Transcatheter valve implantation for patients with aortic stenosis: a position statement from the European Association of Cardio-Thoracic Surgery (EACTS) and the European Society of Cardiology (ESC), in collaboration with the European Association of Percutaneous Cardiovascular Interventions (EAPCI)

Alec Vahanian1*, Ottavio Alfieri2*, Nawwar Al-Attar1, Manuel Antunes3, Jeroen Bax4, Bertrand Cormier5, Alain Cribier6, Peter De Jaegere7, Gerard Fournial8, Arie Pieter Kappetein7, Jan Kovac9, Susanne Ludgate10, Francesco Maisano2, Neil Moat11, Friedrich Mohr12, Patrick Nataf1, Luc Pierard13, José Luis Pomar14, Joachim Schofer15, Pilar Tornos16, Murat Tuzcu17, Ben van

Aims: To critically review the available transcatheter aortic valve implantation techniques and their results, as well as propose recommendations for their use and development.

Methods and results: A committee of experts including European Association of Cardio-Thoracic Surgery and European Society of Cardiology representatives met to reach a consensus based on the analysis of the available data obtained with transcatheter aortic valve implantation and their own experience. The evidence suggests that this technique is feasible and provides haemodynamic and clinical improvement for up to 2 years in patients with severe symptomatic aortic stenosis at high risk or with contraindications for surgery. Questions remain mainly concerning safety and long-term durability, which have to be assessed. Surgeons and cardiologists working as a team should select candidates, perform the procedure, and assess the results. Today, the use of this technique should be restricted to high-risk patients or those with contraindications for surgery. However, this may be extended to lower risk patients if the initial promise holds to be true after careful evaluation.

Conclusion: Transcatheter aortic valve implantation is a promising technique, which may offer an alternative to conventional surgery for high-risk patients with aortic stenosis. Today, careful evaluation is needed to avoid the risk of uncontrolled diffusion.

Preamble

Valve disease is an important public health problem, as it carries a poor prognosis and its prevalence is strongly linked to the phenomenon of population ageing.1 The most frequent native valve disease in Europe is currently aortic stenosis (AS), which is most often seen in elderly patients with comorbidities.2 Valve replacement is the definitive therapy for patients with severe AS who have symptoms or objective consequences such as left ventricular (LV) dysfunction.3,4 Operative mortality is quite low, even in elderly patients when properly selected, and long-term results have been shown to be satisfactory.5,6 However, the risk of surgery may be higher in elderly patients with significant comorbidities. In addition, several registries show that referring physicians often do not propose surgery, as was the case in the Euro Heart Survey with 33% of patients with severe valve disease and severe symptoms not being considered for surgery.7 Thus, despite the good results of valve surgery, there may well be a role for less invasive alternatives.

Balloon aortic valvuloplasty (BAV) is now rarely used, mainly due to its limited long-term efficacy.2

Six years after the first in man,8 transcatheter aortic valve implantation (TAVI) for the treatment of AS currently represents a dynamic field of research and development: two devices have been CE-marked and are being commercialized.

At this point in time, it is important for professional societies to critically review the available TAVI techniques and their results, as well as propose recommendations for their use, development in clinical practice, and further research.9,10 On the 18-19th of November 2007, a committee of experts, including EACTS and ESC representatives, met. The consensus reached is summarized in the current document, which was approved first by the committee members and subsequently by the board of both professional societies involved.

The Committee acknowledges that the conclusions in this document rely on limited data reported mostly in oral communications and few in peer-reviewed journals and are temporally limited by the very nature of the document.

Current techniques and results

Two devices are under clinical investigation for TAVI. One device is the Edwards-Sapien valve (Edwards Lifesciences Inc., CA, USA), which consists of three pericardial leaflets, initially equine and currently bovine, mounted within a tubular, slotted, stainless steel, balloon-expandable stent. It is currently available in 23 and 26 mm sizes, necessitating, for the transfemoral approach, respectively, 22 and 24 Fr introducer sheaths, and for the transapical approach, this measurement was 33 Fr and is now 26 Fr. The other device is the CoreValve Revalving System (CRS TM, CoreValve Inc., Irvine, CA, USA), which has three pericardial leaflets, initially bovine and currently porcine, mounted in a self-expanding, nitinol frame. It is available in 26 and 29 mm sizes, which go through an 18 Fr introducer for transfemoral use, or very recently 21 Fr for the transapical route. (NB: the measurements of prosthesis size quoted by the manufacturers do not correspond exactly to those of surgically implanted prostheses).

