10.4244/EIJV11SVA42 May 19, 2015
Treatment of bioresorbable scaffold failure
Felix C et al
free

10.4244/EIJV13I13A242 Jan 19, 2018
Polymeric bioresorbable coronary scaffolds: the hype is over, but the dream lives on
Pyxaras S and Wijns W
free
Trending articles
151.43

State-of-the-Art

10.4244/EIJ-D-22-00776 Apr 3, 2023
Computed tomographic angiography in coronary artery disease
Serruys PW et al
free
55.9

Clinical research

10.4244/EIJ-D-22-00621 Feb 20, 2023
Long-term changes in coronary physiology after aortic valve replacement
Sabbah M et al
free
54.9

Expert review

10.4244/EIJ-D-21-01010 Jun 24, 2022
Device-related thrombus following left atrial appendage occlusion
Simard T et al
free
43.75

Clinical Research

10.4244/EIJ-D-21-01091 Aug 5, 2022
Lifetime management of patients with symptomatic severe aortic stenosis: a computed tomography simulation study
Medranda G et al
free
39.95

Clinical research

10.4244/EIJ-D-22-00558 Feb 6, 2023
Permanent pacemaker implantation and left bundle branch block with self-expanding valves – a SCOPE 2 subanalysis
Pellegrini C et al
free
X

The Official Journal of EuroPCR and the European Association of Percutaneous Cardiovascular Interventions (EAPCI)

EuroPCR EAPCI
PCR ESC
Impact factor: 7.6
2023 Journal Citation Reports®
Science Edition (Clarivate Analytics, 2024)
Online ISSN 1969-6213 - Print ISSN 1774-024X
© 2005-2024 Europa Group - All rights reserved