1. Drug-eluting and bare metal stents

SVELTE Fixed Wire SES system for treatment of CAD

OPTIMIZE
Objective
to report the clinical efficacy of SVELTE SES system compared with Xience or Promus EES for the treatment of coronary artery obstructions
Study
multicentre, randomised non-inferiority trial (3.58% margin)
Population
patients with non-STEMI or stable CAD
Endpoints
target lesion failure as cardiac death, MI or clinically indicated target lesion revascularisation at 12 months
Conclusion
SVELTE – SES did not meet the prespecified threshold for non- inferiority compared to EES at 1 year
Kereiakes et al. Circulation Cardiovascular Interventions. 2021;14:e010609
Receive our newsletter


The Official Journal of EuroPCR and the European Association of Percutaneous Cardiovascular Interventions (EAPCI)

EuroPCR EAPCI
PCR ESC
Impact factor: 6.2
2022 Journal Citation Reports®
Science Edition (Clarivate Analytics, 2023)
Online ISSN 1969-6213 - Print ISSN 1774-024X
© 2005-2024 Europa Group - All rights reserved