Techniques of implantation

TAVI is currently carried out using two different approaches (retrograde transfemoral and anterograde transapical), which share the same main principles.

– Most teams perform the procedure under general anaesthesia, although sedation and analgesia may suffice for the transfemoral approach.

– Peri-procedural transoesophageal echocardiography (TEE) monitoring is desirable to correctly position the valve as well as to detect complications.

– After crossing the aortic valve, BAV is performed to pre-dilate the native valve and serve as a rehearsal for TAVI. Simultaneous rapid pacing decreases cardiac output, stabilizing the balloon during inflation. Normal blood pressure must be completely recovered between sequences of rapid pacing.

– The following imaging methods can be used to position the prosthesis at the aortic valve:

- Fluoroscopy to assess the level of valve calcification.

- Aortography, using different views, performed at the beginning of the procedure and eventually repeated with the unde-ployed prosthesis in place, to determine the position of the valve and the plane of alignment of the aortic cusps.

- Echocardiography: TEE is helpful, in particular, in cases with moderate calcification. The additional value of three-dimensional real-time TEE is currently being evaluated. According to the limited current experience with intracardiac echography, it does not seem to add to TEE in this setting.

– When positioning is considered correct, the prosthesis is released. Rapid pacing is used at this stage in balloon expandable but not in self-expanding devices.

– Immediately after TAVI, aortography and, whenever available, TEE or, in the absence of TEE, eventually Transthoracic echo- cardiogram (TTE) are performed to assess the location and degree of aortic regurgitation and the patency of the coronary arteries and to rule out complications such as haemopericardium, and aortic dissection. The haemodynamic results are assessed using pressure recordings and/or echocardiography.

– After the procedure, the patients should stay in intensive care for at least 24 h and be closely monitored for several days especially as regards haemodynamics, vascular access, rhythm disturbances (especially late atrioventricular block), and renal function.

The following are the specific issues related to the different approaches.

In the transfemoral approach, close attention should be paid to the vascular access.

The common femoral artery can be either prepared surgically or approached percutaneously. Echo-guided femoral access could be useful. Manipulation of the introductory sheaths should be careful and fluoroscopically guided. Depending on the size of the device, closure of the vascular access can be effected surgically or using a percutaneous closure device.11

For the transapical approach, femoral access and cardiopulmon-ary bypass should be on standby in patients in whom surgical conversion is an option in case of complications. The technique requires an antero-lateral mini-thoracotomy, pericardiotomy, identification of the apex, and then puncture of the left ventricle using a needle through purse-string sutures. Subsequently, an introductory sheath is positioned in the LV, and the prosthesis is implanted using the anterograde route.

Results

Since the first-in-man TAVI by Alain Cribier in 2002, well over 1000 high-risk patients with severe symptomatic AS have been treated using TAVI (as of January 2008).

TRANSFEMORAL APPROACH

Over 400 cases have been performed using the balloon-expandable and another 500 using the self-expandable prosthesis (Company sources January 2008). Reports originate from a limited number of centres worldwide.12-17 The patients treated were mostly >80 years old, at high risk (e.g. Logistic EuroScore >20% in most cases), or with contraindications for surgery.

The overall results can be summarized as follows:

Procedural success is closely linked to experience and is ~90% in experienced centres. A learning curve can also be observed resulting in better patient selection and outcomes. Valve function is good with a final valve area ranging from 1.5 to 1.8 cm2.

