DOI: 10.4244/EIJV14I9A176

The future of vascular restoration therapy

Manel Sabaté*, MD, PhD

The promise

The launch of the Absorb™ (Abbott Vascular, Santa Clara, CA, USA), the world’s first drug-eluting poly-L-lactide acid-based (PLLA) bioresorbable vascular scaffold (BVS), in September 2012 was enthusiastically applauded by the interventional cardiology community. The so-called fourth revolution of coronary interventions not only aspired to maintain the favourable acute results of current second-generation drug-eluting stents, but it also sought to improve the long-term clinical outcomes of patients with coronary artery disease. In particular, BVS technology was destined to be able to improve residual angina pectoris and to restore normal coronary vasomotion. Data from the initial ABSORB cohort B registry showed the return of vasomotor function that correlated with reduction in polymeric strut echogenicity1. Physicians and scientists were able to demonstrate that the BVS could be resorbed and, subsequently, the vessel could be released from the perpetual caging of metallic drug-eluting stents. Against this background, the concept of vascular restoration therapy (VRT) was capitalised by this new technology and the device was used to this end in different scenarios (ranging from simple to highly complex) without the development of any dedicated implantation technique protocol or even without any consensus regarding the type and duration of the antiplatelet regimen to be prescribed. Furthermore, patients themselves started to ask for an Absorb to be implanted in their own coronary arteries as a pill for eternality.

The evidence

Several randomised controlled trials were designed to demonstrate the beneficial effect of this new technology scientifically. Most of them were intended to demonstrate non-inferiority at short-term and midterm follow-up but superiority at long-term follow-up as compared to current-generation drug-eluting stents. Early results of the ABSORB II trial apparently supported the concept of angina relief with the use of BVS2. However, this initial finding was soon tarnished by the increased number of both early and very late scaffold thrombosis observed in the Absorb arm3. Device-related factors (e.g., strut thickness), procedural factors (e.g., implantation technique not following predilatation, sizing and postdilatation [PSP] criteria, implantation of BVS in small vessels, device rupture in complex procedures, etc.), type and duration of antiplatelet regimen, and metabolic factors (“unhealthy” resorption process causing very late vessel collapse) were ultimately identified as potential causes of scaffold thrombosis at different stages after BVS implantation4-6. A number of meta-analyses of all randomised trials corroborated the increased risk of myocardial infarction and thrombosis with the use of Absorb BVS7,8. At this point, BVS technology had not been able to demonstrate non-inferiority as compared to standard metallic drug-eluting stents.

Was the promise of vascular restoration therapy accomplished? In the context of randomised data, vasodilatory response to nitroglycerine as a surrogate of endothelium-independent vasomotion was not improved by BVS at three years after its implantation. Therefore, the primary endpoint of the ABSORB II trial was not met9. Besides, device remnants were observed by intracoronary imaging techniques and identified in histopathological samples up to four years of follow-up10. Therefore, in humans and in the clinical arena there probably exists a delay in the resorption process of the PLLA-based device that may preclude normalising vasoreactivity to stimuli.

Like a house of cards, all hopes fell apart when, as of May 2017, the Absorb BVS was restricted to use only in centres already participating in clinical trials and registries and, ultimately, it was removed from the market.

The future

We should learn from past failures. From the disappointing results of first-generation BVS, there had emerged signs of green shoots that should be taken into consideration for future improvements in this field. Let’s focus first on the device, second on the target, and finally on the concept. The next generation of devices should probably be thinner with initial higher radial force. A dedicated implantation technique should be developed and proven in various clinical settings. The resorption process should be safe, healthy and complete and should not take longer than one year. Target lesions amenable to be treated by this device are those which are soft by nature, with small plaque burden. In this regard, patients suffering from acute coronary syndromes, specifically ST-segment elevation myocardial infarction (STEMI), are those who may benefit most from bioresorbable scaffolds. The TROFI II trial performed in STEMI patients was probably the only randomised trial with reassuring results after BVS implantation up to three-year follow-up11. This trial, however, was not powered for clinical events. Recently, a refined PSP-derived implantation algorithm (BVS STEMI STRATEGY-IT) has been validated in the acute phase in this thrombotic context12,13. Whether this technique may confer any benefit during the resorption phase of the device remains to be proven. Finally, the concept of vascular restoration therapy should be revisited. In my view, it comprises at least two parts: first, the complete and safe disappearance of the device as mentioned above and, second, the restoration of vascular reactivity and endothelial function leading to prevention of neoatherosclerosis development and atherosclerosis progression. While the former may be achieved by refinement in engineering work, the latter may require a holistic view. A small device intended to be used to relieve an obstructive lesion is not able to cure a systemic disease. Endothelial function is affected by numerous factors including hypertension, dyslipidaemia, diabetes mellitus, chronic renal disease, sepsis, therapeutic hypothermia, chronic inflammatory disease, obesity and smoking, among others14-19. The use of preventive measures or pharmacological agents such as statins or ACE inhibitors may ameliorate endothelial dysfunction14,16. In 1985, Ludmer et al demonstrated that coronary artery disease was associated with an abnormal vasoconstrictive response to acetylcholine20. Therefore, the sole presence of atherosclerotic plaques in the vasculature may preclude vasomotion from recovery. Consequently, the implantation of a bioresorbable device in, for example, a diffusely diseased diabetic patient may not be enough to restore endothelial function in the long term. At three years, in the TROFI II trial, vasoconstriction to acetylcholine and vasodilatation to nitroglycerine were more often observed in the segment treated with BVS than in that with a permanent metallic everolimus-eluting stent, despite the fact that footprints of the implanted polymeric scaffolds were still clearly visible in all lesions treated with BVS21 (Figure 1). Interestingly, the microcirculation dependent on the infarct-related artery showed a preserved index of resistance, coronary flow reserve, fractional flow reserve and absolute flow. These data support the concept that patients with a small overall plaque burden (e.g., STEMI with only proximal lesions) may be the best candidates for vascular restoration therapy. Neoatherosclerosis development at the scaffolded segment may also be a matter of concern. It has recently been put forward as the second most frequent underlying mechanism of very late scaffold thrombosis (the first being the unfavourable resorption-related process)22. Besides, in patients with stable chronic angina treated with BVS from the ABSORB-EXTEND registry, neoatherosclerosis appeared to develop between one and five years in terms of in-scaffold lipid-laden neointima, calcification, neovascularisation, and thin-cap fibroatheroma23,24. Notably, these findings were not observed in non-scaffolded segments. Whether chronic inflammatory response during the resorption phase induces such a phenomenon remains to be elucidated but should surely be addressed in any new BVS iteration or new bioresorbable technology (e.g., metallic-based BVS).

