Original Research

DOI: 10.4244/EIJ-D-24-00619

Redo-TAVI with the SAPIEN 3 valve in degenerated calcified CoreValve/Evolut explants

David Meier1,2, MD; Anish Nigade3, MSc; Althea Lai4, BSc; Kyle Dorman3, BSc; Hacina Gill4, BSc; Shahnaz Javani3, MSc; Mariama Akodad5, MD, PhD; David A. Wood4,2, MD; Toby Rogers6,7, MD, PhD; Rishi Puri8, MD, PhD; Keith B. Allen9, MD; Adnan K. Chhatriwalla9, MD; Michael J. Reardon10, MD; Gilbert H.L. Tang11, MD, MSc, MBA; Vinayak N. Bapat12, MD; John G. Webb4,2, MD; Shinichi Fukuhara13, MD; Stephanie L. Sellers4,2, PhD, MSc

Abstract

Background: Redo-transcatheter aortic valve implantation (TAVI) is the treatment of choice for failed transcatheter aortic valves. Currently, implantation of a SAPIEN 3 (S3) is indicated for redo-TAVI in degenerated CoreValve/Evolut (CV/EV) transcatheter aortic valves (TAVs) but is not well understood.

Aims: We aimed to evaluate S3 function following implantation in explanted calcified CV/EV TAVs and to assess the impact of CV/EV pathology on redo-TAVI outcomes.

Methods: Ex vivo hydrodynamic testing was performed per the International Organization for Standardization (ISO) 5840-3 standard on 4 S3 TAVs implanted at node 5 in calcified CV/EV explants. The mean gradient (MG), effective orifice area (EOA), peak velocity, regurgitant fraction (RF), geometric orifice area (GOA), leaflet overhang, leaflet pinwheeling, neoskirt height, and frame deformation were evaluated.

Results: CV/EV explants were calcified and stenotic. Following S3 implantation, the MG and peak velocity decreased. As per the ISO standard, all S3 implants showed adequate EOA, and 3 out of 4 had an RF within the accepted value (<20%). CV/EV leaflet overhang ranged from 25-37%. Calcified leaflets remained stationary throughout the cardiac cycle (difference <9%) and were not pinned in a manner that constrained S3 systolic flow or appeared to prevent selective frame cannulation. The downstream CV/EV GOA was larger than the upstream S3 GOA during systole. S3 frame underexpansion was seen, resulting in leaflet pinwheeling (range 13-30%). Above the neoskirt, calcium protrusion was observed in contact with the S3 leaflets.

Conclusions: S3 implantation at node 5 in calcified CV/EV valves resulted in satisfactory hydrodynamic performance in most configurations tested with stable leaflet overhang throughout the cardiac cycle. The long-term implications of S3 underexpansion, leaflet pinwheeling, and calcium protrusion require future studies.

Sign in to read
the full article

Forgot your password?
No account yet?
Sign up for free!

Create my pcr account

Join us for free and access thousands of articles from EuroIntervention, as well as presentations, videos, cases from PCRonline.com

Volume 20 Number 22
Nov 18, 2024
Volume 20 Number 22
View full issue


Key metrics

Suggested by Cory

Translational research

10.4244/EIJ-D-22-00769 May 12, 2023
A bench study of balloon-expandable valves for the treatment of self-expanding valve failure
Akodad M et al
free

Original Research

10.4244/EIJ-D-24-00367 Nov 18, 2024
Feasibility of redo-TAVI in the self-expanding ACURATE neo2 valve: a computed tomography study
Bieliauskas G et al

Original Research

10.4244/EIJ-D-23-00996 May 10, 2024
Performance and outcomes of the SAPIEN 3 Ultra RESILIA transcatheter heart valve in the OCEAN-TAVI registry
Yamamoto M et al

Original Research

10.4244/EIJ-D-23-00783 Mar 18, 2024
Redo-TAVI with the ACURATE neo2 and Prime XL for balloon-expandable transcatheter heart valve failure
Meier D et al

Clinical research

10.4244/EIJ-D-22-01125 Jul 17, 2023
Feasibility of redo-TAVI in self-expanding Evolut valves: a CT analysis from the Evolut Low Risk Trial substudy
Grubb KJ et al
free
Trending articles
225.43

State-of-the-Art Review

10.4244/EIJ-D-21-00426 Dec 3, 2021
Myocardial infarction with non-obstructive coronary artery disease
Lindahl B et al
free
77.1

State-of-the-Art

10.4244/EIJ-D-23-00840 Sep 2, 2024
Aortic regurgitation: from mechanisms to management
Baumbach A et al
free
47.8

NEW INNOVATION

10.4244/EIJ-D-15-00467 Feb 20, 2018
Design and principle of operation of the HeartMate PHP (percutaneous heart pump)
Van Mieghem NM et al
free
43.4

Clinical research

10.4244/EIJ-D-23-00590 Dec 4, 2023
Prognostic impact of cardiac damage staging classification in each aortic stenosis subtype undergoing TAVI
Nakase M et al
free
33.4

Translational research

10.4244/EIJ-D-23-00308 Nov 17, 2023
Redo-TAVI with SAPIEN 3 in SAPIEN XT or SAPIEN 3 – impact of pre- and post-dilatation on final THV expansion
Meier D et al
free
30.1

Original Research

10.4244/EIJ-D-24-00247 Aug 19, 2024
Changes in absolute coronary flow and microvascular resistance during exercise in patients with ANOCA
Zeitouni M et al
X

The Official Journal of EuroPCR and the European Association of Percutaneous Cardiovascular Interventions (EAPCI)

EuroPCR EAPCI
PCR ESC
Impact factor: 7.6
2023 Journal Citation Reports®
Science Edition (Clarivate Analytics, 2024)
Online ISSN 1969-6213 - Print ISSN 1774-024X
© 2005-2024 Europa Group - All rights reserved