Mortality at 30 days ranges from 5 to 18%. Acute myocardial infarction occurs in 2-11%. Coronary obstruction is rare (<1%). Mild-to-moderate aortic regurgitation, mostly paravalvular, is observed in ~50% of cases. However, the availability of larger prostheses and their more careful matching with the size of the aortic annulus led to the decrease in the incidence of severe aortic regurgitation to ~5%. Prosthesis embolization is rare, ~1%.

Vascular complications, with an incidence ranging from 10 to 15%, remain a significant cause of mortality and morbidity. Stroke ranges from 3 to 9%. Finally, atrioventricular block occurs in 4-8%, necessitating pacemaker implantation in up to 24% with self-expandable devices.

Long-term results up to 2 years (though only 1 year in most studies) are reported in a limited number of patients. They show a survival rate of 70-80% with a significant improvement in clinical condition in most cases. The majority of late deaths are due to comorbidities.

Serial echocardiographic studies have consistently shown good prosthetic valve function with no structural deterioration of valve tissue.

TRANSAPICAL APPROACH

The total experience with transapical aortic valve implantation comprises over 300 patients, also at high risk for conventional surgery, even more so because of concomitant peripheral arterial disease in most cases. Experience currently reported only relates to the balloon-expandable prosthesis,18-21 although, the first in man has been recently performed with the self-expandable device.

The implantation success rate of the transapical procedure is ~90%. Over 70% of cases are done off pump, the figure being 90% in experienced centres, and the rate of peri-operative conversion is 9-12%. Mortality rates range from 9 to 18%. The incidence-of stroke is 0-6%. The quality of the results seems closely related to experience as well as the availability of high-quality imaging in the operating theatre.19

There are currently no direct comparative studies available for the two approaches.

Perspectives

Progress in delivery systems and valve manufacturing could lead to lower profile, repositionable, retrievable, and more durable devices, as well as a wider range of prosthetic valve dimensions. Furthermore, improved imaging, such as online three-dimensional reconstruction and stereotaxis, could facilitate valve placement.

Pending questions

In the light of the current experience, bearing in mind the previously mentioned, inherent limitations of any conclusions, TAVI using both balloon- and self-expandable devices can be said to be feasible. Short- and mid-term haemodynamic results are good up to 2 years. However, the technique remains challenging, in particular, as regards vascular access, device sizing, and positioning. The major concerns as regards to safety are as follows:

– Vascular complications with the transfemoral approach, which should decrease with smaller devices.

– Stroke rate, in particular, when using the transfemoral approach.

– Long-term consequences of paravalvular leaks, even if mild-to-moderate regurgitation is considered not to have significant clinical consequences in the short-term.

– Atrioventricular block, the incidence, timing, and predictors of which have to be identified more precisely.

Mid (short)-term clinical outcome is encouraging, however, long-term durability of these bioprostheses and the incidence of endocarditis or thrombo-embolic events remain key questions, especially if considering lowering the threshold for indication.

The feasibility of subsequent aortic valve intervention is not known.

Recommendations for the development of transcatheter aortic valve implantation

Patient selection

Selection of candidates for TAVI, especially risk assessment, should involve multi-disciplinary consultation between cardiologists, surgeons, imaging specialists, anaesthesiologists, and possibly other specialists if necessary.

TAVI is indicated in patients with calcified pure or predominant AS. It is unlikely that it will be used in patients with pure aortic regurgitation.

Degeneration of an implanted aortic bioprosthesis (valve-in-valve implantation) is an attractive potential indication because of the high risk of re-operation in elderly patients. However, the current experience is too limited to make any recommendations.

The following are the four steps of patient selection:

– confirmation the severity of AS;

– evaluation of symptoms;

– analysis of the risk of surgery and evaluation of life expectancy and quality of life;

– assessment of the feasibility and exclusion of contraindications for TAVI.

CONFIRMATION OF THE SEVERITY OF AORTIC STENOSIS

TAVI should be performed only in severe AS. Echocardiography is the preferred tool to assess the severity of AS according to a combination of measurements of valve area and flow-dependent indices. Low-dose dobutamine echocardiography is useful to differentiate between severe and the rare ‘pseudo severe’ AS in patients with low LV ejection fraction and low gradient.3,4

EVALUATION OF SYMPTOMS

At the present stage, TAVI should only be proposed in patients with severe symptoms that can definitely be attributed to valve disease because of pending questions on safety and valve durability.

