SPECIAL REPORT

DOI: 10.4244/EIJY15M07_05

Direct Flow valve-in-valve implantation in a degenerated mitral bioprosthesis

Giuseppe Bruschi1*, MD, FESC; Aldo Cannata1, MD; Alberto Barosi1, MD; Paola Colombo1, MD, PhD; Francesco Soriano1, MD; Stefano Nava1, MD; Elisa Montrasio2, MD; Luca Botta1, MD, PhD; Maria Pia Gagliardone2, MD; Silvio Klugmann1, MD; Federico De Marco1, MD, PhD

Abstract

Aims: Mitral valve reoperations due to failing bioprostheses, in patients affected by multiple comorbidities, are associated with high morbidity and mortality. Transcatheter techniques may evolve as complementary approaches to surgery in these patients at high risk for surgery.

Methods and results: We describe a case of Direct Flow 25 mm transcatheter valve implantation as valve-in-valve in a degenerated mitral bioprosthesis through a transapical approach in a 63-year-old man affected by dilated cardiomyopathy. The patient was affected by Carpentier-Edwards 29 mm severe regurgitation. The 25 mm Direct Flow bioprosthesis was advanced through the mitral bioprosthesis into the left atrium and then positioned using the three independent positioning wires. Transoesophageal echocardiography evidenced normal Direct Flow function with no paravalvular regurgitation and a low transmitral gradient of 4 mmHg.

Conclusions: Our successful experience, characterised by a Heart Team approach and multidisciplinary patient care, demonstrated the technical feasibility and procedural safety of Direct Flow valve-in-valve mitral implantation.

Transcatheter mitral valve-in-valve implantation into failing surgical bioprosthetic valves has been reported previously. It has evolved in recent years as a promising complementary therapy to avoid repeat cardiac surgery in patients affected by multiple comorbidities and who are considered at high risk for surgical reoperation1,2. Different approaches to access the mitral bioprosthesis or ring have been described, including the transseptal and transatrial techniques. Nevertheless, most experience has been gained using the transapical access3-5.

The current standard when treating degenerated mitral bioprostheses with transcatheter valves is the use of Edwards balloon-expandable devices (Edwards Lifesciences, Irvine, CA, USA). This technique gives good overall results but it is limited by lack of repositionability and retrievability and the issues associated with residual regurgitation and malpositioning. The Direct Flow Medical® Transcatheter Aortic Valve System (Direct Flow Medical Inc., Santa Rosa, CA, USA) has the potential to overcome these limitations.

We describe a case of Direct Flow Medical transcatheter valve implantation as valve-in-valve in a degenerated mitral bioprosthesis through the transapical approach.

The Direct Flow Medical device is the first transcatheter valve device which is not based on a metallic frame technology. The bovine pericardial prosthesis has an inflatable and deflatable support structure that allows precise positioning, repositioning, and retrieval if needed.

Full assessment of the valve performance in its intended final configuration can be carried out with multiple modalities, including complete echographic assessment, before permanent fixation with a durable polymer6.

Case report

A 63-year-old man (height: 185 cm, weight: 135 kg, body mass index: 39.4 kg/m2) affected by dilated cardiomyopathy was admitted to our department for worsening dyspnoea at rest. His symptoms were consistent with New York Heart Association Class IV. In 2008, the patient had undergone coronary artery revascularisation and concomitant mitral valve replacement with a Carpentier-Edwards 29 mm bioprosthesis (Edwards Lifesciences). In 2012, he experienced a non-ST-elevation myocardial infarction treated by percutaneous coronary intervention. His comorbid conditions included obstructive sleep apnoea syndrome treated with nocturnal continuous positive airway pressure therapy, chronic atrial fibrillation, pulmonary hypertension and moderate chronic kidney disease. In recent years, the patient had had several hospital admissions for heart failure with evidence of mitral bioprosthetic dysfunction.