Figure 1. Long-term vasomotor response after BVS implantation. Three-year follow-up of a proximal left anterior descending artery treated with an Absorb BVS scaffold in the STEMI context. At three years, both the scaffold segment and the distal vessel are reactive to acetylcholine (endothelium-dependent vasoconstriction indicative of endothelial dysfunction) and to nitroglycerine (endothelium-independent vasodilation). ACH: acetylcholine; LD: lumen diameter; QCA: quantitative coronary angiography

In summary, vascular restoration therapy is a complex process in which safe scaffold implantation should be the first step. This has to be followed by a healthy non-inflammatory resorption of the device within a prudent period of 12 months. Concomitantly, a systemic treatment targeted to the underlying chronic disease would be necessary to ameliorate omnipresent endothelial dysfunction demonstrated in patients with coronary risk factors and atherosclerosis.

Conflict of interest statement

M. Sabaté received consulting fees from Abbott Vascular.


References

Volume 14 Number 9
Oct 12, 2018
Volume 14 Number 9
View full issue


Key metrics

Suggested by Cory

10.4244/EIJV13I7A115 Sep 20, 2017
Bioresorbable scaffolds: reflections after a setback - losing a battle does not mean losing the war!
Colombo A and Azzalini L
free

10.4244/EIJV16I8A116 Oct 23, 2020
BVS déjà vu: the storm before the calm
Kereiakes D
free

10.4244/EIJV15I16A254 Mar 20, 2020
Bioresorbable scaffolds and STEMI: an ideal setting, but still vacant
Spaulding C
free

10.4244/EIJV13I13A242 Jan 19, 2018
Polymeric bioresorbable coronary scaffolds: the hype is over, but the dream lives on
Pyxaras S and Wijns W
free

10.4244/EIJV15I1A5 May 20, 2019
Should we stop using bioresorbable scaffolds in coronary revascularisation?
Mangieri A and Colombo A
free

10.4244/EIJV16I2A16 Jun 12, 2020
Bioresorbable scaffolds: did we jump the gun?
Waksman R and Forrestal B
free

EXPERT REVIEW

10.4244/EIJ-D-17-00499 Aug 25, 2017
State of the art: the inception, advent and future of fully bioresorbable scaffolds
Katagiri Y et al
free
Trending articles
152.9

Clinical research

10.4244/EIJ-D-20-01125 Oct 20, 2021
An upfront combined strategy for endovascular haemostasis in transfemoral transcatheter aortic valve implantation
Costa G et al
free
47.8

NEW INNOVATION

10.4244/EIJ-D-15-00467 Feb 20, 2018
Design and principle of operation of the HeartMate PHP (percutaneous heart pump)
Van Mieghem NM et al
free
39.1

Clinical research

10.4244/EIJ-D-22-00558 Feb 6, 2023
Permanent pacemaker implantation and left bundle branch block with self-expanding valves – a SCOPE 2 subanalysis
Pellegrini C et al
free
38.95

State-of-the-Art

10.4244/EIJ-D-23-00912 Oct 7, 2024
Optical coherence tomography to guide percutaneous coronary intervention
Almajid F et al
free
X

The Official Journal of EuroPCR and the European Association of Percutaneous Cardiovascular Interventions (EAPCI)

EuroPCR EAPCI
PCR ESC
Impact factor: 7.6
2023 Journal Citation Reports®
Science Edition (Clarivate Analytics, 2024)
Online ISSN 1969-6213 - Print ISSN 1774-024X
© 2005-2024 Europa Group - All rights reserved