ANALYSIS OF THE RISKS OF SURGERY, AND EVALUATION OF LIFE EXPECTANCY AND QUALITY OF LIFE

Decision-making is particularly complex in these elderly patients who represent a heterogeneous population and require balanced and individualized analysis.

The evaluation of the risk of surgery is based on the assessment of cardiac and extra cardiac factors.22 Risk scores, such as the Euro-Score,23 the STS Predicted Risk of Mortality score,24 or the Ambler score,25 are of interest. However, they all share similar limitations: predictive ability is reduced in these high-risk patients who represent only a small proportion of the population from which the scores were elaborated; and high-risk patients form a particularly heterogeneous group in which it is difficult to capture all the comor-bidities. The value of individual scores in this high-risk population has yet to be specifically established.26 The predictive value of these scores for morbidity and long-term results is also unknown.

TAVI should currently be restricted to patients at high-risk or with contraindications for surgery. It is premature to consider using it in patients who are good surgical candidates.

For the Committee, the key element to establish whether patients are at high risk for surgery is clinical judgement, which should be used in association with a more quantitative assessment, based on the combination of several scores (for example expected mortality >20% with the Logistic EuroScore and >10% with STS score). This approach allows the team to take into account risk factors that are not covered in scores but often seen in practice such as chest radiation, previous aorto-coronary bypass with patent grafts, porcelain aorta, liver cirrhosis, etc. Surgical risk estimation should also bear in mind results in the given institution.27

At this stage, TAVI is not recommended for patients who simply refuse surgery on the basis of personal preference.

Life expectancy is most significantly influenced by comorbidities, which should be carefully looked for. In addition to clinical evaluation, semi-quantitative scoring systems, such as those used in geriatrics, may be helpful.28 Today, TAVI is seldom considered in patients <70, however, age alone is not sufficient for its use instead of surgery. TAVI should not be performed in patients whose life expectancy is <1 year, who should be managed conservatively.

ASSESSMENT OF FEASIBILITY AND EXCLUSION OF CONTRAINDICATIONS OF TRANSCATHETER AORTIC VALVE IMPLANTATION

The following steps should be taken to assess the feasibility of TAVI:

Assessment of the coronary anatomy

Coronary angiography should be performed to this end. If associated coronary artery disease requires revascularization, whether to proceed surgically, percutaneously, or in a hybrid manner, as well as the chronology of interventions, should be the subject of individualized discussion based on the patient’s clinical condition and anatomy. TAVI is probably not recommended in patients with severe proximal coronary stenoses not amenable to percutaneous coronary interventions.

The position of the coronary arteries relative to the aortic cusps can be assessed using aortography or multislice computed tomography.29

Measurement of the aortic annulus

Correct sizing of the valve is critical to minimize the potential for paravalvular leakage and to avoid prosthesis migration after placement; however, a gold standard method of measurement has yet to be established. TEE has been found to show larger values than transthoracic echocardiography, thus, it should be performed if borderline values lead to doubt the feasibility of the procedure.30 Multislice computed tomography29 or magnetic resonance imaging could also be used for this purpose. Finally, aortography measurements performed during BAV are also useful.

Echocardiography is the preferred tool for the assessment of the morphology of the LV outflow tract and, before implanting self-expandable devices, the dimensions of the aortic root.

Evaluation of size, tortuosity, and calcification of peripheral arteries

Angiography is the reference; however, multislice computed tomography can also be used. Finally, magnetic resonance imaging is helpful in patients with renal insufficiency.

Some contraindications are general, whereas others are approach- or device-specific.