Echocardiography on admission (Figure 1A) confirmed left ventricular (LV) dysfunction with an LV ejection fraction of 35% and severe bioprosthetic mitral stenosis (valve area, 0.7 cm2; mean gradient, 10 mmHg). This was associated with severe regurgitation and an elevated pulmonary artery systolic pressure of 60 mmHg.

Figure 1. Echocardiographic and CT scan mitral bioprosthesis evaluation. A) Mitral bioprosthesis severe regurgitation. B) CT scan mitral bioprosthesis evaluation at the level of the sewing ring, diameter 25 mm and perimeter of 80.2 mm. C) CT scan evaluation of the space in the left atrium to unsheathe the Direct Flow bioprosthesis. D) Transapical insertion of a Check-Flo Cook sheath for Direct Flow valve deployment.

According to our institutional policy the patient was jointly evaluated by our Heart Team, adopting the same philosophy we use for transcatheter aortic valve implantation. A transcatheter mitral valve-in-valve procedure was considered since the patient was a high-risk surgical candidate due to his comorbidities and previous cardiac surgery (EuroSCORE II: 24.59%, logistic 47%; and Society of Thoracic Surgeons risk of mortality 27.1%, and morbidity and mortality 62%).

After careful patient and procedural evaluation, and on the basis of our own and other authors’ experience of Edwards SAPIEN valve migration or severe paravalvular regurgitation after mitral valve-in-valve procedures1,7,8, we decided to perform a Direct Flow Medical device implantation through the transapical approach. Detailed written informed consent was obtained from our patient.

The procedure was performed in our hybrid suite by our Heart Team composed of cardiac surgeons with expertise in hybrid procedures, interventional cardiologists, cardiac anaesthesiologists and echocardiographists.

The procedure was performed under general anaesthesia and double lumen intubation to allow single lung ventilation. Defibrillator pads were properly placed across the chest wall. During the procedure, unfractionated heparin 100 IU/kg was administered to achieve an activated clotting time of 200 to 250 seconds for the duration of the procedure. This was then reversed with protamine at the end of the operation. A 6 Fr sheath was inserted in the left femoral artery for haemodynamic monitoring and one in the right femoral vein to have the possibility to establish a femoro-femoral emergency cardiopulmonary bypass support in case of haemodynamic collapse due to procedural issues. A standard transapical approach was performed by left anterolateral mini-thoracotomy at the fifth intercostal space; the pericardium was dissected from the epicardium to expose the LV apex. After coaxial trajectory evaluation, between entry site and mitral bioprosthesis (Figure 1B, Figure 1C), two paired orthogonal U-shaped sutures with pledgets were placed to secure the apex. Two temporary epicardial pacing leads were placed on the LV. The left ventricle was accessed using the Seldinger technique through the double sutures, a 12 Fr sheath was inserted and mitral bioprosthesis crossed with a straight wire. After replacement of the soft guidewire by an Amplatz Super-Stiff™ (Cook Medical, Bloomington, IN, USA), positioned across the mitral bioprosthesis into the right lower pulmonary vein, a 24 Fr sheath was advanced to deploy the Direct Flow device (Figure 1D). Mitral prosthesis balloon valvuloplasty was performed under rapid pacing using a Loma Vista Medical TRUE Dilatation™ Balloon Valvuloplasty Catheter, 24 mm (Bard Medical, Covington, GA, USA). A 25 mm Direct Flow bioprosthesis was advanced through the mitral bioprosthesis into the left atrium (Figure 2A). After unsheathing, the valve rings were inflated at 12 atmospheres in the left atrium with a contrast-saline mixture through the positioning wires of the delivery system to allow for valve unfolding. The bioprosthesis was then positioned using the three independent positioning wires with a modified inner curve technique in a similar way to that which is routinely carried out during TAVI with the Direct Flow device9 (Figure 2B-D). Some minimal repositioning performed by deflating only the larger (ventricular) ring and pulling on the positioning wires was carried out in order to position the ventricular ring on the top of the tip of stent posts. Fluoroscopic optimal expansion of the Direct Flow valve inside the Carpentier-Edwards bioprosthesis was confirmed by the symmetric and circular inflation of the atrial ring inside the bioprosthetic stent at the level of the sewing ring (Figure 2E). Transoesophageal echocardiography evidenced normal Direct Flow function (Figure 2F) with no paravalvular regurgitation and a low transmitral gradient of 4 mmHg. The contrast-saline mixture was then exchanged for the polymer to allow permanent valve fixation. Positioning wires were then detached and the delivery system removed. LV sutures were knotted regularly, thoracotomy was closed in the standard fashion, and the patient was transferred to the intensive care unit for recovery. He was extubated 48 hours after the procedure. The patient was discharged home on the ninth postoperative day, thoracotomy wound heal by primary intention.

