DOI: 10.4244/EIJV9I2A47

Percutaneous treatment of paravalvular leaks after transcatheter aortic valve implantation with the CoreValve self-expanding bioprosthesis

Francesco Saia*, MD, PhD; Cinzia Marrozzini, MD; Carolina Moretti, MD; Barbara Bordoni, MD; Cristina Ciuca, MD; Antonio Marzocchi, MD

Mild-to-moderate aortic regurgitation (AR) after transcatheter aortic valve implantation (TAVI) is common and often clinically silent. Higher-degree AR may occur, with significant haemodynamic and clinical consequences. Treatment of this condition is challenging and there are no evidence-based therapeutic options. This image illustrates the successful percutaneous management of clinically relevant paravalvular leaks (PVL) after implantation of the CoreValve bioprosthesis (Medtronic, Minneapolis, MN, USA) with the use of vascular plug devices.

Conflict of interest statement

The authors report no financial relationships or conflicts of interest regarding the content herein.

Figure 1. Examples of successful percutaneous closure of severe paravalvular leak (PVL) after transcatheter aortic valve implantation (TAVI) using two different Amplatzer™ vascular plug (AVP) devices (St. Jude Medical, St. Paul, MN, USA). A) The Amplatzer™ Vascular Plug 4. B) The frame of the bioprosthesis was crossed using an Amplatz Left diagnostic catheter, with the help of a hydrophilic wire. C) Catheter exchange into the left ventricle was performed with a 260 cm hydrophilic wire, and a 6 mm device was advanced through a 6 Fr multipurpose PCI catheter. The distal part of the AVP4 device is visible (arrow). D) Position of the device across the valve frame after its release. E) Closure of a PVL with an AVP2 device (box). The device was released with the proximal disc within the aortic side of the frame and the other two lobes sitting below the severely calcified native valve annulus. F and G) Drawings superimposed on the angiography to highlight the mechanism and the site of aortic regurgitation (F) and the position of the vascular plug (G). The lines show the presence of an angle between the aortic annulus plane (dotted line) and the bioprosthetic valve plane (solid line).

Volume 9 Number 2
Jun 28, 2013
Volume 9 Number 2
View full issue


Key metrics

Trending articles
310.93

State-of-the-Art Review

10.4244/EIJ-D-21-00695 Nov 19, 2021
Transcatheter treatment for tricuspid valve disease
Praz F et al
free
92.2

State-of-the-Art Review

10.4244/EIJ-D-20-01296 Aug 27, 2021
Management of cardiogenic shock
Thiele H et al
free
76.25

State-of-the-Art

10.4244/EIJ-D-23-00840 Sep 2, 2024
Aortic regurgitation: from mechanisms to management
Baumbach A et al
free
56.65

Clinical research

10.4244/EIJ-D-20-01155 Oct 20, 2021
A deep learning algorithm for detecting acute myocardial infarction
Liu W et al
free
35

Original Research

10.4244/EIJ-D-25-00331 May 21, 2025
One-month dual antiplatelet therapy followed by prasugrel monotherapy at a reduced dose: the 4D-ACS randomised trial
Jang Y et al
open access
26

Expert Review

10.4244/EIJ-D-24-00535 May 5, 2025
Catheter-based techniques for pulmonary embolism treatment
Costa F et al
X

The Official Journal of EuroPCR and the European Association of Percutaneous Cardiovascular Interventions (EAPCI)

EuroPCR EAPCI
PCR ESC
Impact factor: 9.5
2024 Journal Citation Reports®
Science Edition (Clarivate Analytics, 2025)
Online ISSN 1969-6213 - Print ISSN 1774-024X
© 2005-2025 Europa Group - All rights reserved