General contraindications for TAVI are as follows:

– Aortic annulus <18 or >25 mm for balloon-expandable and <20 or >27 for self-expandable devices

– Bicuspid valves because of the risk of incomplete deployment of the prosthesis31

– Presence of asymmetric heavy valvular calcification, which may compress the coronary arteries during TAVI.15 The bulk and distribution of calcification in the valve may be assessed by fluoroscopy and multislice computed tomography. Finally, the risk of coronary compression can be anticipated during BAV.

– Aortic root dimension >45 mm at the aorto-tubular junction for self-expandable prostheses.

– Presence of apical LV thrombus.

The specific indications for transfemoral and transapical approaches are not fully established and should be discussed according to patient condition and local expertise.

Contraindications of the transfemoral approach are as follows:

– Iliac arteries: severe calcification, tortuosity, small diameter (<6 to 9 mm according to the device used), previous aorto-femoral bypass.

– Aorta: severe angulation, severe atheroma of the arch, coarctation, aneurysm of the abdominal aorta with protruding mural thrombus.

– Presence of bulky atherosclerosis of the ascending aorta and arch detected by TEE

– Transverse ascending aorta (balloon-expandable device).

Contraindications for the transapical approach are as follows:

– previous surgery of the LV using a patch, such as the Dor procedure;

– calcified pericardium;

– severe respiratory insufficiency;

– non-reachable LV apex.

Performance of transcatheter aortic valve implantation

The performance of TAVI, even more so ab initio, should be restricted to a limited number of high-volume centres, which have both cardiology and cardiac surgery departments, with expertise in structural heart disease intervention and high-risk valvular surgery.

The procedure requires the close cooperation of a team of specialists in valve disease, including clinical cardiologists, echocar-diographists, interventional cardiologists, cardiac surgeons, and anaesthesiologists.

Interventional cardiologists should be experienced in catheter-based valvular interventions, and peripheral access using large devices.

Cardiac surgeons should be experienced in valve surgery and the management of complex cases.

The final organization may vary from centre to centre and according to the type of procedure.

A multidisciplinary team approach and cross-fertilization will be fundamental in the development of these procedures.

Because of the severe clinical condition of these patients, the presence of anaesthesiologists with specific expertise in cardiology is mandatory for peri- and post-operative care in collaboration with cardiologists and cardiovascular surgeons.

A close collaboration with surgeons skilled in vascular access repair and endo-vascular procedures is recommended for transfemoral aortic valve implantation.

The aim of TAVI training is to acquire basic, then advanced, and finally device-specific skills. It will concentrate on knowledge of valve disease (clinical assessment, catheterization techniques, imaging), working in a sterile environment, access management, understanding the equipment, and anticipation and treatment of complications. Training will be through didactic sessions, bench and/or computer simulators, animal laboratories, hands-on training, participation in established workshops, and proctoring during the first cases.

The field of teaching and certification needs to be significantly developed. A new subspecialization may be necessary with a common training pathway for cardiologists and surgeons wishing to practice TAVI.

At this early stage, data are insufficient to define a minimum number of procedures required for competency. Furthermore, individual experience must be continued in order to conserve skills.

As regards the logistics, it is agreed that the operating room and the catheterization laboratory should be as close as possible to guarantee optimum patient safety. The optimal environment for TAVI should be spacious and sterile, and feature high-quality imaging equipment, and haemodynamic monitoring and recording capabilities. However, where to find such an environment is still under debate, as both current surgical and interventional suites are suboptimal. Catheterization laboratories have good quality imaging but may not be sterile and may induce delay in rescue surgery. The opposite is often the case for operating rooms.

A hybrid suite is ideal as it fulfils the role of both an operating room and a catheterization laboratory. However, most institutions do not have a hybrid room available, thus, the organizational and financial implications of acquiring one, which is highly desirable, cannot be overlooked.

It is the role of professional organizations such as the EACTS and the ESC and their national equivalents to set the standards for performance of the procedure and training and accreditation on an individual and institutional level.

Evaluation of the results

Evaluation of the results of TAVI will present a significant challenge because of the unique characteristics of this patient population. Thus, comparison with current outcome standards may not be appropriate. Close collaboration between investigators, regulators, and the industry should be assured early on.