Figure 2. Direct Flow valve-in-valve mitral implantation. A) Direct Flow was unsheathed in the left atrium and valve rings were inflated at 12 atmospheres to unfold the valve. B) The bioprosthesis was then positioned using the three independent positioning wires with a modified inner curve technique. C) Atrial ring was positioned inside the bioprosthetic stent at the level of the sewing ring. D) Direct Flow ventricular ring was positioned on the top of the tip of stent posts of the Carpentier-Edwards bioprosthesis. E) Final Direct Flow position after positioning wires have been detached. F) 3D echocardiographic image of the atrial Direct Flow ring positioned inside the bioprosthetic stent at the level of the sewing ring.

Transcatheter mitral valve-in-valve implantation with the Edwards SAPIEN balloon-expandable device has proved to be an attractive alternative to conventional reoperation for elderly high-risk surgical patients with bioprosthetic degeneration. Nevertheless, in the literature, cases of delayed valve migration have been reported together with cases of significant residual regurgitation, mainly related to the choice of size of the TAVI device or its suboptimal placement unable to prevent embolisation1,7,8. In our case we implanted a Direct Flow 25 mm device in a Carpentier-Edwards Perimount 29 mm bioprosthesis. Valve size was chosen mainly relying on computed tomography sizing (Figure 1B) and not on the internal diameter reported by the manufacturer. The internal diameter of the Perimount at the level of the sewing ring on CT scan measured 25 mm (derived from a perimeter of 80.2 mm). Since the Direct Flow device in the aortic location does not require significant oversizing, we opted for a 1:1 ratio between the inner diameter of the Perimount and that of the atrial ring of the Direct Flow device. Our assumption was that the shape of the Direct Flow device with the unique double ring design of the valve could create a double seal - both around the atrial portion of the bioprosthesis and also at the level of the tip of stent posts of the mitral bioprosthesis. Moreover, we speculated that the larger ventricular ring expanded beyond the tip of the stent posts could prevent delayed atrial embolisation of the TAVI device. This is purely a hypothesis. Obviously no definite conclusions about a rare complication such as delayed migration can be drawn from a single case.

Size choice might appear conservative as we selected a 25 mm device for a device having an inner diameter of 27 mm. Being a retrievable device, we chose to start with a relatively small size. In case of a suboptimal result, the planned strategy would have been that of retrieving it and upsizing to a 27 mm Direct Flow Medical (DFM) device. This did not happen due to the acceptable haemodynamics obtained with the first device. If retrieval had been carried out, it would have been done according to a modified technique to retrieve the device in very limited spaces which has been tested effectively on the bench. Basically, the two main differences from the standard technique are: using a larger sheath so that less force is needed when extracting the device, and opening the retrieve basket inside the sheath and carefully advancing it from its tip to “encage” the valve instead of opening the basket inside the body and pulling the valve inside the basket.

Our successful experience, characterised by a Heart Team approach and multidisciplinary patient care, demonstrated the technical feasibility and procedural safety of Direct Flow valve-in-valve mitral implantation.

Impact on daily practice

A less invasive approach for mitral re-replacement is desirable not only for a high-risk elderly population. Transcatheter mitral valve-in-valve implantation for a dysfunctional biological mitral prosthesis can be performed with low operative mortality. Even if further improvements are needed before widespread clinical application, significant clinical impact can be anticipated, not only because the absolute number of failing mitral biological prostheses will continue to grow, but also because a change in the surgical community should be anticipated, favouring mitral valve replacement with a bioprosthesis to treat functional mitral regurgitation instead of repairing the valve.

