Re: “Transcatheter treatment for refractory angina with the coronary sinus Reducer” by Maayan Konigstein et al

Yoav Paz1*, MD, MHA; Amihay Shinfeld2, MD

We read with interest the paper by Konigstein et al1, and wish to commend the authors for their significant contribution.

In the discussion section the authors suggest that the anti-ischaemic effect of the Coronary Sinus Reducer™ Stent (CSRS) (Neovasc Medical, Inc., Or Yehuda, Israel) is based on the hypothesis described by Camici et al2. We would like to point out that, while Camici et al published this hypothesis in the New England Journal of Medicine in 2007, the initial experiments with the CSRS were performed a decade before. Even Banai et al had used the CSRS before Camici et al published their hypothesis, when they implemented the first-in-man study using this stent3.

Konigstein et al1 support their rationalisation of the anti-ischaemic effect of the CSRS by quoting Ido et al4, a study which examined the effects of coronary sinus occlusion on coronary collateral blood flow and on the distribution of regional myocardial blood flow in dogs. We wish to emphasise that there is a substantial difference between human and dog morphology of veins that drain into the coronary sinus5.

In the mid 1990s we tested a new strategy supporting the ischaemic myocardium. This strategy included: 1) catheterisation of the coronary venous system rather than catheterisation of the coronary arteries, and 2) reduction of the coronary sinus (CS) effective cross area, as opposed to the expansion of a narrowed coronary artery. At that time, the main concept behind this strategy was to rebuild retrograde coronary pressure that would be attenuated by the atherosclerotic disease. In order to test this strategy, we designed and manufactured the first CSRS.

In a preliminary non-ischaemic pig model we succeeded in increasing the mean CS pressure from 7.0 to 24.6 mmHg (p=0.011) after CSRS deployment. Further studies in a non-ischaemic pig model were devoted to macroscopic and histological investigations of the treated hearts, in particular investigating whether any structural or histological damage, such as an infarct, had occurred after CSRS implantation.

While looking for such damage, these studies revealed that eight to 12 weeks of coronary sinus narrowing produced macroscopic epicardial and intramyocardial new blood vessels – neovascularisation. Histopathological analysis described these findings as follows: significant proliferation of small to medium-sized vessels, containing smooth muscle representing coronary collaterals. This was evident in almost all specimens, representing various myocardial anatomical areas, including specimens from the anterior and mid-posterior wall. According to these unpredicted neovascularisation findings after CSRS implantation in a non-ischaemic pig model, we created the name “Neovasc” for this novel CSRS device.

Backwards pressure elevation in the venules and capillaries starts immediately after CSRS deployment and a few weeks before clinical improvement. However, neovascularisation created by CSRS a few weeks following deployment seems to be the main contributing factor for the anti-ischaemic effect, and correlates well with the report by Konigstein et al, in which clinical improvement was, in most cases, reported to have started a few weeks following the procedure.

Conflict of interest statement

Y. Paz and A. Shinfeld are the inventors of the Neovasc Coronary Sinus Reducer Stent. They currently have no commercial or other association with the company whose product is the subject of this letter, or with a company that manufactures comparable products.

Drs Paz and Shinfeld argue that the anti-ischaemic mechanism of action of the coronary sinus Reducer is by promoting neovascularisation. The scientific basis of their theory is a series of preclinical experiments which they say they performed in the mid 1990s.

As far as I comprehend the field of myocardial neovascularisation and myocardial angiogenesis6-10, pressure gradients that are built between blood vessels within the myocardium (as in the case of the Reducer) are responsible for the opening of preformed collaterals and not for the creation of new blood vessels (neovascularisation).

Since the Reducer is implanted in patients with chronic refractory angina, chronic myocardial ischaemia is present long before the Reducer is implanted. The chronic ischaemia activates angiogenic processes, generating neovessels, mainly capillaries, small arterioles, and venules11. Therefore, the ischaemia-induced myocardial neovascularisation process exerts itself long before the Reducer is implanted.

After implantation, the Reducer does not cause any pressure gradient until four to six weeks after implantation, when the metal mesh is covered with tissue. When pressure gradient is built, the changes in vascular tone provide a quick functional adaptation to accommodate rapid changes in metabolic demand by the development of existing collateral vessels, and by rapid changes in vessel diameter and drop in resistance to flow in the subendocardial myocardium11.