The evaluation of new devices or techniques should be performed in accordance with the stipulations of the regulating authorities.32,33 In Europe, TAVI devices are classified in the highest risk group (class 3) as regards the requirements for the putting of new devices on the market.32 Evaluation should adhere to the following pattern: bench testing, in vitro, and animal implantation studies (including bio-compatibility studies), then, only when sufficient data are available, clinical investigation.

Clinical investigation should begin with the proof of concept feasibility studies, followed by prospective clinical investigations to determine safety and performance.34

Randomized trials are the most rigorous design to evaluate safety and efficacy in relation to other treatment modalities. However, this is difficult with TAVI in relation to the most high-risk patients, many of whom are currently not operated on mostly due to non-referral.7 Furthermore, the contemporary natural history of unoperated patients is incompletely known.

The Committee believes that randomized trials are highly desirable once greater experience has been acquired, and only small modifications in the technology used are to be expected. The necessary clear definition of entry criteria for these trials may be hindered by the inadequacy of risk scores in these high-risk patients (see the section ‘Analysing the risks of surgery’ for more details).

In high-risk operable patients, TAVI should be compared with valve replacement, in particular with the perspective of patients who are in better clinical condition in the future. In these patients, efficacy could be tested as non-inferiority vs. surgery, avoiding a too large delta, whereas safety can be tested as superiority to surgery. In non-operable patients, TAVI can be compared with the best available medical treatment using the superiority design for efficacy; however, this is more debatable and probably more difficult to run. A randomized trial, PARTNER US, is currently ongoing where the primary endpoint is all-cause mortality at 1 year. In the future, the extension of TAVI to a lower risk patient population should occur only in the setting of a randomized trial.

As well as enrolment in randomized controlled trials, data should be accumulated in registries. Such registries, which may be regional, national, or international, are mandatory in the evaluation of TAVI before its release in the general medical community. Completeness of the data is essential here. The duration of follow-up to assess safety should be at least 1 year in the pre-market phase and ideally 5 years for post-marketing surveillance. Retrieval studies should also be carried out if appropriate.

Irrespective of the study design, safety and efficacy endpoints must be carefully defined.

Analysis will include the standard major adverse cardiac events: mortality (30 day, in hospital, 1 year), stroke, myocardial infarction, re-hospitalization, reoperation, arrhythmias, and conduction disturbances. All adverse events, anticipated and non-anticipated, should also be reported to the health authority. Quality of life is paramount, in particular, in these populations; thus, improvement should be recorded using quality of life metrics, which should be combined with cost effectiveness parameters. Endpoints assessing valve function should follow the guidelines developed for valve surgery and include structural and non-structural valve dysfunction, endocarditis, thrombo-embolism and bleeding events.

Secondary endpoints, particular to TAVI, should also include valve area and transvalvular gradient measurement, paravalvular leaks, device migration, emergency valve-in-valve implantation, conversion to conventional surgery, and vascular complications.

These trials should be performed according to the standards for clinical trials, i.e. central data collection and analysis, data and safety monitoring board, core laboratory analysis, and auditing and quality control, all respecting the rules of independence from sponsors.

TAVI should currently be performed only in clinical studies or as part of post-marketing registries. Clinical trials should be limited to centres with experience in TAVI and trials, and which actively participate in internal evaluation.

In centres performing TAVI, multi-disciplinary meetings should be held on a regular basis to discuss indications, procedural techniques, and case outcomes. Hospitals should keep proof of close medico-surgical collaboration and maintain a log of all patients referred for TAVI, regardless of final treatment strategy, for continuous evaluation of the programme.35 Furthermore, a series of thorough quality assurance measures should be included.

Conclusion

The currently available results obtained with TAVI suggest that these techniques are feasible and provide haemodynamic and clinical improvement for up to 2 years in patients with severe symptomatic AS at high risk or with contraindications for surgery. Pending questions concern mainly safety and long-term durability. Surgeons and cardiologists must work as a team to select the best candidates, perform the procedure, and, finally, evaluate the results. Today, these techniques are targeted at high-risk patients but they may be extended to the lower risk group in the future, if the initial promise holds true after careful evaluation. We are currently at the stage of evaluation, and a careful commercialization process including training and post-market surveillance is crucial to avoid the risk of uncontrolled diffusion, which is wholly undesirable at this stage of evolution of these techniques.