Conflict of interest statement

G. Bruschi is a consultant for Medtronic and Direct Flow. F. De Marco is a consultant for Direct Flow. The other authors have no conflicts of interest to declare.

Supplementary data

Moving image 1. Mitral bioprosthesis echocardiographic evaluation.

Moving image 2. The Direct Flow valve is advanced through the mitral bioprosthesis in the left atrium and unsheathed.

Moving image 3. The Direct Flow valve was positioned inside mitral bioprosthesis using the three independent positioning wires.

Moving image 4. Final adjustment of the Direct Flow position and valve inflation.

Moving image 5. Final Direct Flow valve position. After polymer fixation, the positioning wires were detached.


References

Volume 11 Number 13
Apr 20, 2016
Volume 11 Number 13
View full issue


Key metrics

On the same subject

Image – Interventional flashlight

10.4244/EIJ-D-20-00426 Sep 18, 2020
Technical considerations for repeat transcatheter aortic valve implantation in a failing Direct Flow Medical transcatheter heart valve
Yap J et al
free

Image – Interventional flashlight

10.4244/EIJ-D-21-00630 Apr 22, 2022
Embolisation of a prosthetic mitral valve fragment during valve-in-valve transcatheter mitral valve implantation
Alnsasra H et al
free

CLINICAL RESEARCH

10.4244/EIJY15M07_01 Apr 20, 2016
First Lotus aortic valve-in-valve implantation to treat degenerated Mitroflow bioprostheses
Castriota F et al
free

INTERVENTIONAL FLASHLIGHT

10.4244/EIJ-D-17-00183 Feb 2, 2018
Going the extra mile: percutaneous transseptal valve-in-valve implantation and leak closure
Ribeiro J et al
free

SPECIAL REPORT

10.4244/EIJ-D-16-00896 Jan 20, 2017
First successful transcatheter double valve replacement from a transapical access and nine-month follow-up
Bauernschmitt R et al
free

SPECIAL REPORT

10.4244/EIJY15M09_08 Jul 20, 2016
First-in-man treatment of a degenerated mitral surgical valve with the mechanical expanding Lotus™ valve
Schäfer U et al
free
Trending articles
338.63

State-of-the-Art Review

10.4244/EIJ-D-21-00904 Apr 1, 2022
Antiplatelet therapy after percutaneous coronary intervention
Angiolillo D et al
free
295.45

Expert consensus

10.4244/EIJ-D-21-00898 Sep 20, 2022
Intravascular ultrasound guidance for lower extremity arterial and venous interventions
Secemsky E et al
free
226.03

State-of-the-Art Review

10.4244/EIJ-D-21-00426 Dec 3, 2021
Myocardial infarction with non-obstructive coronary artery disease
Lindahl B et al
free
209.5

State-of-the-Art Review

10.4244/EIJ-D-21-01034 Jun 3, 2022
Management of in-stent restenosis
Alfonso F et al
free
168.4

Expert review

10.4244/EIJ-D-21-00690 May 15, 2022
Crush techniques for percutaneous coronary intervention of bifurcation lesions
Moroni F et al
free
149.53

State-of-the-Art

10.4244/EIJ-D-22-00776 Apr 3, 2023
Computed tomographic angiography in coronary artery disease
Serruys PW et al
free
103.48

Expert consensus

10.4244/EIJ-E-22-00018 Dec 4, 2023
Definitions and Standardized Endpoints for Treatment of Coronary Bifurcations
Lunardi M et al
free
X

The Official Journal of EuroPCR and the European Association of Percutaneous Cardiovascular Interventions (EAPCI)

EuroPCR EAPCI
PCR ESC
Impact factor: 6.2
2022 Journal Citation Reports®
Science Edition (Clarivate Analytics, 2023)
Online ISSN 1969-6213 - Print ISSN 1774-024X
© 2005-2024 Europa Group - All rights reserved