As we state in our manuscript12, the beneficial effects of the Reducer revolve around the consequence of ischaemia-induced impaired contractility and elevated left ventricular end-diastolic pressure (LVEDP). Elevated LVEDP exerts an external pressure on the subendocardial capillaries and arterioles which increases the resistance to flow in the subendocardium, and worsens subendocardial ischaemia. Heightened CS pressure causes backwards pressure elevation in the venules and capillaries which will result in a dilatation of the capillaries’ and arterioles’ diameter and a reduction in resistance to flow to the subendocardium. Consequently, enhancement of blood flow to the ischaemic subendocardial layers will occur, with improved contractility and reduced LVEDP which will further reduce subendocardial resistance.

Nevertheless, if Drs Paz and Shinfeld have proven otherwise, they should share the results of their scientific work with the readers of EuroIntervention, and they should cite their peer-reviewed scientific publications on the subject, so we can all read and learn.

Conflict of interest statement

S. Banai is the Medical Director of Neovasc Inc, Vancouver, Canada. M. Schwartz is the Director of Clinical Affairs and an employee of Neovasc Inc. The other authors have no conflicts of interest to declare.


References

Volume 11 Number 6
Oct 20, 2015
Volume 11 Number 6
View full issue


Key metrics

Suggested by Cory

10.4244/EIJV17I7A94 Sep 20, 2021
Coronary sinus reducer therapy for refractory angina: is it ready for prime time?
Foley M and Al-Lamee R
free

10.4244/EIJV9I10A196 Feb 20, 2014
Transcatheter treatment for refractory angina with the coronary sinus Reducer
Konigstein M et al
free

SHORT REPORT

10.4244/EIJ-D-17-00626 Feb 20, 2018
Coronary sinus reducer non-responders: insights and perspectives
Baldetti L et al
free

Expert review

10.4244/EIJ-D-18-01180 Feb 20, 2020
Technical aspects in coronary sinus Reducer implantation
Giannini F et al
free

10.4244/EIJV11I1A3 May 19, 2015
Transcoronary sinus catheter interventions: back in the repertoire?
Mohl W et al
free

10.4244/EIJ-D-20-00676L Mar 19, 2021
Letter: Pitfalls in aerobic capacity estimation of a chronic angina population
Tzanis G and Giannini F
free

Clinical research

10.4244/EIJ-D-19-00766 Apr 2, 2021
Effects of coronary sinus Reducer implantation on oxygen kinetics in patients with refractory angina
Zivelonghi C et al
free
Trending articles
225.68

State-of-the-Art Review

10.4244/EIJ-D-21-00426 Dec 3, 2021
Myocardial infarction with non-obstructive coronary artery disease
Lindahl B et al
free
105.78

Expert consensus

10.4244/EIJ-E-22-00018 Dec 4, 2023
Definitions and Standardized Endpoints for Treatment of Coronary Bifurcations
Lunardi M et al
free
77.85

State-of-the-Art

10.4244/EIJ-D-23-00840 Sep 2, 2024
Aortic regurgitation: from mechanisms to management
Baumbach A et al
free
68.7

Clinical research

10.4244/EIJ-D-21-00545 Sep 20, 2022
Coronary lithotripsy for the treatment of underexpanded stents: the international; multicentre CRUNCH registry
Tovar Forero M et al
free
47.8

NEW INNOVATION

10.4244/EIJ-D-15-00467 Feb 20, 2018
Design and principle of operation of the HeartMate PHP (percutaneous heart pump)
Van Mieghem NM et al
free
45.3

Clinical research

10.4244/EIJ-D-18-01126 Aug 29, 2019
New-generation mechanical circulatory support during high-risk PCI: a cross-sectional analysis
Ameloot K et al
free
X

The Official Journal of EuroPCR and the European Association of Percutaneous Cardiovascular Interventions (EAPCI)

EuroPCR EAPCI
PCR ESC
Impact factor: 7.6
2023 Journal Citation Reports®
Science Edition (Clarivate Analytics, 2024)
Online ISSN 1969-6213 - Print ISSN 1774-024X
© 2005-2024 Europa Group - All rights reserved