Conflict of interest. The following authors declared the following conflicts of interest: J.B., research grants from GE Healthcare, BMS Medical Imaging, St Jude, Medtronic, Boston Scientific; A.C., consultant for Edwards Lifesciences; J.K., consultant for Medtronic; F.M., consultant for Edwards Lifesciences and Medtronic, consultant and stockholder for Micardia; P.N., consultant for Sorin; J.L.P., consultant for Edwards Lifesciences and Boston Scientific; A.V., consultant for Edwards Lifesciences; B.v.H., scientific director on a cost effectiveness study concerning PVR sponsored by Edwards Lifesciences; L.K.v.S., founder and shareholder of Smartcanula LLC, Lausanne, Switzerland. All other authors have no conflicts of interest to declare.

Funding

Edwards Lifesciences provided an unrestricted grant for the practical organization of the meeting. No representatives of Edwards Lifesciences attended the meeting or were involved in the elaboration of the document.


References

Volume 4 Number 2
Aug 20, 2008
Volume 4 Number 2
View full issue


Key metrics

On the same subject

Editorial

10.4244/EIJ-E-24-00010 Apr 15, 2024
Timing of revascularisation in acute coronary syndromes with multivessel disease – two sides of the same coin
Stähli B and Stehli J
free

Editorial

10.4244/EIJ-E-24-00006 Apr 15, 2024
The miracle of left ventricular recovery after transcatheter aortic valve implantation
Dauerman H and Lahoud R
free

Research Correspondence

10.4244/EIJ-D-23-01046 Apr 15, 2024
Feasibility and safety of transcaval venoarterial extracorporeal membrane oxygenation in severe cardiogenic shock
Giustino G et al

State-of-the-Art

10.4244/EIJ-D-23-00836 Apr 15, 2024
Renal denervation in the management of hypertension
Lauder L et al
free

Original Research

10.4244/EIJ-D-23-00643 Apr 15, 2024
A study of ChatGPT in facilitating Heart Team decisions on severe aortic stenosis
Salihu A et al
Trending articles
338.63

State-of-the-Art Review

10.4244/EIJ-D-21-00904 Apr 1, 2022
Antiplatelet therapy after percutaneous coronary intervention
Angiolillo D et al
free
295.45

Expert consensus

10.4244/EIJ-D-21-00898 Sep 20, 2022
Intravascular ultrasound guidance for lower extremity arterial and venous interventions
Secemsky E et al
free
226.03

State-of-the-Art Review

10.4244/EIJ-D-21-00426 Dec 3, 2021
Myocardial infarction with non-obstructive coronary artery disease
Lindahl B et al
free
209.5

State-of-the-Art Review

10.4244/EIJ-D-21-01034 Jun 3, 2022
Management of in-stent restenosis
Alfonso F et al
free
168.4

Expert review

10.4244/EIJ-D-21-00690 May 15, 2022
Crush techniques for percutaneous coronary intervention of bifurcation lesions
Moroni F et al
free
149.53

State-of-the-Art

10.4244/EIJ-D-22-00776 Apr 3, 2023
Computed tomographic angiography in coronary artery disease
Serruys PW et al
free
103.48

Expert consensus

10.4244/EIJ-E-22-00018 Dec 4, 2023
Definitions and Standardized Endpoints for Treatment of Coronary Bifurcations
Lunardi M et al
free
X

The Official Journal of EuroPCR and the European Association of Percutaneous Cardiovascular Interventions (EAPCI)

EuroPCR EAPCI
PCR ESC
Impact factor: 6.2
2022 Journal Citation Reports®
Science Edition (Clarivate Analytics, 2023)
Online ISSN 1969-6213 - Print ISSN 1774-024X
© 2005-2024 Europa Group - All rights reserved