IMAGE IN CARDIOLOGY

DOI: 10.4244/EIJV12I9A194

Bilateral, reversible coronary obstruction during aortic valve-in-valve implantation of a repositionable valve system

Fabien Praz1, MD; Aris Moschovitis1, MD; Stefan Stortecky1, MD; Thomas Pilgrim1, MD; Balthasar Eberle2, MD; Stephan Windecker1*, MD

An 80-year-old male patient was admitted to our institution due to new-onset orthopnoea ten years after surgical aortic valve replacement with a 25 mm stentless Freedom Solo bioprosthesis (Sorin Group, Milan, Italy). An echocardiogram revealed severe aortic valve regurgitation owing to a tear in one of the bioprosthetic valve leaflets. In view of the anticipated risk of coronary obstruction associated with this specific type of stentless bioprosthetic valve, TAVI with the use of a fully repositionable 27 mm Lotus™ valve (Boston Scientific, Marlborough, MA, USA) was planned.

Deployment of the Lotus valve was performed aiming at a rather low implantation depth. This led to resolution of the aortic regurgitation. However, immediate haemodynamic compromise (Panel A, arrow point 1) was observed due to the occlusion of both coronary arteries (Panel B). After partial valve retrieval, a contrast injection confirmed restoration of TIMI 3 flow in both coronary arteries resulting in resolution of profound hypotension (Panel A, Panel C, Moving image 1). Repositioning of the Lotus valve at a different depth resulted in a similar haemodynamic compromise due to re-occlusion of the coronary circulation (Moving image 2). As bail-out strategy, a CoreValve® Evolut™ R 29 mm prosthesis (Medtronic, Minneapolis, MN, USA) was successfully deployed (Panel A, arrow point 2). A final aortogram confirmed patency of both coronary arteries (Panel D) with mild paravalvular regurgitation (Panel E, arrow). Furthermore, a sustained rise of the diastolic blood pressure was observed (Panel A).

This report illustrates the value of fully repositionable transcatheter valve prostheses in order to avoid the risk of irreversible coronary obstruction which is increased with some types of surgical stentless bioprostheses.

Conflict of interest statement

T. Pilgrim reports having received research grants to the institution from Biotronik and speaker’s fees from Medtronic and Biotronik. S. Windecker reports receiving research grants to the institution from Abbott, Boston Scientific, Biosensors, Cordis, Medtronic, and St. Jude. The other authors have no conflicts of interest to declare.

Supplementary data

Moving image 1. Restoration of TIMI 3 flow in both coronary arteries after partial valve retrieval.

Moving image 2. Bilateral coronary occlusion despite repositioning of the Lotus valve at a lower depth. Retrieval of the valve.

Supplementary data

To read the full content of this article, please download the PDF.

Restoration of TIMI 3 flow in both coronary arteries after partial valve retrieval

Bilateral coronary occlusion despite repositioning of the Lotus valve at a lower depth. Retrieval of the valve.

Volume 12 Number 9
Oct 20, 2016
Volume 12 Number 9
View full issue


Key metrics

Suggested by Cory

INTERVENTIONAL FLASHLIGHT

10.4244/EIJ-D-17-00850 Jun 14, 2018
Transcatheter aortic valve-in-valve treatment in prior Sorin Freedom Solo valve
Rahman MS et al
free

CLINICAL RESEARCH

10.4244/EIJY15M07_01 Apr 20, 2016
First Lotus aortic valve-in-valve implantation to treat degenerated Mitroflow bioprostheses
Castriota F et al
free

EXPERT REVIEW

10.4244/EIJ-D-18-00667 Sep 7, 2018
Novel strategies in aortic valve-in-valve therapy including bioprosthetic valve fracture and BASILICA
Dvir D et al
free

Image – Interventional flashlight

10.4244/EIJ-D-19-00690 Aug 28, 2020
Transcatheter aortic valve implantation after transcatheter mitral valve implantation
Kuhn E et al
free

State-of-the-Art Review

10.4244/EIJ-D-21-00157 Oct 20, 2021
Transcatheter aortic valve implantation in degenerated surgical aortic valves
Tarantini G et al
free

10.4244/EIJY14M09_04 Mar 20, 2015
How should I treat dislocation of a TAVI SAPIEN prosthesis into the left ventricle?
Tiroch K et al
free
Trending articles
203.25

State-of-the-Art

10.4244/EIJ-D-21-00089 Jun 11, 2021
Intracoronary optical coherence tomography: state of the art and future directions
Ali ZA et al
free
87.2

State-of-the-Art

10.4244/EIJ-D-25-00266 Jan 19, 2026
Lesion stratification with intracoronary imaging
McGarvey M et al
free
47.45

NEW INNOVATION

10.4244/EIJ-D-15-00467 Feb 20, 2018
Design and principle of operation of the HeartMate PHP (percutaneous heart pump)
Van Mieghem NM et al
free
34.8

Original Research

10.4244/EIJ-D-25-01006 Mar 16, 2026
Clinical outcomes and haemodynamic response after blinded stress assessment of moderate aortic stenosis
Eerdekens R et al
22.2

Viewpoint

10.4244/EIJ-D-25-01066 May 4, 2026
Intracoronary imaging guidance for de novo coronary lesion treatment with drug-coated balloons
Amabile N et al
free
20.75

Flashlight

10.4244/EIJ-D-25-01014 Apr 6, 2026
Stent retriever-assisted coronary thrombectomy with continuous aspiration
Liabot Q et al
open access
20

State-of-the-Art

10.4244/EIJ-D-25-00874 Jun 1, 2026
TAVI and coronary interventions: indications, technical considerations, and clinical scenarios
Aquino Bruno H et al
free
19.5

Original Research

10.4244/EIJ-D-26-00032 May 15, 2026
Glucocorticoids to reduce permanent pacemaker implantation after TAVI: the GLUCO-TAVI randomised trial
Fuertes-Kenneally L et al
X

PCR
Impact factor: 9.5
2024 Journal Citation Reports®
Science Edition (Clarivate Analytics, 2025)
Online ISSN 1969-6213 - Print ISSN 1774-024X
© 2005-2026 Europa Group - All rights reserved