INTERVENTIONAL FLASHLIGHT

DOI: 10.4244/EIJ-D-18-00030

Pressure wire assessment of jailed balloon technique efficacy in a patient with complex bifurcated lesion

Hiroyuki Omori*, MD; Yoshiaki Kawase, MD; Toru Tanigaki, MD; Hitoshi Matsuo, MD, PhD

An 85-year-old man was admitted to our hospital with effort-related chest pain. Coronary angiography revealed moderate stenosis in the proximal left anterior descending artery (LAD) and the large side branch (SB) (Medina 0,1,1) (Panel A, Panel B). The jailed balloon technique was applied to protect the SB1. A pressure wire (OptoWire®; Opsens, Quebec, Canada) was used for measuring Pd/Pa during and at the end of the procedure. A 2.0×30 mm Ryujin™ Plus balloon (Terumo Corp., Tokyo, Japan) was used to protect the SB. XIENCE 3.5×23 mm and 3.0×28 mm stents (Abbott Vascular, Santa Clara, CA, USA) were deployed to the LAD lesion (Panel C). The Pd/Pa in the SB immediately after stent implantation was 0.34 (Panel D). The jailed balloon was then withdrawn easily, upon which Pd/Pa changed to 0.60 (Panel E). As the Pd/Pa value was suboptimal, kissing balloon inflation was performed without difficulty to recross a conventional wire. The final values of fractional flow reserve in the LAD and SB were 0.89 and 0.84, respectively. The final angiogram showed TIMI grade 3 flow in both the LAD and SB (Panel F).

The present report shows a beneficial effect of the jailed balloon technique using a pressure wire. The Pd/Pa dramatically increased after balloon removal, indicating recovery of flow in the jailed SB. This demonstrated that the jailed balloon could secure the lumen in the jailed SB.

FFR in an SB can provide significant information. Koo et al reported that only 27% of lesions had FFR <0.75 when SB stenosis was more than 75%2. The DKCRUSH-VI trial reported that there was less SB treatment in the FFR-guided group compared with the angiography-guided group (56% vs. 63%, p=0.07). However, the rate of failure to measure FFR of a jailed SB was 9.4%3.

The advantages of jailed balloon protection are enhanced SB protection by ostium spatial occupation of a jailed balloon and immediate SB flow restoration by eventual balloon dilation4. A potential strength of this technique is that jailed balloon protection may be used to facilitate the jailed pressure wire technique by protecting the pressure wire from both entrapment and risk of damage due to crossover stenting. We have evaluated only one case; however, we will evaluate a series of bifurcation cases in the future.

Conflict of interest statement

The authors have no conflicts of interest to declare.

Volume 14 Number 3
Jun 14, 2018
Volume 14 Number 3
View full issue


Key metrics

Suggested by Cory

Clinical research

10.4244/EIJ-D-20-00334 Aug 6, 2021
Feasibility and efficacy of an ultra-short side branch-dedicated balloon in coronary bifurcation stenting
Murasato Y et al
free

10.4244/EIJV11SVA27 May 19, 2015
When and how to perform the provisional approach for distal LM stenting
Park S et al
free

Image – Interventional flashlight

10.4244/EIJ-D-19-00973 Apr 2, 2021
Relationship between fractional flow reserve value and the amount of subtended myocardium
Tsujimoto M et al
free
Trending articles
225.68

State-of-the-Art Review

10.4244/EIJ-D-21-00426 Dec 3, 2021
Myocardial infarction with non-obstructive coronary artery disease
Lindahl B et al
free
105.78

Expert consensus

10.4244/EIJ-E-22-00018 Dec 4, 2023
Definitions and Standardized Endpoints for Treatment of Coronary Bifurcations
Lunardi M et al
free
77.85

State-of-the-Art

10.4244/EIJ-D-23-00840 Sep 2, 2024
Aortic regurgitation: from mechanisms to management
Baumbach A et al
free
68.7

Clinical research

10.4244/EIJ-D-21-00545 Sep 20, 2022
Coronary lithotripsy for the treatment of underexpanded stents: the international; multicentre CRUNCH registry
Tovar Forero M et al
free
47.8

NEW INNOVATION

10.4244/EIJ-D-15-00467 Feb 20, 2018
Design and principle of operation of the HeartMate PHP (percutaneous heart pump)
Van Mieghem NM et al
free
45.3

Clinical research

10.4244/EIJ-D-18-01126 Aug 29, 2019
New-generation mechanical circulatory support during high-risk PCI: a cross-sectional analysis
Ameloot K et al
free
X

The Official Journal of EuroPCR and the European Association of Percutaneous Cardiovascular Interventions (EAPCI)

EuroPCR EAPCI
PCR ESC
Impact factor: 7.6
2023 Journal Citation Reports®
Science Edition (Clarivate Analytics, 2024)
Online ISSN 1969-6213 - Print ISSN 1774-024X
© 2005-2024 Europa Group - All rights reserved