Abstract
BACKGROUND: Left atrial appendage occlusion (LAAO) is performed increasingly, but long-term follow-up imaging data are lacking.
AIMS: The aim of this study was to evaluate the safety and durability of the Amplatzer Amulet device >4 years after LAAO.
METHODS: This was a prospective observational cohort study including 52 patients implanted with the Amplatzer Amulet device at Aarhus University Hospital, Denmark. A >4-year follow-up cardiac computed tomography (CT) scan after LAAO was performed and compared with the results from the 2-month and 12-month scans. The primary outcome was left atrial appendage (LAA) sealing based on distal LAA contrast patency and peridevice leakage (PDL), stratified into complete occlusion (grade 0 [G0]) and grade 1-3 leakage (G1-3), respectively. Secondary outcomes were low- and high-grade hypoattenuated thickening (HAT), device-related thrombosis (DRT) and device durability.
RESULTS: The median (interquartile range [IQR]) follow-up time from LAAO to the latest CT scan was 5.8 years (4.5; 6.3). At 2-month (n=52), 12-month (n=27) and >4-year CT follow-ups (n=52), rates of both complete occlusion (33%, 37%, 35%) and G2 leaks (52%, 52%, 48%) remained stable. Rates of...